NexGen® 5986-47-02Recalls & reported problems
Zimmer, Inc. · Class 2 · Implant
All models of NexGen (198)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 16
- Deaths reported
- 0
- Injuries reported
- 15
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 16 adverse event reports that list this exact device (DI 00889024218864), received between Jan 14, 2021 and Apr 1, 2026 — classified as 15 injuries, 1 malfunction.
The most frequently reported specific device problem is “Loosening of Implant Not Related to Bone-Ingrowth” (13 reports).
95 other recalls cover devices cleared under the same FDA submission (K173057, K872379, K933785); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Loosening of Implant Not Related to Bone-Ingrowth13
- Insufficient Device Problem Information2
- Adverse Event Without Identified Device or Use Problem1
- Fracture1
- Loss of or Failure to Bond1
Patient problems
- Pain11
- Aseptic loosening7
- Failure of Implant7
- Ambulation or Postural Difficulties2
- Insufficient Information2
- Swelling/ Edema2
- Bone Fracture(s)1
- Erosion1
Recent adverse event reports
The 16 most recent of 16 reports, in the reporter's own words as published by the FDA.
InjuryApr 1, 2026Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THE PATIENT UNDERWENT A RIGHT KNEE ARTHROPLASTY REVISION TO ADDRESS PAIN AND TIBIAL LOOSENING APPROXIMATELY FIVE (5) YEARS POST-OPERATIVELY. DURING THE PROCEDURE, THE TIBIAL PLATE WAS FOUND TO BE GROSSLY LOOSE WITH NO CEMENT ADHERED TO THE TITANIUM SURFACE. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: Aseptic loosening; Pain
MAUDE report 24758596 ↗InjuryMar 20, 2026Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Aseptic loosening; Inadequate Osseointegration; Pain
MAUDE report 24649411 ↗InjuryMar 16, 2026Insufficient Device Problem Information; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Aseptic loosening; Insufficient Information; Pain
MAUDE report 24608712 ↗InjuryFeb 8, 2026Loosening of Implant Not Related to Bone-Ingrowth
NO ADDITIONAL INFORMATION TO REPORT.
Patient problems: Aseptic loosening; Pain
MAUDE report 24288358 ↗InjuryJan 14, 2026Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Aseptic loosening; Osteolysis
MAUDE report 24072731 ↗InjuryNov 10, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Aseptic loosening; Pain
MAUDE report 23510024 ↗InjuryJun 12, 2025Insufficient Device Problem Information; Insufficient Device Problem Information
No further event information available at the time of this report.
Patient problems: Insufficient Information
MAUDE report 22193245 ↗InjuryFeb 27, 2025Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT A PATIENT UNDERWENT AN INITIAL TOTAL KNEE ARTHROPLASTY. APPROXIMATELY THREE (3) YEARS AND THREE (3) MONTHS POST-IMPLANTATION, THE PATIENT BEGAN TO EXPERIENCE PAIN, SWELLING, INSTABILITY, AND DIFFICULT AMBULATING. A DIAGNOSTIC BONE SCAN AND LABS REVEALED RESULTS CONSISTENT WITH LOOSENING OF THE TIBIAL COMPONENT. SUBSEQUENTLY, A REVISION SURGERY WAS PERFORMED TWO (2) MONTHS FOLLOWING THE ONSET OF SYMPTOMS. THE TIBIAL COMPONENT WAS FOUND GROSSLY LOOSE WITH NO ADHESION OF THE CEMENT TO THE IMPLANT. THE FEMORAL COMPONENT AND PATELLA WERE FOUND TO BE WELL FIXED HOWEVER ALL COMPONENTS ASIDE FROM THE PATELLA WERE REVISED. IT WAS REPORTED THAT ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.
Patient problems: Ambulation or Postural Difficulties; Aseptic loosening; Failure of Implant; Pain; Swelling/ Edema
MAUDE report 21479763 ↗InjuryDec 30, 2024Adverse Event Without Identified Device or Use Problem; Adverse Event Without Identified Device or Use Problem
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Impaired Healing
MAUDE report 21034697 ↗InjuryNov 25, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT THE PATIENT HAD A BILATERAL KNEE ARTHROPLASTY. SUBSEQUENTLY, THE PATIENT REPORTED RIGHT KNEE PAIN AND NOISE. THE PATIENT HAD X-RAYS TAKEN AND IT WAS REPORTED THAT THE PATIENT NEEDED A RIGHT KNEE REVISION. APPROXIMATELY 4 YEARS POST-OP, THE PATIENT WAS REVISED DUE TO TIBIAL LOOSENING. DURING THE REVISION THERE WAS A BONE DEFECT NOTED ON THE MEDIAL CONDYLE OF THE TIBIA. THE PATELLA WAS FOUND TO BE IN STABLE CONDITION AND WAS RETAINED. ALL OTHER COMPONENTS WERE REVISED. ATTEMPTS HAVE BEEN MADE AND ALL AVAILABLE INFORMATION HAS BEEN PROVIDED.
