Affixus® Natural Nail® 47-2486-038-40Recalls & reported problems

Zimmer GmbH · Class 2 · Implant

All models of Affixus Natural Nail (55)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Quick summary

This Zimmer Affixus Natural Nail intramedullary nail is used to fix fractures inside long bones. The 53 reports mostly describe screws backing out or shifting out of position after surgery, sometimes with pain and, in some cases, a second operation to reinsert the screws. There are 41 injuries and no deaths. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
53
Deaths reported
0
Injuries reported
41
Recalls naming it
0

Summary

The FDA's MAUDE database contains 53 adverse event reports that list this exact device (DI 00889024505438), received between Jun 26, 2020 and May 20, 2026 — classified as 41 injuries, 12 malfunctions.

The most frequently reported specific device problem is “Migration” (40 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 3 reportsJan 2024: 2 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 3 reportsSep 2024: 0 reportsOct 2024: 3 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 1 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 1 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 1 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration40
  • Migration or Expulsion of Device18
  • Patient Device Interaction Problem7
  • Loosening of Implant Not Related to Bone-Ingrowth3
  • Adverse Event Without Identified Device or Use Problem1
  • Device Slipped1
  • Device-Device Incompatibility1

Patient problems

  • Failure of Implant29
  • Pain14
  • Insufficient Information11
  • No Clinical Signs, Symptoms or Conditions11
  • Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available4
  • Unspecified Tissue Injury3
  • No Code Available1

Recent adverse event reports

The 25 most recent of 53 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 20, 2026Migration

    IT WAS REPORTED THAT TWO PROXIMAL SCREWS OF AN INTRAMEDULLARY NAIL BACKED OUT APPROXIMATELY TWO MONTHS POSTOPERATIVELY. THE PATIENT WAS MONITORED AND NO REVISION WAS PLANNED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25237096 ↗
  • MalfunctionFeb 5, 2026Migration

    IT WAS REPORTED THAT ONE PROXIMAL SCREW BACKED OUT APPROXIMATELY 2 MONTHS POSTOPERATIVELY FOLLOWING AN INTRAMEDULLARY NAILING PROCEDURE. THE PATIENT WAS MONITORED, AND NO REVISION WAS PLANNED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24274676 ↗
  • MalfunctionOct 31, 2025Migration; Migration

    DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 23437333 ↗
  • InjuryJan 8, 2025Migration; Migration

    IT WAS REPORTED THAT AN ANN NAIL SURGERY WAS PERFORMED. AFTER 1 MONTH FROM THE INITIAL SURGERY, THE SURGEON FOUND THAT 3 PROXIMAL SCREWS HAD MIGRATED OUT OF THE PROPER POSITION. THE PATIENT COMPLAINED OF PAIN AND REVISION SURGERY WAS PERFORMED. IN THE REVISION SURGERY, THE SCREWS THAT HAD COME LOOSE WERE REINSERTED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Pain

    MAUDE report 21106005 ↗
  • InjuryOct 24, 2024Migration; Migration

    IT WAS REPORTED THAT AN INITIAL OPERATION WAS PERFORMED WITH ANN NAIL SYSTEM AND AFTER 3 WEEKS, SURGEON FOUND TWO OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. THE SURGEON KEEPS AN EYE ON THE PATIENT CONDITION AS WELL AS NO REVISION WILL BE PLANNED SO FAR. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 20522091 ↗
  • InjuryOct 24, 2024Migration; Migration

    IT WAS REPORTED THAT AN INITIAL OPERATION WAS PERFORMED WITH ANN NAIL SYSTEM AND AFTER 3 WEEKS, SURGEON FOUND TWO OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. THE SURGEON KEEPS AN EYE ON THE PATIENT CONDITION AS WELL AS NO REVISION WILL BE PLANNED SO FAR. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 20522078 ↗
  • MalfunctionOct 8, 2024Migration; Migration

    DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 20398421 ↗
  • InjuryAug 27, 2024Migration; Migration

    DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Pain

    MAUDE report 20077723 ↗
  • InjuryAug 22, 2024Migration; Migration

    DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Failure of Implant; Pain

    MAUDE report 20038826 ↗
  • InjuryAug 15, 2024Migration; Migration

    IT WAS REPORTED THAT AFTER 2 MONTHS FROM THE INITIAL SURGERY, THE SURGEON FOUND THAT THE #1 AND #3 PROXIMAL SCREWS HAD MIGRATED OUT OF THE PROPER POSITION. THE PATIENT COMPLAINED OF PAIN AND REVISION SURGERY WAS PERFORMED TO REMOVE THE MIGRATED SCREWS. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant; Pain

    MAUDE report 19994475 ↗
  • InjuryJun 20, 2024Migration; Migration

    IT WAS REPORTED PATIENT UNDERWENT REVISION APPROXIMATELY 4 WEEKS FROM THE INITIAL SURGERY, SURGEON FOUND ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. THE PATIENT COMPLAINED OF PAIN AND A REVISION WAS PERFORMED TO REMOVE THE MIGRATED SCREW. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant; Pain

