Skip to content

Zimmer® Natural Nail® System 47-2485-090-10Recalls & reported problems

Zimmer GmbH · ZNN CMN LAG SCREW 10.5X90 · Class 2 · Implant

All models of Zimmer Natural Nail System (57)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
10
Deaths reported
0
Injuries reported
10
Recalls naming it
0

Summary

The FDA's MAUDE database contains 10 adverse event reports that list this exact device (DI 00889024298439), received between Dec 19, 2018 and Mar 20, 2025 — classified as 10 injuries.

The most frequently reported specific device problem is “Fracture” (7 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 1 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture7
  • Insufficient Device Problem Information2
  • Appropriate Device Problem Term/Code Not Available1
  • Break1
  • Separation Problem1

Patient problems

  • Failure of Implant3
  • Bone Fracture(s)2
  • No Code Available2
  • Extraskeletal ossification1
  • Foreign Body In Patient1
  • No Clinical Signs, Symptoms or Conditions1
  • No Information1
  • Pain1

Recent adverse event reports

The 10 most recent of 10 reports, in the reporter's own words as published by the FDA.

  • InjuryMar 20, 2025Fracture; Fracture

    DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Foreign Body In Patient

    MAUDE report 21653766 ↗
  • InjuryMay 23, 2024Separation Problem; Separation Problem

    IT WAS REPORTED THAT PATIENT HAD AN INITIAL OPEN REDUCTION INTERNAL FIXATION ON THE LEFT SIDE. SUBSEQUENTLY, APPROXIMATELY 8 MONTHS POST-IMPLANTATION, REVISION WAS NEEDED DUE TO EXIT OF THE CERVICAL SCREW FROM THE JOINT AND SINKING INTO THE ACETABULUM. DURING THE REVISION IT WAS NOTED THAT THE ACETABULAR ROOF WAS FRACTURED. ALL COMPONENTS WERE EXCHANGED WITH COMPETITOR PRODUCT. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: Bone Fracture(s); Failure of Implant; Pain

    MAUDE report 19383577 ↗
  • InjuryMar 3, 2022Fracture; Fracture

    IT WAS REPORTED THAT: AFTER 2 MONTHS OF IMPLANTATION THERE WAS A BREAK IN THE FITTING OF THE PROXIMAL FIXATION SCREW OF THE FEMORAL NECK TOGETHER WITH THE FEMORAL NAIL, WHICH RESULTED A REVISION.

    Patient problems: Bone Fracture(s); Failure of Implant

    MAUDE report 13658386 ↗
  • InjuryFeb 3, 2022Fracture; Fracture

    IT WAS REPORTED THAT: NAIL SNAPPED AT LAG SCREW JUNCTION.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13440250 ↗
  • InjuryJul 30, 2021Fracture; Insufficient Device Problem Information; Fracture; Insufficient Device Problem Information

    DURING THE EXPLANTATION OF THE METAL NAILS IT WAS FOUND THAT THE PATIENT'S CALLUS WAS COVERED WITH THE SCREW TAIL CAP. THE INTERMEDULLARY NAIL COULD NOT BE REMOVED AND WAS LEFT INSIDE THE PATIENT. THE SURGEON USED THE STANDARD LAG SCREW REMOVAL TOOL, BUT THE TAIL WAS BROKEN OFF AND HE HAD TO GIVE UP. THE SURGERY BEGAN AT 9AM AND ENDED AT 2-3PM.

    Patient problems: Extraskeletal ossification

    MAUDE report 12252729 ↗
  • InjuryJul 13, 2021Break; Fracture; Break; Fracture

    NO EVENT UPDATE. INVESTIGATION RESULTS ARE NOW AVAILABLE.

    Patient problems: Failure of Implant

    MAUDE report 12157909 ↗
  • InjuryDec 30, 2019Fracture; Fracture

    INVESTIGATION DONE.

    Patient problems: No Code Available

    MAUDE report 9529045 ↗
  • InjuryJun 13, 2019Insufficient Device Problem Information; Insufficient Device Problem Information

    NO EVENT UPDATE.

    Patient problems: Unspecified Infection

    MAUDE report 8694489 ↗
  • InjuryJun 11, 2019Fracture; Fracture

    NO CHANGE TO PREVIOUSLY REPORTED EVENT.

    Patient problems: No Code Available

    MAUDE report 8685947 ↗
  • InjuryDec 19, 2018Appropriate Device Problem Term/Code Not Available; Appropriate Device Problem Term/Code Not Available

    IT WAS REPORTED THAT TWO ZNN LAG SCREW GOT DAMAGED DURING THE SURGERY RESULTING IN 40 MINUTES DELAY OF SURGERY.

    Patient problems: No Information

    MAUDE report 8177932 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page