CREO 1067.4745Recalls & reported problems

Globus Medical, Inc. · 7.5 x 45mm Cannulated Screw, Modular, CREO AMP · Class 2

All models of CREO (100)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889095023879), received between Mar 5, 2024 and Apr 21, 2026 — classified as 1 injury, 3 malfunctions.

The most frequently reported specific device problem is “Use of Device Problem” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 3 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520263
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Use of Device Problem2
  • Device Slipped1
  • Migration1

Patient problems

  • Spinal Column Injury2
  • Insufficient Information1
  • Unspecified Musculoskeletal problem1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • InjuryApr 21, 2026Migration

    IT WAS REPORTED THAT ON (B)(6) 2025 THE PATIENT UNDERWENT A SPINAL PROCEDURE USING CREO MIS. THEN ON (B)(6) 2026 THE REVISION SURGERY WAS PERFORMED DUE TO THE LOOSENING OF PEDICLE SCREWS AT L5. THE FIXATION AREA WAS EXTENDED FROM L2 TO ILIAC. THE SALE REP COMMENTED THAT THE PATIENT'S BONE QUALITY WAS POOR. THE LOCKING CAPS WERE OBTAINED, BUT THE LOOSE MIS SCREWS WERE NOT.

    Patient problems: Unspecified Musculoskeletal problem

    MAUDE report 24940788 ↗
  • MalfunctionMar 5, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT A REVISION SURGERY WAS PERFORMED TO REPLACE A CREO AMP 7.5X45MM MODULAR CANNULATED SCREW THAT MIGRATED POST OPERATIVELY. THIS EVENT OCCURRED IN JAPAN.

    Patient problems: Spinal Column Injury

    MAUDE report 18837673 ↗
  • MalfunctionMar 5, 2024Use of Device Problem; Use of Device Problem

    IT WAS REPORTED THAT A CREO AMP MODULAR SCREW WAS FOUND TO HAVE DAMAGED THREADS DURING SURGERY. THIS EVENT OCCURRED IN JAPAN.

    Patient problems: Spinal Column Injury

    MAUDE report 18837333 ↗
  • MalfunctionMar 5, 2024Device Slipped; Device Slipped

    IT WAS REPORTED THERE WAS A REVISION SURGERY NEEDED TO REPLACE A CREO SCREW THAT WAS FOUND LOOSE AT L5 POST OPERATIVELY. THIS EVENT OCCURRED IN JAPAN.

    Patient problems: Insufficient Information

    MAUDE report 18834290 ↗

Other Orthosis, spinal pedicle fixation, for degenerative disc disease devices

All Orthosis, spinal pedicle fixation, for degenerative disc disease devices →