Sesamoid ® 20-8000-000-11Recalls & reported problems
Orthosoft Inc · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 7
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 00889024563636), received between Aug 2, 2021 and Feb 10, 2026 — classified as 7 injuries, 2 malfunctions.
The most frequently reported specific device problem is “Fracture” (7 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture7
- Material Fragmentation2
Patient problems
- Foreign Body In Patient8
- No Clinical Signs, Symptoms or Conditions2
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
InjuryFeb 10, 2026Fracture
IT WAS REPORTED THAT DURING A ROBOT-ASSISTED TOTAL KNEE ARTHROPLASTY, A FIXATION FLUTED PIN (3.2 MM X 80 MM) FRACTURED DURING REMOVAL FROM THE TIBIA, RESULTING IN A RETAINED FRAGMENT. INTRA-OPERATIVELY, THE PIN BROKE AS THE SURGEON ATTEMPTED TO REMOVE IT; A PORTION WAS RETRIEVED AND THE REMAINING SEGMENT EMBEDDED IN THE TIBIA WAS LEFT IN PLACE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 24306009 ↗InjuryAug 7, 2025Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Foreign Body In Patient
MAUDE report 22737799 ↗InjuryOct 17, 2024Fracture; Fracture
IT WAS REPORTED THAT THE FIX TIBIAL PINS FRACTURED DURING A ROSA TKA PROCEDURE. THE TIPS BROKE WHILE INSERTING THE PINS INTO THE TIBIA. THE BROKEN TIPS WERE LEFT INSIDE THE PATIENT'S BONE, AND A CONTROL X-RAY WAS TAKEN AT THE END OF THE SURGICAL PROCEDURE. DUE DILIGENCE IS COMPLETE AS MULTIPLE ATTEMPTS WERE MADE; HOWEVER, NO FURTHER INFORMATION IS AVAILABLE. AS NO ADDITIONAL INFORMATION OR PRODUCT IS AVAILABLE, WE ARE UNABLE TO PROVIDE FURTHER INFORMATION.
Patient problems: Foreign Body In Patient
MAUDE report 20472561 ↗MalfunctionApr 3, 2024Fracture; Fracture
IT WAS REPORTED PATIENT UNDERWENT AN INITIAL ROSA KNEE PROCEDURE. AFTER THE PROCEDURE, THE TRACKERS WERE REMOVED, AND A SMALL PIECE OF THE 80 MM FIX FLUTED PIN HAD BROKEN OFF AND WAS BELIEVED TO HAVE REMAINED IN THE PATIENT¿S POSTERIOR CORTEX. THE BROKEN TIP MEASURED AROUND 1MM. THE SURGEON EXPRESSED NO CONCERN AND DID NOT FEEL AN XR WAS NEEDED. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Foreign Body In Patient; No Clinical Signs, Symptoms or Conditions
MAUDE report 19033178 ↗InjuryNov 29, 2023Fracture; Fracture
IT WAS REPORTED WHEN REMOVING THE 80MM PIN TOWARD THE END OF THE SURGERY FROM THE TIBIA, THE TIP SNAPPED OFF AND REMAINED IN THE BONE. APPROXIMATELY 5MM OF THE TIP REMAINED IN THE PATIENT. ATTEMPTS HAVE BEEN MADE AND ADDITIONAL INFORMATION ON THE REPORTED EVENT IS UNAVAILABLE AT THIS TIME.
Patient problems: Foreign Body In Patient
MAUDE report 18229176 ↗MalfunctionApr 25, 2023Fracture; Fracture
IT WAS REPORTED DURING AN INITIAL KNEE PROCEDURE, THE SURGEON PLACED THE PINS FOR THE NAVITRACKER INTO THE TIBIA. SUBSEQUENTLY, THE 80MM PIN BROKE OFF IN THE PROCESS. THE PIN WAS COMPLETELY REMOVED FROM THE TIBIA. A COMPETITOR MACHINE WITH QUICK-RELEASE FASTENER WAS USED TO PLACE THE PIN. NO PATIENT HARM OR IMPACT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16807332 ↗InjuryMay 25, 2022Material Fragmentation; Material Fragmentation
IT WAS REPORTED DURING TKA USING ROBOTIC ASSISTED INSTRUMENTATION, A PIN SNAPPED WITHIN THE PATIENT. PROPER SURGICAL TECHNIQUE WAS USED WHEN INSERTING THE PINS INTO THE TIBIA. FIRST PIN WAS PLACE PROXIMALLY AND SECOND PIN DISTAL TO THAT. THE DISTAL PIN WENT THROUGH THE NEAR CORTEX FINE AND PROCEEDED TO 'STOP' PROGRESSING INTO THE BONE. WHEN THE SURGEON ATTEMPTED TO BACK THE PIN OUT AND INSPECT WHAT OCCURRED. HE NOTICED THAT THE PIN WAS BROKEN. IT WAS NOTED THAT THE PATIENT DID NOT HAVE UNUSUALLY DENSE BONE NOR WEAK BONE. THIS EVENT OCCURRED AT THE TIME OF PLACING THE TRACKERS PRIOR TO REGISTRATION. THERE WAS A 30-45 MINUTE DELAY IN PROCEDURE. ADDITIONAL SKIN INCISION AND PIN REQUIRED. THE SURGEON ATTEMPTED TO REMOVE THE PIN INTRA-OPERATIVELY BUT WAS UNABLE TO DO SO AS THE PIN ANCHORED INTO POSTERIOR CORTICAL BONE. NO KNOWN IMPACT TO THE PATIENT, AT THIS TIME THE PIN REMAINS IMPLANTED IN PATIENT. ATTEMPTS FOR ADDITIONAL INFORMATION HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: Foreign Body In Patient
MAUDE report 14498905 ↗InjuryFeb 18, 2022Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT DURING A TKA UTILIZING ROSA KNEE INSTRUMENTATION THE TIP OF THE FIXATION PIN BROKE OFF UPON INSERTION INTO THE TIBIA. THE COMPONENT THAT BROKE OFF WAS RETAINED IN THE PATIENTS TIBIA AND WAS NOT REMOVED. NEW PINS WERE OPENED AND THE PROCEDURE WAS COMPLETED SUCCESSFULLY. THERE WAS NO ADDITIONAL PATIENT IMPACT REPORTED AND NO FURTHER INFORMATION IS AVAILABLE.
Patient problems: Foreign Body In Patient
MAUDE report 13562081 ↗InjuryAug 2, 2021Fracture; Fracture
NO ADDITIONAL EVENT INFORMATION TO REPORT AT THIS TIME.
Patient problems: Foreign Body In Patient
MAUDE report 12262575 ↗
Other Orthopedic Stereotaxic Instrument devices
- REAL INTELLIGENCE ROB10024Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall924 reports
- NA 209063Mako Surgical Corp. · Orthopedic stereotaxic instrument2 recalls602 reports
- RESTORIS, RIO 212760Mako Surgical Corp. · Orthopedic stereotaxic instrument605 reports
- REAL INTELLIGENCE ROB10013Smith & Nephew, Inc. · Orthopedic stereotaxic instrument1 recall528 reports
- VELYS 4515-70-105Depuy (Ireland) · Orthopedic stereotaxic instrument578 reports
- VELYS 4515-70-106Depuy (Ireland) · Orthopedic stereotaxic instrument564 reports
- NA 209999Mako Surgical Corp. · Orthopedic stereotaxic instrument522 reports
- Mazor X System TPL0059-06Mazor Robotics Ltd · Orthopedic Stereotaxic Instrument2 recalls2 deaths392 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page