Dermatome 00-8800-000-10Recalls & reported problems

Zimmer Orthopaedic Surgical Products, Inc. · Dermatome Blades · Class 1

All models of Dermatome (5)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Quick summary

This Zimmer dermatome blade, used to harvest skin grafts, has 109 adverse event reports, with no deaths and 59 injuries. Most reports describe the blade failing to cut properly, and some mention cuts to patients' skin. Reports fell to 1 in the last 12 months from 7 in the 12 months before. 1 recall is open, started in September 2023, over skin grafts that come out thin and uneven and may need additional harvests.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
109
Deaths reported
0
Injuries reported
59
Recalls naming it
1

Summary

The FDA's MAUDE database contains 109 adverse event reports that list this exact device (DI 00889024375895), received between Sep 20, 2022 and Apr 7, 2026 — classified as 59 injuries, 50 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (95 reports).

1 FDA recall record name this device's identifier, 1 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 7 reportsDec 2023: 1 reportsJan 2024: 5 reportsFeb 2024: 2 reportsMar 2024: 1 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 2 reportsAug 2024: 0 reportsSep 2024: 16 reportsOct 2024: 0 reportsNov 2024: 1 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 1 reportsJun 2025: 2 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports20242025202616
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Sep 5, 2023Class IIOpen, Classified

    Skin grafts thin and non-uniform when using the affected blades. The issue would be identified at the time of use and may result in the need for additional harvests to adequately cover the area

    Z-0069-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported problems

Device problems

  • Failure to Cut95
  • Patient Device Interaction Problem58
  • Material Deformation8
  • Material Twisted/Bent7
  • Dull, Blunt6
  • Positioning Problem6
  • Delivered as Unsterile Product5
  • Vibration4

Patient problems

  • No Clinical Signs, Symptoms or Conditions55
  • Laceration(s)50
  • Skin Tears17
  • Abrasion9
  • Unspecified Tissue Injury4
  • Insufficient Information3
  • Physical Asymmetry2

Recent adverse event reports

The 25 most recent of 109 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 7, 2026Fracture; Material Twisted/Bent

    IT WAS REPORTED THAT BEFORE SURGERY THE DERMATOME BLADE WAS NOTED TO BE BENT AND PLASTIC PART WAS FRACTURED. NO PATIENT INVOLVEMENT. DILIGENCE IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24800679 ↗
  • MalfunctionOct 15, 2025Delivered as Unsterile Product; Delivered as Unsterile Product

