PathFinder NXT® 3572-1Recalls & reported problems
Biomet Spine LLC · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 5
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 00889024336513), received between Aug 21, 2024 and Dec 12, 2024 — classified as 5 malfunctions.
The most frequently reported specific device problem is “Calibration Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Calibration Problem3
- Mechanical Problem2
Patient problems
- No Clinical Signs, Symptoms or Conditions5
Recent adverse event reports
The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.
MalfunctionDec 12, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT THREE CLOSURE TOPS STRIPPED DURING FINAL TIGHTENING INTRA-OPERATIVELY. THE STRIPPED CLOSURE TOPS WERE REMOVED AND REPLACED. IT WAS SUSPECTED THAT THE TORQUE WRENCH WAS OVER-TORQUEING AND CONTRIBUTED TO THE DAMAGE. THERE WAS A 30 MINUTE DELAY BUT NO PATIENT IMPACT. THIS IS REPORT FOUR OF FOUR FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20918176 ↗MalfunctionNov 26, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT DURING THE FINAL TIGHTENING OF A SET SCREW, THE TORQUE WRENCH CLICKING SOUND WAS SLUGGISH AND THE SET SCREW WAS DAMAGED. THE SET SCREW WAS REPLACED, HOWEVER THE SECOND AND THIRD SET SCREWS WERE DAMAGED IN THE SAME WAY. THE BONE SCREW WAS ALSO DAMAGED, AND REMOVED AND REPLACED WITH A NEW SCREW, AND THE FINAL TIGHTENING OF THE SET SCREW WAS SUCCESSFUL. THIS CAUSED A 15 MINUTE DELAY TO THE PROCEDURE, HOWEVER THERE WAS NO PATIENT IMPACT. THIS IS REPORT THREE OF FIVE FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20787877 ↗MalfunctionNov 20, 2024Calibration Problem; Calibration Problem
IT WAS REPORTED THAT INTRA-OPERATIVELY, A PATHFINDER TORQUE RATCHET HANDLE WAS NOT ABLE TO TIGHTEN CLOSURE TOPS PROPERLY, STRIPPING THE THREADS ON FIVE CLOSURE TOPS AND STRIPPING THE INNER DRIVE HEXAGON ON THREE MORE. A PEDICLE SCREW WAS ALSO REPORTED TO HAVE DISASSEMBLED DURING THE SURGERY. THERE WAS NO PATIENT HARM, HOWEVER THERE WAS A 30 MINUTE PROCEDURAL DELAY AS A RESULT OF THIS. THIS IS REPORT TEN OF TEN FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20737165 ↗MalfunctionAug 22, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A SURGEON COMPLAINED THAT TWO T-HANDLE TORQUE RATCHETING HANDLES DID NOT APPLY ENOUGH TORQUE INTRA-OPERATIVELY. THE SURGERY WAS ABLE TO BE COMPLETED USING THE SAME TOOL WITH GREATER FORCE, AND THERE NO PATIENT IMPACT; HOWEVER, THERE WAS A 10 MINUTE DELAY ON SURGERY. THIS IS REPORT TWO OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20025721 ↗MalfunctionAug 21, 2024Mechanical Problem; Mechanical Problem
IT WAS REPORTED THAT A SURGEON COMPLAINED THAT TWO T-HANDLE TORQUE RATCHETING HANDLES DID NOT APPLY ENOUGH TORQUE INTRA-OPERATIVELY. THE SURGERY WAS ABLE TO BE COMPLETED USING THE SAME TOOL WITH GREATER FORCE, AND THERE NO PATIENT IMPACT; HOWEVER, THERE WAS A 10 MINUTE DELAY ON SURGERY. THIS IS REPORT ONE OF TWO FOR THIS EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20020634 ↗
Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices
- EXPEDIUM VERSE 199721000Medos International Sàrl · Orthosis, spinal pedicle fixation, for degenerative disc disease387 reports
- Sequoia® 3301-1Zimmer Spine, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease184 reports
- CD HORIZON® Spinal System 8590855Medtronic Sofamor Danek, Inc. · Orthosis, spinal pedicle fixation, for degenerative disc disease179 reports
- Vitality® 07.02011.001Biomet Spine LLC · Orthosis, spinal pedicle fixation, for degenerative disc disease3 recalls0 reports
- EXPEDIUM VERSE 199721001SMedos International Sàrl · Thoracolumbosacral pedicle screw system124 reports
- LineSider LS-N6SA7550Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls1 report
- LineSider LS-N6SA6545Integrity Implants Inc. · Thoracolumbosacral Pedicle Screw System2 recalls0 reports
- EXPEDIUM VERSE 199721000SMedos International Sàrl · Thoracolumbosacral pedicle screw system97 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page