Alliance® SBGL3701Recalls & reported problems
Zimmer, Inc. · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 11
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 11 adverse event reports that list this exact device (DI 00889024575202), received between May 13, 2021 and May 29, 2026 — classified as 1 injury, 10 malfunctions.
The most frequently reported specific device problem is “Fracture” (11 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fracture11
Patient problems
- No Clinical Signs, Symptoms or Conditions10
- Insufficient Information1
Recent adverse event reports
The 11 most recent of 11 reports, in the reporter's own words as published by the FDA.
InjuryMay 29, 2026Fracture
IT WAS REPORTED THAT A PORTION OF THE AUGMENT REAMER DRIVER FRACTURED INSIDE THE PATIENT DURING A PROCEDURE. AN ADDITIONAL INCISION WAS MADE IN ORDER TO REMOVE THE FRACTURED PORTION. ATTEMPTS HAVE BEEN MADE AN NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: Insufficient Information
MAUDE report 25331779 ↗MalfunctionMay 22, 2026Fracture
IT WAS REPORTED THAT DURING REAMING FOR AN AUGMENT, THE TIP OF THE REAMER DRIVER SHEARED OFF INTO THE AUGMENT REAMER. DURING INTRA-OPERATIVE PREPARATION, WHILE THE SURGEON WAS REAMING FOR AN AUGMENT, THE OFF-AXIS REAMER TIP FRACTURED AND REMAINED WITHIN THE AUGMENT REAMER ASSEMBLY. THE SURGEON HAD COMPLETED THE REAMING PORTION AND WAS ABLE TO FINISH THE CASE AS PLANNED. THERE WAS NO KNOWN IMPACT OR CONSEQUENCE TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25256116 ↗MalfunctionMar 31, 2026Fracture
IT WAS REPORTED THAT THE BALL HEX REAMER DRIVER WAS FRACTURED DURING A PROCEDURE. NO PATIENT IMPACT WAS REPORTED. ATTEMPTS HAVE BEEN MADE AND NO FURTHER EVENT INFORMATION IS AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24743094 ↗MalfunctionMar 20, 2026Fracture
IT WAS REPORTED THAT DURING GLENOID AUGMENT REAMING THE TIP OF A BALL HEX AUGMENT REAMER BROKE. THE EVENT OCCURRED INTRA-OPERATIVELY; ALL FRAGMENTS WERE RETRIEVED, AND A NEW INSTRUMENT TRAY WAS OPENED TO REPLACE THE BROKEN INSTRUMENT AND PROCEED. THERE WERE NO CONSEQUENCES OR IMPACT TO THE PATIENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24650147 ↗MalfunctionJan 6, 2026Fracture
IT WAS REPORTED THAT THE ROUNDED HEX TIP OF A REAMER DRIVER BROKE AS THE SURGEON FINISHED REAMING FOR A GLENOID AUGMENT. THE EVENT OCCURRED INTRA-OPERATIVELY DURING GLENOID PREPARATION; THE DRIVER WAS USED WITH POWER, WHICH WAS BELIEVED TO HAVE CONTRIBUTED TO THE BREAKAGE. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23975472 ↗MalfunctionNov 21, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A PROCEDURE, THE HEAD OF THE REAMER SNAPPED WHILE REAMING FOR THE AUGMENTED ALLIANCE IMPLANTS. NOTHING FELL INTO THE PATIENT AND NO HARM WAS REPORTED AS A RESULT OF THE EVENT. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23621901 ↗MalfunctionJul 21, 2025Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22560490 ↗MalfunctionJun 24, 2025Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22319665 ↗MalfunctionJun 9, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11966307 ↗MalfunctionMay 24, 2021Fracture; Fracture
IT WAS REPORTED THAT THE OFF AXIS REAMER HIT THE DELTOID RETRACTOR WHILE REAMING ON POWER. CAUSED THE OFF AXIS DRIVER HEAD TO SNAP OFF. ATTEMPTS HAVE BEEN MADE AND NO FURTHER INFORMATION HAS BEEN PROVIDED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11871946 ↗MalfunctionMay 13, 2021Fracture; Fracture
NO FURTHER EVENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11821097 ↗
Other PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED devices
- DJO SURGICAL 508-32-204Encore Medical, L.P. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall544 reports
- Comprehensive Reverse Shoulder 010000589Biomet Orthopedics, LLC · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED1 recall142 reports
- DELTA XTEND 130738203Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED175 reports
- DJO SURGICAL 509-00-432Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented1 recall117 reports
- Alliance® SBGL3007Zimmer, Inc. · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED162 reports
- DELTA XTEND 130760138Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED146 reports
- DELTA XTEND 130790030Depuy (Ireland) · PROSTHESIS, SHOULDER, SEMI-CONSTRAINED, METAL/POLYMER CEMENTED132 reports
- DJO SURGICAL 509-00-436Encore Medical, L.P. · Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented124 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page