Affixus® Natural Nail® 47-2486-050-40Recalls & reported problems
Zimmer GmbH · Class 2 · Implant
All models of Affixus Natural Nail (55)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 14
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 14 adverse event reports that list this exact device (DI 00889024505490), received between Oct 6, 2020 and Oct 31, 2025 — classified as 12 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Migration” (10 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration10
- Migration or Expulsion of Device6
Patient problems
- Failure of Implant5
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions3
- No Code Available2
- Fall1
- Loss of Range of Motion1
- Pain1
Recent adverse event reports
The 14 most recent of 14 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 31, 2025Migration; Migration
DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 23437342 ↗InjurySep 17, 2025Migration; Migration
IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN NAIL. ONE WEEK AFTER THE INITIAL SURGERY, SURGEON FOUND ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. TWO WEEKS AFTER THE INITIAL SURGERY, ALL OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION AND SUBSEQUENTLY UNDERWENT A REVISION SURGERY, DURING WHICH, THE SURGEON CONFIRMED THAT THE CORELOCK HAD BECOME LOOSE. THEN THE SCREW DEPTH WAS READJUSTED, THE CORELOCK WAS RETIGHTENED, AND THE REVISION WAS COMPLETED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 23074155 ↗InjurySep 17, 2025Migration; Migration
IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN NAIL. ONE WEEK AFTER THE INITIAL SURGERY, SURGEON FOUND ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. TWO WEEKS AFTER THE INITIAL SURGERY, ALL OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION AND SUBSEQUENTLY UNDERWENT A REVISION SURGERY, DURING WHICH, THE SURGEON CONFIRMED THAT THE CORELOCK HAD BECOME LOOSE. THEN THE SCREW DEPTH WAS READJUSTED, THE CORELOCK WAS RETIGHTENED, AND THE REVISION WAS COMPLETED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 23074027 ↗MalfunctionJul 24, 2025Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL SURGERY WITH ANN NAIL. SUBSEQUENTLY, 9 DAYS POST IMPLANTATION, SURGEON FOUND THREE OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. THE SURGEON WILL KEEP AN EYE ON THE PATIENT CONDITION AS WELL, AS NO REVISION WILL BE PLANNED SO FAR. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22587079 ↗InjuryApr 28, 2025Migration; Migration
IT WAS REPORTED THAT A PATIENT HAD AN INITIAL PROCEDURE WITH THE AFFIXUS HUMERAL NAIL, AFTER WHICH, PRESENTS LIMITED MOVEMENT DUE TO LOOSENING OF TWO SCREWS. PATIENT IS PENDING FOR SCREW REMOVAL ON UNKNOWN DATE. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Failure of Implant; Loss of Range of Motion
MAUDE report 21920452 ↗InjuryJul 5, 2023Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN OPERATION WITH A PROXIMAL HUMERUS NAIL APPROXIMATELY THREE MONTHS AGO. SIX WEEKS AFTER THE INITIAL PROCEDURE, THE SURGEON FOUND #1 AND #3 PROXIMAL SCREWS WERE BACKED OUT FROM THEIR CORRECT POSITION. THE PATIENTS CONDITION WILL BE MONITORED AND NO REVISION IS CURRENTLY PLANNED. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE, NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17255672 ↗InjuryOct 25, 2022Migration or Expulsion of Device; Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION SURGERY DUE TO MIGRATION OF THE SCREW.
Patient problems: Failure of Implant
MAUDE report 15665247 ↗InjuryOct 14, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION DUE TO SCREW MIGRATION 3-MONTHS POST-IMPLANTATION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 15602997 ↗InjuryOct 14, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT UNDERWENT REVISION DUE TO SCREW MIGRATION 3-MONTHS POST-IMPLANTATION. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 15602996 ↗InjuryJul 25, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT THE PATIENT HAD A SHOULDER SURGERY, AND APPROXIMATELY 6 MONTHS LATER A REVISION SURGERY WAS DONE DUE TO THE SCREW LOOSENING AND UNSCREWING. THE ANTERIOR OBLIQUE WAS ALSO LOOSE BUT DID NOT UNSCREWED LIKE THE TRANSVERSE ONE. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.
Patient problems: Failure of Implant
MAUDE report 15093005 ↗InjuryFeb 23, 2022Migration or Expulsion of Device; Migration or Expulsion of Device
IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN NAIL SYSTEM. AFTER 4 WEEKS FROM THE INITIAL, SURGEON FOUND THREE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. THE SURGEON KEEP AN EYE ON THE PATIENT CONDITION AS WELL AS NO REVISION WILL BE PLANNED SO FAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13583778 ↗InjuryApr 28, 2021Migration or Expulsion of Device; Migration; Migration or Expulsion of Device; Migration
InjuryOct 7, 2020Migration; Migration
NO EVENT UPDATE. INVESTIGATION RESULTS ARE NOW AVAILABLE.
Patient problems: No Code Available
MAUDE report 10640620 ↗InjuryOct 6, 2020Migration; Migration
PATIENT WAS IMPLANTED ON AN UNKNOWN SIDE AND UNDERWENT REVISION SURGERY DUE TO BACKED OUT SCREW POST 4 WEEKS OF INITIAL SURGERY.
Patient problems: No Code Available
MAUDE report 10633248 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page