NexGen® 5880-30-12Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of NexGen (198)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 3
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00889024199248), received between Apr 27, 2020 and Nov 17, 2025 — classified as 3 injuries.
The most frequently reported specific device problem is “Fracture” (2 reports).
4 other recalls cover devices cleared under the same FDA submission (K013385, K173057); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture2
- Device Dislodged or Dislocated1
- Loosening of Implant Not Related to Bone-Ingrowth1
- Noise, Audible1
- Unstable1
Patient problems
- Failure of Implant2
- Pain2
- Joint Dislocation1
- Loss of Range of Motion1
Recent adverse event reports
The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.
InjuryNov 17, 2025Fracture; Fracture
NO FURTHER EVENT INFORMATION AT TIME OF THIS REPORT.
Patient problems: Failure of Implant; Loss of Range of Motion; Pain
MAUDE report 23572276 ↗InjuryDec 15, 2022Fracture; Fracture
IT WAS REPORTED THAT PATIENT WHO UNDERWENT A PREVIOUS REVISION STARTED TO FEEL A PAIN APPROXIMATELY 1 YEAR 7 MONTHS POST REVISION. AT THE DIAGNOSIS, A NON-IMPLANT RELATED TUMOR OUT OF JOINT WAS SUSPECTED. A PROCEDURE WAS PERFORMED TO REMOVE THE TUMOR AND ADMINISTRATE ANTI-BIOTICS. IT WAS FOUND AT THAT TIME THE HINGE POST WAS FRACTURED AND WILL NEED REVISION TO CORRECT. NO PATIENT INFORMATION AVAILABLE.
Patient problems: Failure of Implant; Pain
MAUDE report 15990637 ↗InjuryApr 27, 2020Unstable; Device Dislodged or Dislocated; Noise, Audible; Loosening of Implant Not Related to Bone-Ingrowth; Unstable; Device Dislodged or Dislocated; Noise, Audible; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED PATIENT UNDERWENT INITIAL LEFT ROTATING HINGE TOTAL KNEE ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT UNDERWENT IRRIGATION AND DEBRIDEMENT WITH POLY EXCHANGE TWO WEEKS LATER DUE TO WOUND COMPLICATIONS. WITHOUT A KNOWN TRAUMATIC EVENT, THE PATIENT EXPERIENCED CRACKING AND INSTABILITY IN THE JOINT AND PRESENTED TO THE ER. X-RAYS SHOWED PROSTHESIS LUXATION WITH A LOOSENED COUPLING MECHANISM. THE PATIENT UNDERWENT A SECOND REVISION DURING WHICH THE POLY WAS EXCHANGED WITHOUT COMPLICATION APPROXIMATELY TWO MONTHS POST IMPLANTATION. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE.
Patient problems: Joint Dislocation
MAUDE report 10000571 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K013385, K173057. One submission often covers many models, so check the product and lot details in each record.
- Z-2092-2023 · May 19, 2023 · TerminatedNexGen Precoat Stemmed Tibial Plate Size 5
- Z-0321-2020 · Oct 10, 2019 · TerminatedVarious Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800
- Z-1448-04 · Aug 3, 2004 · TerminatedZimmer brand NEXGEN Complete Knee Solution Rotating Hinge Knee Tibial Component, precoat, nonmodular, size 3, sterile, for cemented use only; Cat. No. 5880-03-02.
- Z-1447-04 · Aug 3, 2004 · TerminatedZimmer brand NEXGEN Complete Knee Solution Rotating Hinge Knee Tibial Component, precoat, nonmodular, size 2, sterile, for cemented use only; Cat. No. 5880-02-02.
Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices
- ELEOS™ 25001208EOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER3 deaths193 reports
- ELEOS™ THSMWRS01MOnkos Surgical, Inc. · PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER190 reports
- "UNITED" USTAR II Hip System 1115-1413United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls3 reports
- "UNITED" USTAR II Hip System 1115-1213United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1417United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- "UNITED" USTAR II Hip System 1115-1217United Orthopedic Corp. · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer3 recalls0 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/12WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
- Endo-Model Modular Knee Prosthesis System 15-3816/11WALDEMAR LINK GmbH & Co. KG · Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer2 recalls6 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.