Patient problems: Erosion; Failure of Implant; Pain
MAUDE report 20775267 ↗InjuryNov 20, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Joint effusion; Pain; Swelling/ Edema
MAUDE report 20735250 ↗InjuryOct 10, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 20418426 ↗InjurySep 9, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Ambulation or Postural Difficulties; Bone Fracture(s); Extraskeletal ossification; Failure of Implant; Necrosis; Pain; Synovitis
MAUDE report 20171809 ↗MalfunctionJun 4, 2024Loosening of Implant Not Related to Bone-Ingrowth; Loosening of Implant Not Related to Bone-Ingrowth
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant; Pain
MAUDE report 19457567 ↗InjuryAug 23, 2021Fracture; Fracture
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL KNEE ARTHROPLASTY ON AN UNKNOWN DATE. SUBSEQUENTLY, THE PATIENT UNDERWENT REVISION DUE TO A BROKEN TIBIAL PLATE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 12353781 ↗InjuryJan 14, 2021Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth; Loss of or Failure to Bond; Loosening of Implant Not Related to Bone-Ingrowth
IT WAS REPORTED THAT A PATIENT WAS REVISED APPROXIMATELY FIVE YEARS AND ONE MONTH POST IMPLANTATION DUE TO PAIN CAUSED BY TIBIAL LOOSENING. THERE IS NO ADDITIONAL INFORMATION AVAILABLE AT THIS TIME.
Patient problems: Pain
MAUDE report 11174747 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K173057, K872379, K933785. One submission often covers many models, so check the product and lot details in each record.
- Z-2092-2023 · May 19, 2023 · TerminatedNexGen Precoat Stemmed Tibial Plate Size 5
- Z-0469-2023 · Dec 6, 2022 · Open, ClassifiedNexGen Option Stemmed Tibial Component Size 6, Item Number 00-5986-047-02
- Z-0471-2023 · Dec 6, 2022 · Open, ClassifiedNexGen Option Stemmed Tibial Component Size 8, Item Number 00-5986-057-02
- Z-0468-2023 · Dec 6, 2022 · Open, ClassifiedNexGen Option Stemmed Tibial Component Size 5, Item Number 00-5986-047-01
- Z-0470-2023 · Dec 6, 2022 · Open, ClassifiedNexGen Option Stemmed Tibial Component Size 7, Item Number 00-5986-057-01
- Z-0467-2023 · Dec 6, 2022 · Open, ClassifiedNexGen Option Stemmed Tibial Component Size 4, Item Number 00-5986-037-02
- Z-0466-2023 · Dec 6, 2022 · Open, ClassifiedNexGen Option Stemmed Tibial Component Size 3, Item Number 00-5986-037-01
- Z-0321-2020 · Oct 10, 2019 · TerminatedVarious Knee Systems for knee joint arthroplasty, knee prosthesis, Item Nos. 00522000100 00522000200 00522000300 00522000400 00522006304 00522006504 00522006600 00522006800
- Z-1042-2018 · Nov 29, 2017 · TerminatedCR OPTION FEM COMP SIZE C/L CR OPTION FEM COMP SIZE C/R CR OPTION FEM COMP SIZE D/L CR OPTION FEM COMP SIZE D/R CR OPTION FEM COMP SIZE E/L CR OPTION FEM COMP SIZE E/R CR OPTION FE
- Z-1041-2018 · Nov 29, 2017 · TerminatedCR PRECOAT FEM COMP SIZE BMR CR PRECOAT FEM COMP SIZE C/R CR PRECOAT FEM COMP SIZE D/R CR PRECOAT FEM COMP SIZE E/R CR PRECOAT FEM COMP SIZE F/R CR PRECOAT FEM COMP SIZE G/L CR PRE
- Z-1039-2018 · Nov 29, 2017 · TerminatedLPS PRECOAT SRF HDN FEM SZ D-L LPS PRECOAT SRF HDN FEM SZ D-R LPS PRECOAT SRF HDN FEM SZ E-L LPS POROUS SRF HDN FEM SZ D-LT LPS POROUS SRF HDN FEM SZ E-LT This device is indicated
- Z-1040-2018 · Nov 29, 2017 · TerminatedLPS PRECOAT FEM SIZE D-RT LPS PRECOAT FEM SIZE E-RT LPS POROUS FEM SIZE F-RT This device is indicated for patients with severe knee pain and disability
- Z-1038-2018 · Nov 29, 2017 · TerminatedLPS POROUS FEM IMPLANT SZ G-L LPS PRECOAT FEM IMPLANT SZ G-L This device is indicated for patients with severe knee pain and disability
- Z-2115-2017 · Apr 28, 2017 · TerminatedNexGen Offset Stem Extension, sizes 14mm and 17mm For use in total arthroplasty.
- Z-1151-2016 · Jan 11, 2016 · TerminatedProduct 53 consists of all product under: product code: JWH and same usage: Item no: 597002501 NEXGEN PRECOAT PEG TIB PL 597002502 NEXGEN PRECOAT PEG TIB PL 597003501 NEXGEN PR
…and 80 more.
Other PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER devices
- ATTUNE 1506-00-005Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER589 reports
- ATTUNE 1506-00-004Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER580 reports
- JOURNEY 74012812Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER502 reports
- JOURNEY 74011856Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER438 reports
- PERSONA™ 42-5099-094-00Zimmer, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER406 reports
- JOURNEY 74011857Smith & Nephew, Inc. · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER388 reports
- ATTUNE 1506-00-007Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER288 reports
- ATTUNE 1506-00-003Depuy (Ireland) · PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL/POLYMER271 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page