Skip to content

NexGenRecalls & reported problems across all models

Zimmer, Inc. · 198 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

1 open recall names at least one NexGen model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
842
Deaths reported
0
Injuries reported
600
Recalls
4

We found 198 models of NexGen by Zimmer, Inc. in the FDA's device registry (GUDID). Together they are named in 842 adverse event reports in the FDA's MAUDE database, received between Sep 10, 2019 and Sep 30, 2026 — classified as 600 injuries, 242 malfunctions.

The most frequently reported specific device problem is “Fracture” (207 reports).

4 FDA recall records name at least one of these models by its device identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 6 reportsDec 2023: 15 reportsJan 2024: 5 reportsFeb 2024: 8 reportsMar 2024: 17 reportsApr 2024: 24 reportsMay 2024: 10 reportsJun 2024: 20 reportsJul 2024: 13 reportsAug 2024: 29 reportsSep 2024: 16 reportsOct 2024: 24 reportsNov 2024: 19 reportsDec 2024: 8 reportsJan 2025: 12 reportsFeb 2025: 12 reportsMar 2025: 15 reportsApr 2025: 6 reportsMay 2025: 10 reportsJun 2025: 6 reportsJul 2025: 10 reportsAug 2025: 9 reportsSep 2025: 7 reportsOct 2025: 24 reportsNov 2025: 13 reportsDec 2025: 33 reportsJan 2026: 6 reportsFeb 2026: 14 reportsMar 2026: 6 reportsApr 2026: 5 reportsMay 2026: 0 reportsJun 2026: 1 reportsJul 2026: 4 reportsAug 2026: 10 reportsSep 2026: 10 reportsOct 2026: 0 reports20242025202633
Last 36 months. The most recent month may be incomplete.

NexGen recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Mar 28, 2024Open, Classified

    Out of specification violation of devices that results in a gap existing within the fixture that presses the fiber metal pad to the substrate, causing the pad to not fully bond to the substrate.

    Z-1608-2024 · Root cause: Process change control · names 4 of these models · Recall details · FDA record ↗

  2. May 19, 2023Terminated

    Device outer packaging was incorrectly labeled.

    Z-2092-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗

  3. Oct 31, 2018Terminated

    Screw is missing from the package

    Z-0630-2019 · Root cause: Process control · Recall details · FDA record ↗

  4. Apr 28, 2017Terminated

    Zimmer Biomet is conducting a lot-specific medical device field action for two lots of the NexGen Offset Stem Extension implants. Product complaints indicated that the size etched on the part and labeled on the product packaging were incorrect.

    Z-2115-2017 · Root cause: Under Investigation by firm · names 2 of these models · Recall details · FDA record ↗

Most reported NexGen problems

Device problems

  • Adverse Event Without Identified Device or Use Problem280
  • Fracture207
  • Loosening of Implant Not Related to Bone-Ingrowth198
  • Insufficient Device Problem Information34
  • Migration31
  • Difficult to Insert23
  • Mechanical Jam23
  • Detachment of Device or Device Component20

Patient problems

  • Pain324
  • No Clinical Signs, Symptoms or Conditions217
  • Failure of Implant179
  • Swelling/ Edema110
  • Ambulation or Postural Difficulties80
  • Aseptic loosening63
  • Insufficient Information63
  • Loss of Range of Motion55

NexGen models

198 models, 100 per page, most reported first. Each model has its own page with its reports, recalls and device details.

Recent NexGen adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Zimmer, Inc. products

All Zimmer, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page