Skip to content

NexGen® 5880-60-12Recalls & reported problems

Zimmer, Inc. · Class 2 · Implant

All models of NexGen (198)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
8
Deaths reported
0
Injuries reported
8
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00889024199422), received between Apr 29, 2020 and Aug 5, 2026 — classified as 8 injuries.

The most frequently reported specific device problem is “Detachment of Device or Device Component” (2 reports).

4 other recalls cover devices cleared under the same FDA submission (K013385, K173057); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 1 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 1 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 1 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Adverse Event Without Identified Device or Use Problem4
  • Detachment of Device or Device Component2
  • Device Dislodged or Dislocated1
  • Fracture1
  • Loosening of Implant Not Related to Bone-Ingrowth1

Patient problems

  • Pain4
  • Failure of Implant3
  • Joint Laxity2
  • Ambulation or Postural Difficulties1
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
  • Fall1
  • Fever1
  • Fluid Discharge1

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • InjuryAug 5, 2026Adverse Event Without Identified Device or Use Problem

    It was reported the patient underwent a right total knee arthroplasty and was subsequently revised approximately 1 year and 6 months post-implantation due to instability and mechanical loosening. Subsequently, 3 months after the revision, the patient began experiencing pain, swelling, stiffness, fever in the knee, and limitations in normal activities following the procedure. The devices remain implanted. Attempts have been made and no further information has been provided.

    Patient problems: Ambulation or Postural Difficulties; Pain; Swelling/ Edema

    MAUDE report 50143753 ↗
  • InjuryDec 22, 2025Device Dislodged or Dislocated; Device Dislodged or Dislocated

    IT WAS REPORTED THAT A PATIENT UNDERWENT A TOTAL KNEE ARTHROPLASTY. APPROXIMATELY FIVE (5) MONTHS POST-IMPLANTATION, THE COUPLED KNEE PROSTHESIS HAD BECOME UNCOUPLED, RESULTING IN KNEE DISLOCATION. THE FEMORAL, TIBIAL, AND RETROPATELLAR COMPONENTS WERE FOUND TO BE WELL-ANCHORED AND INTACT HOWEVER THE AXLE BOLT HAD DISENGAGED FROM THE RETAINING BUSHING. SUBSEQUENTLY, THE PATIENT UNDERWENT REVISION SURGERY TO REPLACE THE INLAY AND LOOSE AXIS WITHOUT COMPLICATION. DUE DILIGENCE IS COMPLETE AS MULTIPLE ATTEMPTS HAVE BEEN MADE HOWEVER ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: Joint Dislocation; Joint Laxity; Pain

    MAUDE report 23874073 ↗
  • InjuryJun 12, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL RIGHT KNEE SURGERY ON AN UNKNOWN DATE. SUBSEQUENTLY, THE PATIENT FELL AND THE RHK AXLE PIN CAME LOOSE AND THE PATIENT WAS REVISED. THE AXLE PIN AND INLAY WERE REPLACED. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.

    Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Failure of Implant; Fall

    MAUDE report 22197468 ↗
  • InjuryApr 29, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THAT A PATIENT HAD AN INITIAL LEFT TOTAL KNEE ARTHROPLASTY WITH COMPETITOR PRODUCT WHICH WAS FULLY REVISED SIX YEARS POST IMPLANTATION DUE TO OSTEOLYSIS AND ASEPTIC LOOSENING. SUBSEQUENTLY, THE PATIENT REPORTED PAIN, FEVER, AND DRAINAGE FROM THE INCISION SITE THREE WEEKS POSTOP. THE PATIENT UNDERWENT AN UNCOMPLICATED INCISION AND DRAINAGE WITH POLY EXCHANGE. CULTURES WERE OBTAINED, BUT NO RESULTS HAVE BEEN PROVIDED.

    Patient problems: Fever; Fluid Discharge; Pain; Wound Dehiscence

    MAUDE report 21926502 ↗
  • InjurySep 11, 2024Fracture; Fracture

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant

    MAUDE report 20193138 ↗
  • InjuryJun 7, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem

    IT WAS REPORTED THE PATIENT UNDERWENT A RIGHT KNEE REVISION APPROXIMATELY FIVE (5) YEARS POST-IMPLANTATION DUE TO PAIN AND LAXITY. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Joint Laxity; Pain

    MAUDE report 19483545 ↗
  • InjuryDec 13, 2023Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth

    IT WAS REPORTED PATIENT UNDERWENT A REVISION PROCEDURE POST IMPLANTATION DUE TO UNKNOWN REASON. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant; Insufficient Information

    MAUDE report 18317787 ↗
  • InjuryApr 29, 2020Detachment of Device or Device Component; Detachment of Device or Device Component

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: No Code Available

    MAUDE report 10008862 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K013385, K173057. One submission often covers many models, so check the product and lot details in each record.

  • Z-2092-2023 · May 19, 2023 · Terminated
    NexGen Precoat Stemmed Tibial Plate Size 5
  • Z-0321-2020 · Oct 10, 2019 · Terminated
    Various Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800
  • Z-1447-04 · Aug 3, 2004 · Terminated
    Zimmer brand NEXGEN Complete Knee Solution Rotating Hinge Knee Tibial Component, precoat, nonmodular, size 2, sterile, for cemented use only; Cat. No. 5880-02-02.
  • Z-1448-04 · Aug 3, 2004 · Terminated
    Zimmer brand NEXGEN Complete Knee Solution Rotating Hinge Knee Tibial Component, precoat, nonmodular, size 3, sterile, for cemented use only; Cat. No. 5880-03-02.

Other PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices

All PROSTHESIS, KNEE, FEMOROTIBIAL, CONSTRAINED, CEMENTED, METAL/POLYMER devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page