INFINITY 33600030Recalls & reported problems
Wright Medical Technology, Inc. · Resection Adjustment Block · Class 2
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 4
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 00889797004008), received between May 9, 2022 and Oct 18, 2022 — classified as 1 injury, 3 malfunctions.
The most frequently reported specific device problem is “Mechanical Jam” (2 reports).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Recalls naming this device
- Mar 6, 2024Open, Classified
Stryker has identified an issue that impacts specific lots of Infinity" Resection Guide Adjustment Blocks. The parts within these three lots were found to have been missing an internal screw in the finished instrument assembly. The missing screw is used to lockout the Medial/Lateral adjustment on the INFINITY Resection Adjustment Guide.
Z-1477-2024 · Root cause: Process change control · FDA record ↗
Most reported problems
Device problems
- Mechanical Jam2
- Break1
- Device Difficult to Setup or Prepare1
- Fracture1
- Material Twisted/Bent1
Patient problems
- No Clinical Signs, Symptoms or Conditions3
- Insufficient Information1
Recent adverse event reports
The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 18, 2022Break; Fracture; Break; Fracture
IT WAS REPORTED THAT THE PHYSICIAN USED A SCREWDRIVER TO LOCK IN THE PURPLE SCREW, THE CONNECTION BETWEEN THE PURPLE SCREW AND THE BOTTOM WAS BROKEN.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15625192 ↗MalfunctionOct 14, 2022Material Twisted/Bent; Mechanical Jam; Material Twisted/Bent; Mechanical Jam
THE ADJUSTMENT BLOCK WAS NOT EASY TO TURN THE KNOBS ON, WHEN ADJUSTED WITH SCREWDRIVER THE GREEN KNOBS AXEL BENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15611309 ↗MalfunctionAug 9, 2022Mechanical Jam; Mechanical Jam
IT WAS NOTICED THE RESECTION ADJUSTMENT BLOCK WOULDN'T TRANSLATE PROXIMAL DISTAL AT THE END OF THE PROCEDURE. THE PURPLE KNOB IS LOCKED AND WON¿T REALLY TURN WITH SIGNIFICANT FORCE APPLIED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15195593 ↗InjuryMay 9, 2022Device Difficult to Setup or Prepare; Device Difficult to Setup or Prepare
IT WAS REPORTED THAT THE SURGEON WAS PERFORMING A CONVENTIONAL INFINITY TOTAL ANKLE REPLACEMENT WENT THAT WRONG. AFTER INITIAL SIZING AND CHECKING, THE CUTS WERE MADE. THEN THE CUTS LOOKED DIFFERENT FROM WHAT WE SIZED SO WE WENT BACK TO CHECK THE SIZING AGAIN AND THEN THE SIZING WAS COMPLETELY DIFFERENT FROM THE PREVIOUS PICTURES AND THE CUTS WERE TOO WRONG SO WE COULDN¿T PROCEED WITH AN INFINITY PROCEDURE. THE PATIENT WAS CLOSED UP AND A WEEK LATER THEY WENT IN FOR A ANKLE ARTHRODESIS.
Patient problems: Insufficient Information
MAUDE report 14337006 ↗
Other PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER devices
- Vantage 350-22-03Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls2 reports
- Vantage 350-21-04Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer2 recalls0 reports
- INFINITY 33630023Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER60 reports
- INFINITY 33600020Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER1 recall9 reports
- INFINITY 33650003Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER55 reports
- Vantage 350-22-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall5 reports
- INFINITY 33630022Wright Medical Technology, Inc. · PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER54 reports
- Vantage 350-21-02Exactech, Inc. · Prosthesis, Ankle, Semi-Constrained, Cemented, Metal/Polymer1 recall4 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.