Affixus® Natural Nail® 47-2496-221-08Recalls & reported problems
Zimmer GmbH · Class 2 · Implant
All models of Affixus Natural Nail (55)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 6
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 00889024505933), received between Jul 26, 2021 and Aug 15, 2024 — classified as 5 injuries, 1 malfunction.
The most frequently reported specific device problem is “Separation Problem” (3 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Separation Problem3
- Adverse Event Without Identified Device or Use Problem1
- Mechanical Problem1
- Migration1
- Migration or Expulsion of Device1
- Use of Device Problem1
Patient problems
- Failure of Implant4
- Pain3
- Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available1
- Insufficient Information1
- No Clinical Signs, Symptoms or Conditions1
Recent adverse event reports
The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.
InjuryAug 15, 2024Separation Problem; Separation Problem
IT WAS REPORTED THAT AFTER 2 MONTHS FROM THE INITIAL SURGERY, THE SURGEON FOUND THAT THE #1 AND #3 PROXIMAL SCREWS HAD MIGRATED OUT OF THE PROPER POSITION. THE PATIENT COMPLAINED OF PAIN AND REVISION SURGERY WAS PERFORMED TO REMOVE THE MIGRATED SCREWS. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Failure of Implant; Pain
MAUDE report 19994491 ↗InjuryJun 20, 2024Separation Problem; Separation Problem
IT WAS REPORTED PATIENT UNDERWENT REVISION APPROXIMATELY 4 WEEKS FROM THE INITIAL SURGERY, SURGEON FOUND ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. THE PATIENT COMPLAINED OF PAIN AND A REVISION WAS PERFORMED TO REMOVE THE MIGRATED SCREW. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Failure of Implant; Pain
MAUDE report 19579343 ↗MalfunctionJun 5, 2024Separation Problem; Separation Problem
IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH (B)(6). AFTER 8 WEEKS POST-IMPLANTATION, SURGEON FOUND ONE OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. THE SURGEON WILL KEEP AN EYE ON THE PATIENT CONDITION AS WELL, AS NO REVISION HAS BEEN PLANNED. DILIGENCE IS COMPLETED AND NO FURTHER INFORMATION ON THE REPORTED EVENT HAS BEEN PROVIDED.
Patient problems: Failure of Implant
MAUDE report 19470351 ↗InjuryFeb 9, 2023Migration or Expulsion of Device; Adverse Event Without Identified Device or Use Problem; Migration or Expulsion of Device; Adverse Event Without Identified Device or Use Problem
IT WAS REPORTED THAT TWO (2) WEEKS AFTER INITIAL SURGERY, THE SURGEON FOUND ONE OF THE PROXIMAL SCREW WAS BACKED OUT FROM THE PROPER POSITION. REVISION SURGERY IS NOT PLANNED, AND THE PATIENT'S CONDITION WILL CONTINUE TO BE MONITORED. DUE DILIGENCE IS IN PROGRESS FOR THIS COMPLAINT; TO DATE NO ADDITIONAL INFORMATION OR PRODUCT HAS BEEN RECEIVED. .
Patient problems: Failure of Implant; No Clinical Signs, Symptoms or Conditions
MAUDE report 16331866 ↗InjuryFeb 23, 2022Use of Device Problem; Use of Device Problem
IT WAS REPORTED THAT OPERATION WAS PERFORMED WITH ANN NAIL SYSTEM. AFTER 2 MONTHS FROM THE INITIAL, SURGEON FOUND A SCREW OF THE PROXIMAL SCREWS WAS BACKED OUT FROM THE PROPER POSITION. THE SURGEON KEEP AN EYE ON THE PATIENT CONDITION AS WELL AS NO REVISION WILL BE PLANNED SO FAR.
Patient problems: Insufficient Information
MAUDE report 13583791 ↗InjuryJul 26, 2021Mechanical Problem; Migration; Mechanical Problem; Migration
NO EVENT UPDATE. INVESTIGATION RESULTS ARE NOW AVAILABLE.
Patient problems: Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available; Pain
MAUDE report 12223776 ↗
Other ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES devices
- NA 03.033.001Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES265 reports
- GAMMA 3125-1180SStryker GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES249 reports
- NA 03.037.026Synthes GmbH · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES225 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-075Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls2 reports
- PRECICE® Intramedullary Limb Lengthening System LSC5-045Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- PRECICE® Intramedullary Limb Lengthening System LSC5-050Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls1 report
- TRIGEN 71674067Smith & Nephew, Inc. · ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES201 reports
- PRECICE Intramedullary Limb Lengthening System LSB5-040Nuvasive Specialized Orthopedics, Inc. · Rod, Fixation, Intramedullary And Accessories4 recalls0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