Virage® 07.01702.007Recalls & reported problems

Zimmer Biomet Inc · Class 2

Adverse event reports
13
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00889024327009), received between Dec 14, 2021 and Aug 7, 2024 — classified as 2 injuries, 11 malfunctions.

The most frequently reported specific device problem is “Fracture” (11 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 3 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture11
  • Crack1
  • Packaging Problem1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11
  • Fall2

Recent adverse event reports

The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.

  • MalfunctionAug 7, 2024Fracture; Fracture

    IT WAS REPORTED THAT A VIRAGE SET SCREW WOULD NOT SEAT IN IT'S MATING SCREW, RESULTING IN STRIPPED THREADS. ADDITIONALLY, THE SCREW TULIP HEAD SHEARED OFF. THE DEVICES WERE REMOVED AND REPLACED AND THERE WAS NO PATIENT IMPACT. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19934476 ↗
  • MalfunctionJul 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT AFTER FINAL TIGHTENING, THE VIRAGE CONSTRUCT WAS INSPECTED AND TWO VIRAGE SCREWS WERE FOUND TO HAVE FRACTURED. THEY WERE REPLACED WITH SCREWS OF A LARGER DIAMETER AND THE SURGERY WAS COMPLETED WITHOUT PATIENT HARM, ALTHOUGH THERE WAS A 15 MINUTE DELAY TO THE PROCEDURE. THIS IS REPORT ONE OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19866854 ↗
  • MalfunctionJul 30, 2024Fracture; Fracture

    IT WAS REPORTED THAT AFTER FINAL TIGHTENING, THE VIRAGE CONSTRUCT WAS INSPECTED AND TWO VIRAGE SCREWS WERE FOUND TO HAVE FRACTURED. THEY WERE REPLACED WITH SCREWS OF A LARGER DIAMETER AND THE SURGERY WAS COMPLETED WITHOUT PATIENT HARM, ALTHOUGH THERE WAS A 15 MINUTE DELAY TO THE PROCEDURE. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19866826 ↗
  • MalfunctionJul 18, 2024Fracture; Fracture

    IT WAS REPORTED THAT THE TULIP HEAD OF VIRAGE OCT POLYAXIAL SCREW BROKE OFF INTRA-OPERATIVELY. THE DOCTOR REMOVED IT AND REPLACED IT WITH ANOTHER 3.5X14 POLYAXIAL SCREW. THERE WAS NO PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19776937 ↗
  • MalfunctionNov 17, 2023Fracture; Fracture

    IT WAS REPORTED THAT A VIRAGE SCREW FRACTURE WAS DISCOVERED BY POSTOPERATIVE CT SCAN AND CONFIRMED BY X-RAY THAT MAY HAVE OCCURRED DURING THE FINAL CLOSURE OF THE SURGERY, ALTHOUGH THIS IS NOT CERTAIN. A REVISION SURGERY WAS NOT PERFORMED, AND THE PATIENT IS BEING TREATED WITH A HALO VEST. THE PATIENT IS ASYMPTOMATIC.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18162831 ↗
  • MalfunctionOct 19, 2023Fracture; Fracture

    IT WAS REPORTED THAT DURING THE FINAL TIGHTENING OF A VIRAGE SCREW AT LEFT C5, THE HOUSING BASE BROKE. THE SCREW WAS REMOVED AND REPLACED TO COMPLETE THE PROCEDURE. THERE WAS A 15-MINUTE DELAY AND NO IMPACT ON THE PATIENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17967008 ↗
  • MalfunctionJun 7, 2023Packaging Problem; Packaging Problem

    IT WAS REPORTED THAT A VIRAGE 3.5X20MM SCREW WAS PACKAGED AS A 3.5X14MM SCREW. THERE WAS NO REPORTED PATIENT INVOLVEMENT

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17082537 ↗
  • MalfunctionMay 10, 2023Fracture; Fracture

    IT WAS REPORTED THAT THE LOWER HOUSING ON A VIRAGE SCREW FRACTURED INTRA-OPERATIVELY. THE SCREW WAS REMOVED AND REPLACED TO COMPLETE THE PROCEDURE WITHOUT PATIENT IMPACTS.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16904598 ↗
  • MalfunctionOct 7, 2022Crack; Crack

    IT WAS REPORTED THAT TWO VIRAGE SCREWS BROKE AT THE WELD AND THREE OTHERS CRACKED AT THE WELD INTRA-OPERATIVELY. ON ONE SCREW, THE SURGEON PUSHED ON IT FROM ABOVE AFTER THE CLOSURE TOP WAS TIGHTENED AND THE LOWER HOUSING BROKE INTO TWO PIECES AT THE WELDS. THEN THE SURGEON NOTICED THE OTHER FOUR SCREWS THAT EITHER BROKE OR CRACKED. ALL OF THE SCREWS WERE REMOVED AND REPLACED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED PATIENT IMPACTS. THIS IS REPORT THREE OF FIVE FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15566469 ↗
  • InjuryApr 28, 2022Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT FELL AND A SUBSEQUENT X-RAY FOUND THAT TWO VIRAGE SCREWS WERE BROKEN. A REVISION SURGERY WAS PERFORMED TO REMOVE AND REPLACE THE SCREWS WITH 0.5MM LARGER DIAMETER VIRAGE SCREWS. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: Fall

    MAUDE report 14246571 ↗
  • InjuryApr 28, 2022Fracture; Fracture

    IT WAS REPORTED THAT A PATIENT FELL AND A SUBSEQUENT X-RAY FOUND THAT TWO VIRAGE SCREWS WERE BROKEN. A REVISION SURGERY WAS PERFORMED TO REMOVE AND REPLACE THE SCREWS WITH 0.5MM LARGER DIAMETER VIRAGE SCREWS. THIS IS REPORT ONE OF TWO FOR THIS EVENT.

    Patient problems: Fall

    MAUDE report 14246484 ↗
  • MalfunctionDec 14, 2021Fracture; Fracture

    IT WAS REPORTED THAT TWO SCREWS BROKE APART AFTER BEING FINAL TIGHTENED INTRA-OPERATIVELY. THEY SCREWS WERE REMOVED AND REPLACED TO COMPLETE THE PROCEDURE WITHOUT PATIENT IMPACTS. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13003953 ↗
  • MalfunctionDec 14, 2021Fracture; Fracture

    IT WAS REPORTED THAT TWO SCREWS BROKE APART AFTER BEING FINAL TIGHTENED INTRA-OPERATIVELY. THEY SCREWS WERE REMOVED AND REPLACED TO COMPLETE THE PROCEDURE WITHOUT PATIENT IMPACTS. THIS IS REPORT ONE OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13003952 ↗

Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices

All APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.