Affixus® Natural Nail® 47-2496-160-07Recalls & reported problems
Zimmer GmbH · Class 2 · Implant
All models of Affixus Natural Nail (55)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 6
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 00889024505773), received between Jun 8, 2021 and Mar 30, 2026 — classified as 6 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Migration” (4 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Migration4
- Separation Problem2
- Malposition of Device1
- Mechanical Problem1
- Migration or Expulsion of Device1
- Use of Device Problem1
Patient problems
- Failure of Implant3
- Insufficient Information3
- No Clinical Signs, Symptoms or Conditions2
- Pain2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Malposition of Device
IT WAS REPORTED FROM A RETROSPECTIVE CLINICAL STUDY THAT A PATIENT HAD AN INITIAL INTRAMEDULLARY NAILING OF THE RIGHT HUMERUS. SUBSEQUENTLY, THE PATIENT UNDERWENT REMOVAL OF ALL HARDWARE APPROXIMATELY 7 MONTHS LATER DUE TO NAIL PROMINENCE. X-RAYS REVEALED RADIOLOGICAL PROTRUSION WITHOUT CLINICAL SYMPTOMS. THE PATIENT REMAINED SATISFIED THROUGHOUT THE STUDY, AND THE BONE HAD HEALED PRIOR TO EXPLANTATION. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24720093 ↗MalfunctionJul 1, 2025Separation Problem; Separation Problem
IT WAS REPORTED THAT THE PATIENT HAD AN INITIAL SURGERY WITH ANN NAIL SYSTEM. SUBSEQUENTLY, 42 DAYS POST IMPLANTATION, SURGEON FOUND THAT 3 PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. THE SURGEON KEEPS AN EYE ON THE PATIENT CONDITION AS WELL AS NO REVISION WILL BE PLANNED SO FAR. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 22370765 ↗InjuryJan 8, 2025Separation Problem; Separation Problem
IT WAS REPORTED THAT AN ANN NAIL SURGERY WAS PERFORMED. AFTER 1 MONTH FROM THE INITIAL SURGERY, THE SURGEON FOUND THAT 3 PROXIMAL SCREWS HAD MIGRATED OUT OF THE PROPER POSITION. THE PATIENT COMPLAINED OF PAIN AND REVISION SURGERY WAS PERFORMED. IN THE REVISION SURGERY, THE SCREWS THAT HAD COME LOOSE WERE REINSERTED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Failure of Implant; Pain
MAUDE report 21105998 ↗InjurySep 15, 2023Migration; Migration
IT WAS REPORTED THAT AFTER A YEAR POST RIGHT-SIDE IMPLANTATION SURGEON FOUND ONE OF THE SCREWS BACKED OUT FROM THE PROPER POSITION. REVISION SURGERY WAS PERFORMED FIVE MONTHS LATER TO REMOVE THE MIGRATED SCREW. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 17754137 ↗InjuryApr 25, 2023Migration; Migration
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Failure of Implant
MAUDE report 16804868 ↗InjuryMay 18, 2022Migration; Migration
IT WAS REPORTED THAT ONE OF THE PROXIMAL SCREWS BACKED OUT FROM THE PROPER POSITION ONE WEEK AFTER INITIAL IMPLANTATION. THE SURGEON IS KEEPING AN EYE ON THE PATIENT'S CONDITION. NO REVISION SURGERY PLANNED SO FAR. NO HARM REPORTED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 14433193 ↗InjuryFeb 23, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN NAIL SYSTEM. AFTER 4 WEEKS FROM THE INITIAL, SURGEON FOUND THREE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION. THE SURGEON KEEP AN EYE ON THE PATIENT CONDITION AS WELL AS NO REVISION WILL BE PLANNED SO FAR.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13583785 ↗InjuryJun 8, 2021Mechanical Problem; Migration or Expulsion of Device; Migration; Mechanical Problem; Migration or Expulsion of Device; Migration
NO EVENT UPDATE. INVESTIGATION RESULTS ARE AVAILABLE.
Patient problems: Insufficient Information; Pain
MAUDE report 11957629 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page