Skip to content

DermatomeRecalls & reported problems across all models

Zimmer Orthopaedic Surgical Products, Inc. · 5 models in the FDA device registry

Data updated Oct 16, 2026 from official FDA records · Editorial policy

2 open recalls name at least one Dermatome model. Check whether yours is affected in the recall details, and contact the manufacturer or your healthcare provider.

Adverse event reports
1,597
Deaths reported
0
Injuries reported
197
Recalls
2

We found 5 models of Dermatome by Zimmer Orthopaedic Surgical Products, Inc. in the FDA's device registry (GUDID). Together they are named in 1,597 adverse event reports in the FDA's MAUDE database, received between Jul 2, 2019 and Sep 30, 2026 — classified as 197 injuries, 1,400 malfunctions.

The most frequently reported specific device problem is “Failure to Cut” (559 reports).

2 FDA recall records name at least one of these models by its device identifier, 2 still open.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event. Widely used products collect more reports, so totals can't be used to compare products or estimate failure rates. Only reports that name a model's exact device identifier are counted. What reports can and can't tell you · How this data works

Reports received per month, all models

Nov 2023: 41 reportsDec 2023: 20 reportsJan 2024: 38 reportsFeb 2024: 17 reportsMar 2024: 17 reportsApr 2024: 26 reportsMay 2024: 27 reportsJun 2024: 27 reportsJul 2024: 22 reportsAug 2024: 32 reportsSep 2024: 39 reportsOct 2024: 48 reportsNov 2024: 43 reportsDec 2024: 29 reportsJan 2025: 33 reportsFeb 2025: 32 reportsMar 2025: 36 reportsApr 2025: 38 reportsMay 2025: 24 reportsJun 2025: 26 reportsJul 2025: 56 reportsAug 2025: 29 reportsSep 2025: 29 reportsOct 2025: 27 reportsNov 2025: 26 reportsDec 2025: 19 reportsJan 2026: 17 reportsFeb 2026: 14 reportsMar 2026: 21 reportsApr 2026: 23 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 10 reportsSep 2026: 7 reportsOct 2026: 0 reports20242025202656
Last 36 months. The most recent month may be incomplete.

Dermatome recalls

FDA recall records that name at least one of these models by its device identifier, newest first. Each recall usually covers specific models or lots, so check the details.

  1. Nov 24, 2025Open, Classified

    The devices may have a misaligned thickness control bar.

    Z-0924-2026 · Root cause: Process control · Recall details · FDA record ↗

  2. Sep 5, 2023Open, Classified

    Skin grafts thin and non-uniform when using the affected blades. The issue would be identified at the time of use and may result in the need for additional harvests to adequately cover the area

    Z-0069-2024 · Root cause: Process control · Recall details · FDA record ↗

Most reported Dermatome problems

Device problems

  • Failure to Cut559
  • Unstable234
  • Patient Device Interaction Problem182
  • Appropriate Device Problem Term/Code Not Available170
  • Power Problem128
  • Calibration Problem80
  • Intermittent Loss of Power79
  • Material Deformation69

Patient problems

  • No Clinical Signs, Symptoms or Conditions1,253
  • Laceration(s)181
  • Insufficient Information149
  • Skin Tears38
  • Abrasion33
  • Unspecified Tissue Injury30
  • Graft failure18
  • No Consequences Or Impact To Patient7

Dermatome models

5 models, most reported first. Each model has its own page with its reports, recalls and device details.

Recent Dermatome adverse event reports

The 15 most recent reports across all models, in the reporter's own words as published by the FDA.

Other Zimmer Orthopaedic Surgical Products, Inc. products

All Zimmer Orthopaedic Surgical Products, Inc. devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page