Skip to content

InCompass® 2155-1Recalls & reported problems

Zimmer Spine, Inc. · Class 2

Adverse event reports
11
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00889024398993), received between Aug 29, 2022 and Jan 22, 2026 — classified as 11 malfunctions.

The most frequently reported specific device problem is “Fracture” (9 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 3 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 2 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 1 reportsJan 2026: 2 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture9
  • Material Deformation1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions11

Recent adverse event reports

The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJan 22, 2026Material Twisted/Bent

    IT WAS REPORTED THAT A PATHFINDER REMOVAL DRIVER TIP BENT DURING A REVISION CASE. THERE WAS NO PROCEDURAL IMPACT NOR PATIENT HARM, AND THE REVISION PROCEDURE WAS NOT RELATED TO HIGHRIDGE PRODUCT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24148397 ↗
  • MalfunctionJan 22, 2026Fracture

    IT WAS REPORTED THAT A PATHFINDER REMOVAL DRIVER TIP FRACTURED DURING A REVISION CASE. THERE WAS NO PROCEDURAL IMPACT NOR PATIENT HARM, AND THE REVISION PROCEDURE WAS NOT RELATED TO HIGHRIDGE PRODUCT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24148349 ↗
  • MalfunctionDec 22, 2025Fracture; Fracture

    IT WAS REPORTED THAT THE TIPS OF THREE SCREW DRIVERS (ONE UNIVERSAL AND TWO UNIVERSAL WITH STANDARD CONNECTIONS) BROKE INTRA-OPERATIVELY WHILE ATTEMPTING TO REMOVE A SCREW THAT HAD BEEN IMPLANTED FOR 16 YEARS. THE SCREW WAS ABLE TO BE REMOVED USING AN ALTERNATIVE METHOD SINCE THE DRIVER ENGAGEMENT FEATURE HAD STRIPPED. THERE WERE NO REPORTED IMPACTS TO THE PATIENT. THIS IS REPORT ONE OF THREE FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23871615 ↗
  • MalfunctionSep 11, 2025Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE TIP OF PATHFINDER NXT UNIVERSAL DRIVER TIP DEFORMED DURING REMOVAL OF PATHFINDER HARDWARE DUE TO ADJACENT SEGMENT BREAKDOWN AND EXTENSION OF CONSTRUCT. THERE WAS NO REPORTED PATIENT IMPACT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23032063 ↗
  • MalfunctionMay 21, 2025Fracture; Fracture

    IT WAS REPORTED THAT A DISTRIBUTOR DISCOVERED TWO PATHFINDER UNIVERSAL DRIVERS WITH FRACTURED TIPS ON A SHELF IN THE SPD DEPARTMENT AFTER THEY WERE RETURNED FROM A HOSPITAL. NO INFORMATION REGARDING PATIENT IMPACT WAS PROVIDED WITH THE RETURN. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22059628 ↗
  • MalfunctionMay 21, 2025Fracture; Fracture

    IT WAS REPORTED THAT A DISTRIBUTOR DISCOVERED TWO PATHFINDER UNIVERSAL DRIVERS WITH FRACTURED TIPS ON A SHELF IN THE SPD DEPARTMENT AFTER THEY WERE RETURNED FROM A HOSPITAL. NO INFORMATION REGARDING PATIENT IMPACT WAS PROVIDED WITH THE RETURN. THIS IS REPORT ONE OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22059611 ↗
  • MalfunctionJun 12, 2024Fracture; Fracture

    IT WAS REPORTED THAT FIVE PATHFINDER AND SEQUOIA DRIVERS AND REVISION TOOLS WERE DISCOVERED WITH FRACTURED TIPS AT A DISTRIBUTOR. THERE WAS NO REPORTED PATIENT IMPACT. THIS IS REPORT FIVE OF FIVE FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19522322 ↗
  • MalfunctionJun 12, 2024Fracture; Fracture

    IT WAS REPORTED THAT FIVE PATHFINDER AND SEQUOIA DRIVERS AND REVISION TOOLS WERE DISCOVERED WITH FRACTURED TIPS AT A DISTRIBUTOR. THERE WAS NO REPORTED PATIENT IMPACT. THIS IS REPORT FOUR OF FIVE FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19522312 ↗
  • MalfunctionJun 12, 2024Fracture; Fracture

    IT WAS REPORTED THAT FIVE PATHFINDER AND SEQUOIA DRIVERS AND REVISION TOOLS WERE DISCOVERED WITH FRACTURED TIPS AT A DISTRIBUTOR. THERE WAS NO REPORTED PATIENT IMPACT. THIS IS REPORT THREE OF FIVE FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19522303 ↗
  • MalfunctionAug 29, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A REVISION SURGERY TO EXTEND THE CONSTRUCT, TWO INCOMPASS DRIVERS FRACTURED WHEN TRYING TO REMOVE THE PREVIOUSLY-INSTALLED CLOSURE TOPS. THERE WERE NO PATIENT IMPACTS. THIS IS REPORT ONE OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15310926 ↗
  • MalfunctionAug 29, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING A REVISION SURGERY TO EXTEND THE CONSTRUCT, TWO INCOMPASS DRIVERS FRACTURED WHEN TRYING TO REMOVE THE PREVIOUSLY-INSTALLED CLOSURE TOPS. THERE WERE NO PATIENT IMPACTS. THIS IS REPORT TWO OF TWO FOR THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15310925 ↗

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.