Skip to content

Straight In/Out Pt Line Kit 902-00037Recalls & reported problems

Belmont Instrument Corporation · w/ 4.4L Reservoir · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.

Adverse event reports
3
Deaths reported
0
Injuries reported
0
Recalls naming it
1

Summary

The FDA's MAUDE database contains 3 adverse event reports that list this exact device (DI 00896128002213), received between Jan 6, 2021 and Jun 28, 2024 — classified as 3 malfunctions.

The most frequently reported specific device problem is “Fluid/Blood Leak” (2 reports).

1 FDA recall record name this device's identifier, 1 still open.

2 other recalls cover devices cleared under the same FDA submission (K152208); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Recalls naming this device

  1. Mar 31, 2025Class IIOpen, Classified

    Potential crack in the female connector located on the Heat Exchanger in the set, which will lead to a fluid leak through the crack during priming of the rapid infuser.

    Z-1675-2025 · Root cause: Process design · Recall details · FDA record ↗

Most reported problems

Device problems

  • Fluid/Blood Leak2
  • Leak/Splash1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3

Recent adverse event reports

The 3 most recent of 3 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 28, 2024Leak/Splash; Leak/Splash

    BELMONT RECEIVED MEDWATCH #MW5154726 ON MAY 30TH, 2024 HAVING BELOW EVENT DESCRIPTION: DURING THE HIPEC THE HEAT EXCHANGER DISPOSABLE SET BEGAN TO LEAK SHORTLY AFTER THE SECOND DOSE OF MITOMYCIN-C WAS ADMINISTERED. UPON QUICKLY NOTICING THE LEAK, THE HIPEC WAS PAUSED AND THE DISPOSABLE SET THAT WAS LEAKING WAS EXCHANGED WITH A NEW ONE. IT APPEARED TO BE LEAKING FROM A CONNECTION THAT COMES PREMADE / GLUED FROM THE MANUFACTURER- THE RED PIECE OF TUBING THAT IS CONNECTED TO THE TOP OF THE ROUND METAL HEAT EXCHANGER ITSELF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19637032 ↗
  • MalfunctionSep 25, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE USER FACILITY REPORTED THAT THE HYPERTHERMIA PUMP DISPOSABLE SET LEAKED AT THE HEAT EXCHANGER DURING USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12530889 ↗
  • MalfunctionJan 6, 2021Fluid/Blood Leak; Fluid/Blood Leak

    THE USER FACILITY REPORTED THAT THE HYPERTHERMIA PUMP DISPOSABLE SET LEAKED WHILE IN USE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11127915 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K152208. One submission often covers many models, so check the product and lot details in each record.

  • Z-1676-2025 · Mar 31, 2025 · Open, Classified
    Straight Inflow/Bifurcated Outflow Line Kit. Model Number: 902-00038. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Straight Inflow/Outflow Patient L
  • Z-1674-2025 · Mar 31, 2025 · Open, Classified
    Hyperthermia Pump Procedure Kit. Model Number: 902-00045. Sterile, single-use, disposable set used with the Hyperthermia Pump. The Hyperthermia Procedure kit (P/N 902-00045) co

Other Warmer, Thermal, Infusion Fluid devices

All Warmer, Thermal, Infusion Fluid devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page