    MAUDE report 19579275 ↗
  • InjuryJan 30, 2024Migration; Migration

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant; Pain

    MAUDE report 18606745 ↗
  • InjuryJan 5, 2024Migration; Migration

    IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN APPROXIMATELY 1 MONTH AGO. AFTER 4 DAYS FROM THE INITIAL, SURGEON FOUND ALL PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. NO REOPERATION DATE HAS BEEN SET, BUT THE ANN WILL BE REMOVED, AND THE RSA WILL BE PERFORMED. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 18455903 ↗
  • MalfunctionDec 21, 2023Migration; Migration

    IT WAS REPORTED THAT AFTER 2 WEEKS FROM THE INITIAL PROCEDURE, SURGEON FOUND THAT THE PROXIMAL SCREW WAS BACKED OUT FROM THE PROPER POSITION. THE SURGEON KEEPS AN EYE ON THE PATIENT CONDITION AS WELL AS NO REVISION WILL BE PLANNED SO FAR. NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18377002 ↗
  • MalfunctionDec 19, 2023Migration; Migration

    DILIGENCE IS COMPLETE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.

    Patient problems: Failure of Implant

    MAUDE report 18357514 ↗
  • InjuryDec 11, 2023Migration; Migration

    IT WAS REPORTED THAT THE PATIENT HAD A REVISION SURGERY DUE TO PROXIMAL SCREWS THAT WERE BACKED OUT FROM THE PROPER POSITION, APPROXIMATELY ONE (1) MONTH AFTER IMPLANTATION. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: Failure of Implant; Pain

    MAUDE report 18307484 ↗
  • InjuryNov 13, 2023Migration; Migration

    IT WAS REPORTED THERE WAS A REVISION SURGERY 1 MONTH AFTER IMPLANTATION DUE TO ALL PROXIMAL SCREWS HAVING BACKED OUT FROM THE PROPER POSITION. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: Failure of Implant; Pain

    MAUDE report 18118696 ↗
  • InjuryJun 22, 2023Migration; Migration

    IT WAS REPORTED THAT FOUR (4) WEEKS AFTER IMPLANTATION OF SHOULDER NAIL, ONE OF THE PROXIMAL SCREWS HAS BACKED OUT FROM PROPER POSITION. IT IS NOT KNOWN WHICH ONE EXACTLY BACKED OUT. NO REVISION IS PLANNED SO FAR. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17179746 ↗
  • InjuryJun 22, 2023Migration; Migration

    IT WAS REPORTED THAT FOUR (4) WEEKS AFTER IMPLANTATION OF SHOULDER NAIL, ONE OF THE PROXIMAL SCREWS HAS BACKED OUT FROM PROPER POSITION. IT IS NOT KNOWN WHICH ONE EXACTLY BACKED OUT. NO REVISION IS PLANNED SO FAR. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.

    Patient problems: Failure of Implant

    MAUDE report 17179743 ↗
  • InjuryJun 20, 2023Migration; Migration

    IT WAS REPORTED THAT THE TWO PROXIMAL SCREWS BACKED OUT FROM THE PROPER POSITION ABOUT FOUR (4) MONTHS AFTER THE INITIAL SURGERY. THE SURGEON WILL KEEP AN EYE ON THE PATIENT'S CONDITION. NO REVISION WILL BE PLANNED SO FAR. NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant

    MAUDE report 17160297 ↗
  • InjuryJun 13, 2023Migration; Migration

    IT WAS REPORTED THAT AFTER TWO WEEKS FROM THE INITIAL SURGERY, TWO PROXIMAL SCREWS BACKED OUT FROM THE PROPER POSITION. NO CONSEQUENCES OR IMPACT TO THE PATIENT. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 17115989 ↗
  • InjuryApr 26, 2023Migration; Migration

    IT WAS REPORTED THAT ONE MONTH AFTER INITIAL RIGHT HUMERAL NAIL IMPLANTATION ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM ITS PROPER POSITION. NO INTERVENTION IS PLANNED AT THIS TIME. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: Failure of Implant

    MAUDE report 16813090 ↗
  • InjuryApr 25, 2023Migration; Migration

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant

    MAUDE report 16804865 ↗
  • InjuryApr 25, 2023Migration; Migration

    IT WAS REPORTED THAT FOUR MONTHS AFTER INITIAL LEFT HUMERAL NAIL IMPLANTATION ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM ITS PROPER POSITION. NO INTERVENTION IS PLANNED AT THIS TIME. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.

    Patient problems: Failure of Implant

    MAUDE report 16804710 ↗
  • InjuryApr 6, 2023Migration; Migration

    NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.

    Patient problems: Failure of Implant

    MAUDE report 16687531 ↗