NexGen® 5980-47-01Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of NexGen (198)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 8
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00889024217607), received between Oct 1, 2019 and Dec 8, 2025 — classified as 8 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Difficult to Insert” (3 reports).
1 FDA recall record name this device's identifier.
84 other recalls cover devices cleared under the same FDA submission (K173057, K933785); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Recalls naming this device
- May 19, 2023Terminated
Device outer packaging was incorrectly labeled.
Z-2092-2023 · Root cause: Under Investigation by firm · Recall details · FDA record ↗
Most reported problems
Device problems
- Adverse Event Without Identified Device or Use Problem3
- Difficult to Insert3
- Loosening of Implant Not Related to Bone-Ingrowth3
- Fracture2
- Detachment of Device or Device Component1
- Loss of or Failure to Bond1
- Mechanical Problem1
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions4
- Pain4
- Damage to Ligament(s)2
- Failure of Implant2
- Ambulation or Postural Difficulties1
- Blister1
- Loss of Range of Motion1
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
InjuryDec 8, 2025Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Blister; Rash
MAUDE report 23744918 ↗MalfunctionMar 27, 2025Mechanical Problem; Mechanical Problem
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21706702 ↗MalfunctionApr 7, 2023Difficult to Insert; Difficult to Insert
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16701396 ↗InjurySep 12, 2022Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth; Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY TWO YEARS ONE AND A HALF MONTHS POST IMPLANTATION DUE TO PATIENT COMORBIDITIES, LIMITED RANGE OF MOTION, PAIN, SWELLING, AND LOOSENING. THE TIBIA COMPONENTS WERE REVISED WITH DIFFICULTY AND THE OSTEOTOMY WITH TIBIAL TUBERCLE WAS REAPPROXIMATED WITH CABLE WIRES. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE INFORMATION IS AVAILABLE.
Patient problems: Loss of Range of Motion; Pain; Swelling/ Edema
MAUDE report 15402920 ↗MalfunctionMar 9, 2022Positioning Failure; Difficult to Insert; Positioning Failure; Difficult to Insert
UPON FURTHER INVESTIGATION, IT WAS DETERMINED THAT THIS ITEM WAS REPORTED IN ERROR. THE ARTICULAR SURFACE IS DOVETAILED. THIS IS A USER ERROR WITH THIS PART. TIBIAL PLATE WAS ABLE TO ASSEMBLE WITH A DIFFERENT DOVE TAIL. THE INITIAL REPORT WAS SUBMITTED IN ERROR AND SHOULD BE VOIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13713732 ↗InjuryOct 29, 2021Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 12719992 ↗MalfunctionOct 20, 2021Packaging Problem; Packaging Problem
AFTER OPENING THE PROPERLY SEALED EXTERNAL PACKAGING, IT WAS NOTED THE IMPLANT WAS NOT CORRECTLY CONTAINED INSIDE THE INNER STERILE PACKAGING. THERE WAS NO PATIENT INVOLVEMENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12669775 ↗InjuryJun 18, 2021Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth; Adverse Event Without Identified Device or Use Problem; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THE PATIENT UNDERWENT A LEFT KNEE REVISION APPROXIMATELY FOUR YEARS POST-OPERATIVELY DUE TO TIBIAL PAIN AND POSSIBLE LOOSENING. ATTEMPTS TO OBTAIN ADDITIONAL INFORMATION HAVE BEEN MADE; HOWEVER, NO MORE IS AVAILABLE.
Patient problems: Damage to Ligament(s); Failure of Implant; Pain
MAUDE report 12024630 ↗InjuryJun 10, 2021Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
UPON REASSESSMENT OF THE REPORTED EVENT, IT WAS IDENTIFIED THAT THIS DEVICE DID NOT CAUSE OR CONTRIBUTE TO THE REPORTED EVENT. THIS ISSUE HAS BEEN REPORTED UNDER 3007963827-2021-00129 AND 0001822565-2021-01619.
Patient problems: Damage to Ligament(s)
MAUDE report 11976614 ↗InjuryJun 9, 2021Detachment of Device or Device Component; Detachment of Device or Device Component
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Unspecified Infection
MAUDE report 11967746 ↗MalfunctionDec 17, 2020Difficult to Insert; Difficult to Insert
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 11029364 ↗InjuryFeb 3, 2020Fracture; Unstable; Fracture; Unstable
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Pain
MAUDE report 9661058 ↗InjuryOct 1, 2019Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT THAT THE PATIENT WAS REVISED DUE TO IMPLANT FRACTURE APPROXIMATELY 2 YEARS POST IMPLANTATION. THE PATIENT STARTED EXPERIENCING SEVERE PAIN IN HIS KNEE AND WAS UNABLE TO WALK. NO ADDITIONAL PATIENT CONSEQUENCES WERE REPORTED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Ambulation or Postural Difficulties; Osteopenia/ Osteoporosis; Pain
MAUDE report 9142095 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K173057, K933785. One submission often covers many models, so check the product and lot details in each record.
