Skip to content

Affixus® Natural Nail® 47-2486-054-40Recalls & reported problems

Zimmer GmbH · Class 2 · Implant

All models of Affixus Natural Nail (55)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
3
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024505513), received between Aug 9, 2021 and Sep 17, 2025 — classified as 3 injuries, 1 malfunction.

The most frequently reported specific device problem is “Migration” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 1 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Migration3
  • Migration or Expulsion of Device2

Patient problems

  • Insufficient Information2
  • No Clinical Signs, Symptoms or Conditions2
  • Failure of Implant1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjurySep 17, 2025Migration; Migration

    IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN NAIL. ONE WEEK AFTER THE INITIAL SURGERY, SURGEON FOUND ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. TWO WEEKS AFTER THE INITIAL SURGERY, ALL OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION AND SUBSEQUENTLY UNDERWENT A REVISION SURGERY, DURING WHICH, THE SURGEON CONFIRMED THAT THE CORELOCK HAD BECOME LOOSE. THEN THE SCREW DEPTH WAS READJUSTED, THE CORELOCK WAS RETIGHTENED, AND THE REVISION WAS COMPLETED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: Insufficient Information

    MAUDE report 23074173 ↗
  • MalfunctionJul 24, 2025Migration; Migration

    IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL SURGERY WITH ANN NAIL. SUBSEQUENTLY, 9 DAYS POST IMPLANTATION, SURGEON FOUND THREE OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. THE SURGEON WILL KEEP AN EYE ON THE PATIENT CONDITION AS WELL, AS NO REVISION WILL BE PLANNED SO FAR. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22586998 ↗
  • InjuryNov 25, 2021Migration or Expulsion of Device; Migration or Expulsion of Device

    INVESTIGATION RESULTS ARE NOW AVAILABLE.

    Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions

    MAUDE report 12873598 ↗
  • InjuryAug 9, 2021Migration or Expulsion of Device; Migration; Migration or Expulsion of Device; Migration

    IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED ON AN UNKNOWN SIDE. ONE MONTH FROM THE INITIAL SURGERY THE SURGEON NOTICED THAT ONE OF THE PROXIMAL SCREWS HAD MIGRATED FROM ITS POSITION. A REVISION HAS NOT YET TAKEN PLACE. THE TORQUE DRIVER WAS USED TO ENGAGE THE CORELOCK, FURTHERMORE NON-TORQUE DRIVER WAS USED TO TIGHTEN MORE AFTER USING THE TORQUE DRIVER.

    Patient problems: Insufficient Information

    MAUDE report 12294072 ↗

Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices

All ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page