    NO ADDITIONAL INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23295389 ↗
  • MalfunctionJul 17, 2025Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT DURING SURGERY, CUTTING THE GRAFT DOES NOT WORK PROPERLY. THERE WAS NO PATIENT HARM OR SURGICAL DELAY REPORTED. DURING DEVICE EVALUATION IT WAS DISCOVERED THAT THE BLADE MAY HAVE CONTRIBUTED TO THE REPORTED EVENT. DUE DILIGENCE IS COMPLETE, NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22531451 ↗
  • MalfunctionJun 27, 2025Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT DURING ROUTINE HARVESTING OF A SKIN GRAFT, THE DERMATOME FAILED TO EXCISE A GRAFT EXCEPT ONLY ALONG ONE SMALL EDGE OF THE DERMATOME BLADE. A REPEAT ATTEMPT PROVED THE SAME RESULTS. THE ENTIRE DERMATOME SET (PRESSURIZED N2 TUBING, DERMATOME HEAD, BLADE, & GUARD) WERE CHANGED TO A NEW SET AND BROUGHT ONTO THE FIELD. A REPEAT ATTEMPT REVEALED THE SAME RESULTS. CONFERRING WITH A PARTNER ATTENDING IN TROUBLESHOOTING DID NOT IMPROVE THE PROBLEM. FINALLY, A NEW DERMATOME BLADE FROM A DIFFERENT LOT NUMBER (#66646709) WAS USED IN THE DEVICE AND THE DEVICE FUNCTIONED APPROPRIATELY. IT WAS ASSUMED THAT BLADES FROM LOT#66609275 ARE DEFECTIVE THOUGH OUTWARDLY NOT APPEARING ABNORMAL. ALL DERMATOME BLADES FROM THAT LOT WERE TAKEN OUT OF SERVICE. THERE WAS A SURGICAL DELAY OF AN UNKNOWN AMOUNT. THERE WAS NO PATIENT HARM. DUE DILIGENCE IS IN PROGRESS, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22346587 ↗
  • MalfunctionJun 27, 2025Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT DURING ROUTINE HARVESTING OF A SKIN GRAFT, THE DERMATOME FAILED TO EXCISE A GRAFT EXCEPT ONLY ALONG ONE SMALL EDGE OF THE DERMATOME BLADE. A REPEAT ATTEMPT PROVED THE SAME RESULTS. THE ENTIRE DERMATOME SET (PRESSURIZED N2 TUBING, DERMATOME HEAD, BLADE, & GUARD) WERE CHANGED TO A NEW SET AND BROUGHT ONTO THE FIELD. A REPEAT ATTEMPT REVEALED THE SAME RESULTS. CONFERRING WITH A PARTNER ATTENDING IN TROUBLESHOOTING DID NOT IMPROVE THE PROBLEM. FINALLY, A NEW DERMATOME BLADE FROM A DIFFERENT LOT NUMBER (#66646709) WAS USED IN THE DEVICE AND THE DEVICE FUNCTIONED APPROPRIATELY. IT WAS ASSUMED THAT BLADES FROM LOT#66609275 ARE DEFECTIVE THOUGH OUTWARDLY NOT APPEARING ABNORMAL. ALL DERMATOME BLADES FROM THAT LOT WERE TAKEN OUT OF SERVICE. THERE WAS A SURGICAL DELAY OF AN UNKNOWN AMOUNT. THERE WAS NO PATIENT HARM. DUE DILIGENCE IS IN PROGRESS, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22346551 ↗
  • MalfunctionMay 1, 2025Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT DURING SURGERY THE DERMATOME PRODUCED A SKIN GRAFT WITH UNEVEN THICKNESS. IT WAS REPORTED THAT THERE WAS NO HARM TO THE PATIENT OR ANY SURGICAL DELAY. NO ADDITIONAL INFORMATION AVAILABLE IS INDICATED. DILIGENCE IS COMPLETE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21945580 ↗
  • InjuryApr 28, 2025Failure to Cut; Failure to Cut

    IT WAS REPORTED DURING THE PROCEDURE THAT THERE WAS AN UNEVEN GRAFT THICKNESS. THERE IS NO HARM OR DELAY REPORTED. DUE DILIGENCE IS IN PROGRESS AND NO ADDITIONAL INFORMATION IS AVAILABLE TILL DATE.

    Patient problems: Skin Tears; Unspecified Tissue Injury

    MAUDE report 21918372 ↗
  • InjuryNov 11, 2024Failure to Cut; Patient Device Interaction Problem; Failure to Cut; Patient Device Interaction Problem

    THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s)

    MAUDE report 20658847 ↗
  • InjurySep 13, 2024Failure to Cut; Patient Device Interaction Problem; Failure to Cut; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS INADEQUATE. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20211232 ↗
  • InjurySep 12, 2024Failure to Cut; Patient Device Interaction Problem; Failure to Cut; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS INADEQUATE. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20208662 ↗
  • InjurySep 12, 2024Failure to Cut; Material Deformation; Patient Device Interaction Problem; Failure to Cut; Material Deformation; Patient Device Interaction Problem

    THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20207559 ↗
  • InjurySep 12, 2024Failure to Cut; Patient Device Interaction Problem; Failure to Cut; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS INADEQUATE. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20207553 ↗
  • InjurySep 12, 2024Failure to Cut; Patient Device Interaction Problem; Failure to Cut; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS INADEQUATE. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20207390 ↗
  • InjurySep 12, 2024Failure to Cut; Patient Device Interaction Problem; Failure to Cut; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS INADEQUATE. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20207193 ↗
  • InjurySep 12, 2024Failure to Cut; Material Deformation; Patient Device Interaction Problem; Failure to Cut; Material Deformation; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS BOWING. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20206967 ↗
  • InjurySep 12, 2024Failure to Cut; Patient Device Interaction Problem; Failure to Cut; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS INADEQUATE. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20206805 ↗
  • InjurySep 12, 2024Failure to Cut; Patient Device Interaction Problem; Failure to Cut; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS INADEQUATE. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20206100 ↗
  • InjurySep 12, 2024Failure to Cut; Patient Device Interaction Problem; Failure to Cut; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS INADEQUATE. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20205855 ↗
  • InjurySep 12, 2024Failure to Cut; Material Deformation; Patient Device Interaction Problem; Failure to Cut; Material Deformation; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS BOWING. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20205499 ↗
  • InjurySep 12, 2024Failure to Cut; Material Deformation; Patient Device Interaction Problem; Failure to Cut; Material Deformation; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS BOWING. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20202863 ↗
  • InjurySep 12, 2024Failure to Cut; Material Deformation; Patient Device Interaction Problem; Failure to Cut; Material Deformation; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE BLADE WAS BOWING. THE TAKEN GRAFT WAS DAMAGED AS IT WAS NOT ABLE TO TAKE IN ONE STRIP. SUTURES WERE NOT REQUIRED AT THE HARVEST SITE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20200553 ↗
  • InjurySep 12, 2024Failure to Cut; Material Deformation; Patient Device Interaction Problem; Failure to Cut; Material Deformation; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE ZIMMER BLADE WAS BOWING CREATING A GAP IN THE MIDDLE. WHEN CORRECTED, THE GAP WAS ON EITHER SIDE OF THE BLADE. WHEN SKIN WAS REMOVED IT IS DONE SO UNEVENLY AND IN STRIPS RATHER THAN IN ONE PIECE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. THE HARVEST SITE DID NOT REQUIRE SUTURES. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20200266 ↗
  • InjurySep 12, 2024Failure to Cut; Material Deformation; Patient Device Interaction Problem; Failure to Cut; Material Deformation; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE ZIMMER BLADE WAS BOWING CREATING A GAP IN THE MIDDLE. WHEN CORRECTED, THE GAP WAS ON EITHER SIDE OF THE BLADE. WHEN SKIN WAS REMOVED IT IS DONE SO UNEVENLY AND IN STRIPS RATHER THAN IN ONE PIECE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. THE HARVEST SITE DID NOT REQUIRE SUTURES. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20200217 ↗
  • InjurySep 12, 2024Failure to Cut; Material Deformation; Patient Device Interaction Problem; Failure to Cut; Material Deformation; Patient Device Interaction Problem

    IT WAS REPORTED THAT DURING SURGERY, THE ZIMMER BLADE WAS BOWING CREATING A GAP IN THE MIDDLE. WHEN CORRECTED, THE GAP WAS ON EITHER SIDE OF THE BLADE. WHEN SKIN WAS REMOVED IT IS DONE SO UNEVENLY AND IN STRIPS RATHER THAN IN ONE PIECE. THE PATIENT REQUIRED ADDITIONAL TIME UNDER ANESTHESIA AND ADDITIONAL SKIN GRAFTING WAS REQUIRED TO COMPLETE THE SURGERY. THE HARVEST SITE DID NOT REQUIRE SUTURES. ANOTHER PRODUCT WAS NOT USED TO COMPLETE THE SURGERY. DUE DILIGENCE IS COMPLETE, THERE IS NO ADDITIONAL INFORMATION AVAILABLE.

    Patient problems: Laceration(s); Skin Tears

    MAUDE report 20200175 ↗
  • MalfunctionJul 3, 2024Failure to Cut; Failure to Cut

    IT WAS REPORTED THAT DURING SURGERY, THE UNIT DID NOT CUT PROPERLY. IT ONLY CUTS ON OUTSIDE OF UNIT, LEAVES SPACES THROUGHOUT THE MIDDLE. THERE WAS NO PATIENT HARM OR INJURY. THERE WAS NO SURGICAL DELAY. THERE WAS NO SURGICAL/MEDICAL INTERVENTION. THE TAKEN GRAFT WAS DAMAGED HOWEVER AN ADDITIONAL GRAFT WAS NOT REQUIRED TO COMPLETE THE SURGERY. ANOTHER DEVICE WAS NOT USED TO FINISH SURGERY. DURING DEVICE EVALUATION, IT WAS DISCOVERED THAT THE CUTTER MAY HAVE CONTRIBUTED TO THE REPORTED EVENT. DUE DILIGENCE IS COMPLETE, NO FURTHER INFORMATION IS AVAILABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19667026 ↗

Other Dermatome devices

All Dermatome devices →