Affixus® Natural Nail® 47-2486-054-40Recalls & reported problems
Zimmer GmbH · Class 2 · Implant
All models of Affixus Natural Nail (55)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 3
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024505513), received between Aug 9, 2021 and Sep 17, 2025 — classified as 3 injuries, 1 malfunction.
The most frequently reported specific device problem is “Migration” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration3
- Migration or Expulsion of Device2
Patient problems
- Insufficient Information2
- No Clinical Signs, Symptoms or Conditions2
- Failure of Implant1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
InjurySep 17, 2025Migration; Migration
IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN NAIL. ONE WEEK AFTER THE INITIAL SURGERY, SURGEON FOUND ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. TWO WEEKS AFTER THE INITIAL SURGERY, ALL OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION AND SUBSEQUENTLY UNDERWENT A REVISION SURGERY, DURING WHICH, THE SURGEON CONFIRMED THAT THE CORELOCK HAD BECOME LOOSE. THEN THE SCREW DEPTH WAS READJUSTED, THE CORELOCK WAS RETIGHTENED, AND THE REVISION WAS COMPLETED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 23074173 ↗MalfunctionJul 24, 2025Migration; Migration
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL SURGERY WITH ANN NAIL. SUBSEQUENTLY, 9 DAYS POST IMPLANTATION, SURGEON FOUND THREE OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. THE SURGEON WILL KEEP AN EYE ON THE PATIENT CONDITION AS WELL, AS NO REVISION WILL BE PLANNED SO FAR. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22586998 ↗InjuryNov 25, 2021Migration or Expulsion of Device; Migration or Expulsion of Device
INVESTIGATION RESULTS ARE NOW AVAILABLE.
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 12873598 ↗InjuryAug 9, 2021Migration or Expulsion of Device; Migration; Migration or Expulsion of Device; Migration
IT WAS REPORTED THAT THE PATIENT WAS IMPLANTED ON AN UNKNOWN SIDE. ONE MONTH FROM THE INITIAL SURGERY THE SURGEON NOTICED THAT ONE OF THE PROXIMAL SCREWS HAD MIGRATED FROM ITS POSITION. A REVISION HAS NOT YET TAKEN PLACE. THE TORQUE DRIVER WAS USED TO ENGAGE THE CORELOCK, FURTHERMORE NON-TORQUE DRIVER WAS USED TO TIGHTEN MORE AFTER USING THE TORQUE DRIVER.
Patient problems: Insufficient Information
MAUDE report 12294072 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page