Virage® 07.01702.007Recalls & reported problems
Zimmer Biomet Inc · Class 2
- Adverse event reports
- 13
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 13 adverse event reports that list this exact device (DI 00889024327009), received between Dec 14, 2021 and Aug 7, 2024 — classified as 2 injuries, 11 malfunctions.
The most frequently reported specific device problem is “Fracture” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture11
- Crack1
- Packaging Problem1
Patient problems
- No Clinical Signs, Symptoms or Conditions11
- Fall2
Recent adverse event reports
The 13 most recent of 13 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 7, 2024Fracture; Fracture
IT WAS REPORTED THAT A VIRAGE SET SCREW WOULD NOT SEAT IN IT'S MATING SCREW, RESULTING IN STRIPPED THREADS. ADDITIONALLY, THE SCREW TULIP HEAD SHEARED OFF. THE DEVICES WERE REMOVED AND REPLACED AND THERE WAS NO PATIENT IMPACT. THIS IS REPORT TWO OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19934476 ↗MalfunctionJul 30, 2024Fracture; Fracture
IT WAS REPORTED THAT AFTER FINAL TIGHTENING, THE VIRAGE CONSTRUCT WAS INSPECTED AND TWO VIRAGE SCREWS WERE FOUND TO HAVE FRACTURED. THEY WERE REPLACED WITH SCREWS OF A LARGER DIAMETER AND THE SURGERY WAS COMPLETED WITHOUT PATIENT HARM, ALTHOUGH THERE WAS A 15 MINUTE DELAY TO THE PROCEDURE. THIS IS REPORT ONE OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19866854 ↗MalfunctionJul 30, 2024Fracture; Fracture
IT WAS REPORTED THAT AFTER FINAL TIGHTENING, THE VIRAGE CONSTRUCT WAS INSPECTED AND TWO VIRAGE SCREWS WERE FOUND TO HAVE FRACTURED. THEY WERE REPLACED WITH SCREWS OF A LARGER DIAMETER AND THE SURGERY WAS COMPLETED WITHOUT PATIENT HARM, ALTHOUGH THERE WAS A 15 MINUTE DELAY TO THE PROCEDURE. THIS IS REPORT TWO OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19866826 ↗MalfunctionJul 18, 2024Fracture; Fracture
IT WAS REPORTED THAT THE TULIP HEAD OF VIRAGE OCT POLYAXIAL SCREW BROKE OFF INTRA-OPERATIVELY. THE DOCTOR REMOVED IT AND REPLACED IT WITH ANOTHER 3.5X14 POLYAXIAL SCREW. THERE WAS NO PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19776937 ↗MalfunctionNov 17, 2023Fracture; Fracture
IT WAS REPORTED THAT A VIRAGE SCREW FRACTURE WAS DISCOVERED BY POSTOPERATIVE CT SCAN AND CONFIRMED BY X-RAY THAT MAY HAVE OCCURRED DURING THE FINAL CLOSURE OF THE SURGERY, ALTHOUGH THIS IS NOT CERTAIN. A REVISION SURGERY WAS NOT PERFORMED, AND THE PATIENT IS BEING TREATED WITH A HALO VEST. THE PATIENT IS ASYMPTOMATIC.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18162831 ↗MalfunctionOct 19, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING THE FINAL TIGHTENING OF A VIRAGE SCREW AT LEFT C5, THE HOUSING BASE BROKE. THE SCREW WAS REMOVED AND REPLACED TO COMPLETE THE PROCEDURE. THERE WAS A 15-MINUTE DELAY AND NO IMPACT ON THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17967008 ↗MalfunctionJun 7, 2023Packaging Problem; Packaging Problem
IT WAS REPORTED THAT A VIRAGE 3.5X20MM SCREW WAS PACKAGED AS A 3.5X14MM SCREW. THERE WAS NO REPORTED PATIENT INVOLVEMENT
