- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889024505919), received between Dec 21, 2023 and Jan 26, 2026 — classified as 2 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Separation Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Separation Problem3
- Migration1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- Failure of Implant1
- Insufficient Information1
- Pain1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 26, 2026Separation Problem
IT WAS REPORTED THAT THE PATIENT UNDERWENT AN INITIAL PROCEDURE WITH AN INTRAMEDULLARY NAIL, AFTER WHICH ONE PROXIMAL SCREW BACKED OUT APPROXIMATELY 2 WEEKS POSTOPERATIVELY. THE PATIENT REMAINED UNDER CLINICAL OBSERVATION AND NO REVISION HAS BEEN PLANNED TO DATE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24176236 ↗InjurySep 17, 2025Separation Problem; Separation Problem
IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN NAIL. ONE WEEK AFTER THE INITIAL SURGERY, SURGEON FOUND ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. TWO WEEKS AFTER THE INITIAL SURGERY, ALL OF THE PROXIMAL SCREWS WERE BACKED OUT FROM THE PROPER POSITION AND SUBSEQUENTLY UNDERWENT A REVISION SURGERY, DURING WHICH, THE SURGEON CONFIRMED THAT THE CORELOCK HAD BECOME LOOSE. THEN THE SCREW DEPTH WAS READJUSTED, THE CORELOCK WAS RETIGHTENED, AND THE REVISION WAS COMPLETED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Insufficient Information
MAUDE report 23074147 ↗InjurySep 10, 2024Separation Problem; Separation Problem
IT WAS REPORTED THAT A PATIENT HAD A REVISION SURGERY ONE WEEK AFTER INITIAL IMPLANTATION DUE TO A PROXIMAL SCREW THAT WAS MIGRATED OUT FROM THE PROPER POSITION. THE PATIENT COMPLAINED OF PAIN AND A REVISION SURGERY WAS PERFORMED TO REMOVE THE MIGRATED SCREW. A NEW SCREW WAS ALSO INSERTED AT #1, AND ANOTHER NEW SCREW WAS INSERTED AT #4. DUE DILIGENCE HAS BEEN COMPLETED FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED.
Patient problems: Failure of Implant; Pain
MAUDE report 20179169 ↗MalfunctionDec 21, 2023Migration; Migration
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18379001 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page