- Z-0321-2020 · Oct 10, 2019 · TerminatedVarious Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800 0052200
- Z-1042-2018 · Nov 29, 2017 · TerminatedCR OPTION FEM COMP SIZE C/L CR OPTION FEM COMP SIZE C/R CR OPTION FEM COMP SIZE D/L CR OPTION FEM COMP SIZE D/R CR OPTION FEM COMP SIZE E/L CR OPTION FEM COMP SIZE E/R CR OPTION FE
- Z-1039-2018 · Nov 29, 2017 · TerminatedLPS PRECOAT SRF HDN FEM SZ D-L LPS PRECOAT SRF HDN FEM SZ D-R LPS PRECOAT SRF HDN FEM SZ E-L LPS POROUS SRF HDN FEM SZ D-LT LPS POROUS SRF HDN FEM SZ E-LT This device is indicated
- Z-1038-2018 · Nov 29, 2017 · TerminatedLPS POROUS FEM IMPLANT SZ G-L LPS PRECOAT FEM IMPLANT SZ G-L This device is indicated for patients with severe knee pain and disability
- Z-1040-2018 · Nov 29, 2017 · TerminatedLPS PRECOAT FEM SIZE D-RT LPS PRECOAT FEM SIZE E-RT LPS POROUS FEM SIZE F-RT This device is indicated for patients with severe knee pain and disability
- Z-1041-2018 · Nov 29, 2017 · TerminatedCR PRECOAT FEM COMP SIZE BMR CR PRECOAT FEM COMP SIZE C/R CR PRECOAT FEM COMP SIZE D/R CR PRECOAT FEM COMP SIZE E/R CR PRECOAT FEM COMP SIZE F/R CR PRECOAT FEM COMP SIZE G/L CR PRE
- Z-2115-2017 · Apr 28, 2017 · TerminatedNexGen Offset Stem Extension, sizes 14mm and 17mm For use in total arthroplasty.
- Z-1159-2016 · Jan 11, 2016 · TerminatedProduct 61 consists of all product under product code: JWH and same usage: Item no: 598003712 STD PRECOAT TIBIA PLATE S for use in total knee arthroplasty
- Z-1147-2016 · Jan 11, 2016 · TerminatedProduct 49 consists of all product under product code: JWH and same usage: Item no: 597206529 NEXGEN ALL-POLY PATELLA, 597206532 NEXGEN ALL-POLY PATELLA, 597206535 NEXGEN ALL-POL
- Z-1151-2016 · Jan 11, 2016 · TerminatedProduct 53 consists of all product under: product code: JWH and same usage: Item no: 597002501 NEXGEN PRECOAT PEG TIB PL 597002502 NEXGEN PRECOAT PEG TIB PL 597003501 NEXGEN PR
- Z-0835-2016 · Jan 11, 2016 · TerminatedNexGen Knee CR Option Fem NexGen Knee CR Porous Fem NexGen Knee CR Precoat Fem NexGen Knee LPS Porous Fem NexGen Knee LPS Precoat Fem prosthesis, knee, patellofemorotibial, semi-co
- Z-0590-2016 · Dec 3, 2015 · TerminatedNexGEN System REF 5980-37-01 Tibial Component Size Stemmed 3 Precoat, and REF 5986-37-01 Nonaugmentable Tibial Component Option CR/PS/LPS Size Stemmed 3. Tibial component of im
- Z-0449-2016 · Dec 2, 2015 · TerminatedLabel reading in part as: "NexGen System, Complete Knee Solution; Reference Number 5980-37-01; Tibial Component, Precoat, Stemmed Size 3; Lot Number:62460264. Indicated for patie
- Z-2552-2014 · Jul 22, 2014 · TerminatedKnee Stem Extensions & Impactor Sleeve: STEM IMPLANT 20MMDX145MM STEM IMPLANT 22MMDX145MM STEM IMPLANT 24MMDX145MM SHARP FLUTED ST 20MMDX75MM SHARP FLUTED ST 22MMDX75MM SHARP FLTD
- Z-0480-2014 · Oct 31, 2013 · TerminatedNexGen Complete Knee Solutions Stemmed Tibial Component Precoat NexGen Tibial Baseplates are part of the NexGen systems of semi-constrained, non-linked condylar knee prosthesis. T
…and 69 more.
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page