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17082537 ↗MalfunctionMay 10, 2023Fracture; Fracture
IT WAS REPORTED THAT THE LOWER HOUSING ON A VIRAGE SCREW FRACTURED INTRA-OPERATIVELY. THE SCREW WAS REMOVED AND REPLACED TO COMPLETE THE PROCEDURE WITHOUT PATIENT IMPACTS.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16904598 ↗MalfunctionOct 7, 2022Crack; Crack
IT WAS REPORTED THAT TWO VIRAGE SCREWS BROKE AT THE WELD AND THREE OTHERS CRACKED AT THE WELD INTRA-OPERATIVELY. ON ONE SCREW, THE SURGEON PUSHED ON IT FROM ABOVE AFTER THE CLOSURE TOP WAS TIGHTENED AND THE LOWER HOUSING BROKE INTO TWO PIECES AT THE WELDS. THEN THE SURGEON NOTICED THE OTHER FOUR SCREWS THAT EITHER BROKE OR CRACKED. ALL OF THE SCREWS WERE REMOVED AND REPLACED TO COMPLETE THE PROCEDURE. THERE WERE NO REPORTED PATIENT IMPACTS. THIS IS REPORT THREE OF FIVE FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15566469 ↗InjuryApr 28, 2022Fracture; Fracture
IT WAS REPORTED THAT A PATIENT FELL AND A SUBSEQUENT X-RAY FOUND THAT TWO VIRAGE SCREWS WERE BROKEN. A REVISION SURGERY WAS PERFORMED TO REMOVE AND REPLACE THE SCREWS WITH 0.5MM LARGER DIAMETER VIRAGE SCREWS. THIS IS REPORT TWO OF TWO FOR THIS EVENT.
Patient problems: Fall
MAUDE report 14246571 ↗InjuryApr 28, 2022Fracture; Fracture
IT WAS REPORTED THAT A PATIENT FELL AND A SUBSEQUENT X-RAY FOUND THAT TWO VIRAGE SCREWS WERE BROKEN. A REVISION SURGERY WAS PERFORMED TO REMOVE AND REPLACE THE SCREWS WITH 0.5MM LARGER DIAMETER VIRAGE SCREWS. THIS IS REPORT ONE OF TWO FOR THIS EVENT.
Patient problems: Fall
MAUDE report 14246484 ↗MalfunctionDec 14, 2021Fracture; Fracture
IT WAS REPORTED THAT TWO SCREWS BROKE APART AFTER BEING FINAL TIGHTENED INTRA-OPERATIVELY. THEY SCREWS WERE REMOVED AND REPLACED TO COMPLETE THE PROCEDURE WITHOUT PATIENT IMPACTS. THIS IS REPORT TWO OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13003953 ↗MalfunctionDec 14, 2021Fracture; Fracture
IT WAS REPORTED THAT TWO SCREWS BROKE APART AFTER BEING FINAL TIGHTENED INTRA-OPERATIVELY. THEY SCREWS WERE REMOVED AND REPLACED TO COMPLETE THE PROCEDURE WITHOUT PATIENT IMPACTS. THIS IS REPORT ONE OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13003952 ↗
Other APPLIANCE, FIXATION, SPINAL INTERLAMINAL devices
- EXPEDIUM 179702000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL439 reports
- EXPEDIUM VERSE 199721001Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL227 reports
- XIA 3 48230000Stryker Corporation · APPLIANCE, FIXATION, SPINAL INTERLAMINAL193 reports
- VIPER 186715000Medos International Sàrl · APPLIANCE, FIXATION, SPINAL INTERLAMINAL1 death167 reports
- NA 04.614.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL80 reports
- EXPEDIUM VIPER 196789480Depuy Spine, LLC · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- NA 04.615.014Synthes GmbH · APPLIANCE, FIXATION, SPINAL INTERLAMINAL72 reports
- Arsenal Deformity Spinal Fixation System 47127Alphatec Spine, Inc. · APPLIANCE, FIXATION, SPINAL INTERLAMINAL66 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.