Zimmer® Natural Nail™ System 00-2490-003-08Recalls & reported problems
Zimmer GmbH · CM TARGETING GUIDE - SMALL · Class 2
All models of Zimmer Natural Nail System (57)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00889024276239), received between Jul 21, 2023 and Jul 21, 2023 — classified as 1 injury.
The most frequently reported specific device problem is “Break” (1 report).
5 other recalls cover devices cleared under the same FDA submission (K091566, K231114); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break1
Patient problems
- Foreign Body In Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
InjuryJul 21, 2023Break; Break
IT WAS REPORTED THAT A LAG SCREW REAMER FRACTURED WHILE USING IT WITH A DAMAGED TARGETING GUIDE. THE PIECE OF FRACTURED LAG SCREW REAMER WAS RETAINED IN THE PATIENT¿S FEMORAL HEAD. NO FURTHER PATIENT IMPACT HAS BEEN REPORTED. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.
Patient problems: Foreign Body In Patient
MAUDE report 17369631 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K091566, K231114. One submission often covers many models, so check the product and lot details in each record.
- Z-1793-2021 · May 7, 2021 · TerminatedZimmer Natural Nail- ZNN Cephalomedullary Short Nail 11.5 mm X 21.5 cm 130 CCD Left Ti-6Al-4V Alloy-is intended for temporary fracture fixation and stabilization of the bone Item
- Z-1792-2021 · May 7, 2021 · TerminatedZimmer Natural Nail -ZNN Cephalomedullary Short Nail 11.5 mm X 21.5 cm 125 CCD Left Ti-6Al-4V Alloy- intended for temporary fracture fixation and stabilization of the bone Item Nu
- Z-0337-2020 · Oct 10, 2019 · TerminatedCephalomedullary Nail (CMN), Various sizes, Item Nos. 47249318009 47249318010 47249318011 47249318013 47249318014 47249318109 47249318110 47249318111 47249318113 472493181
- Z-0119-2016 · Sep 3, 2015 · TerminatedZimmer Natural Nail CM Long. Intramedullary Fixation Rod. The Zimmer Natural Nail System is intended for temporary fracture fixation and stabilization of the bone.
- Z-0824-2011 · Oct 29, 2010 · TerminatedZimmer Natural Nail System Cephalomedullary Lag Screw Inserter Long, Catalog 00-2490-003-50, Zimmer, Warsaw, IN. A reusable surgical instrument used for the insertion of a Lag Scr
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page