Level 1 HL-90Recalls & reported problems
Icu Medical, Inc. · Class 2
Quick summary
This ICU Medical Level 1 fluid warmer heats infusion fluids before they reach a patient. All 991 reports are malfunction reports, often describing fluid leaks, alarm problems, and heating faults, and some describe a fire or burns on the device. There are no deaths or injuries. Reports fell from 221 in the prior 12 months to 14 in the last 12. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 991
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 991 adverse event reports that list this exact device (DI 50695085820909), received between Nov 7, 2017 and Aug 24, 2026 — classified as 991 malfunctions.
The most frequently reported specific device problem is “Fluid/Blood Leak” (167 reports).
11 other recalls cover devices cleared under the same FDA submission (K911383); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Fluid/Blood Leak167
- Device Alarm System136
- Temperature Problem73
- Insufficient Heating72
- Defective Alarm51
- Failure to Pump46
- Overheating of Device46
- No Audible Alarm44
Patient problems
- No Clinical Signs, Symptoms or Conditions725
- Insufficient Information222
- No Information46
- No Patient Involvement35
- No Known Impact Or Consequence To Patient9
- No Consequences Or Impact To Patient3
- Electric Shock1
Recent adverse event reports
The 25 most recent of 991 reports, in the reporter's own words as published by the FDA.
MalfunctionAug 24, 2026Overheating of Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50419551 ↗MalfunctionJul 8, 2026Overheating of Device
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25752615 ↗MalfunctionMay 18, 2026Fire
IT WAS REPORTED THAT A FIRE OCCURRED WITHIN THE DEVICE DURING PRIMING. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25201227 ↗MalfunctionMar 17, 2026Electrical Power Problem
THE UNIT WAS RETURNED BECAUSE THE UNIT HAD TRIPPED THE LIM (ELECTRICAL SAFETY ALARM) ALARM DUE TO HIGH ELECTRICAL OUTPUT, AND THERE WAS NO ERROR CODES APPEARED. IT HAD OCCURRED DURING SETUP. THE RESULTS OF THE INVESTIGATION FOUND THAT THE DEVICE HAD A CONFIRMED ELECTRICAL ISSUE AND RELATED TO THE MOTOR. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24616514 ↗MalfunctionMar 16, 2026Crack; Fluid/Blood Leak
IT WAS REPORTED THAT THE DISTILLED WATER SYSTEM CONTAINED BLOOD AND THE TRANSPARENT COVER WAS CRACKED. THE EVENT WAS NOTICED IN THE TRAUMA SURGERY OPERATING ROOM. THERE WAS NO REPORTED PATIENT INVOLVEMENT OR HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24605183 ↗MalfunctionMar 3, 2026Electrical /Electronic Property Problem
IT WAS REPORTED THAT THE POWER SWITCH WAS FAULTY AND THERE WERE BURNS ON THE PINS. THE POWER CORD AND FRONT CASE OF THE PUMP WAS LOOSE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24496161 ↗MalfunctionFeb 23, 2026Melted
IT WAS REPORTED THAT THE UNIT SMELLS LIKE PLASTIC WHEN TUNED ON, THE TRACES ON THE MAIN BOARD HAVE MELTED. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24414181 ↗MalfunctionFeb 4, 2026Device Markings/Labelling Problem
IT WAS REPORTED THAT THE DEVICE WAS NOT MAINTAINING TEMPERATURE. DURING INVESTIGATION IDENTIFIED DAMAGE TO THE DEVICE SERIAL NUMBER LABEL, THEREFORE MAKING THIS INVESTIGATION REPORTABLE. THE EVENT HAD OCCURRED DURING ROUTINE PREVENTIVE MAINTENANCE INSPECTION AND THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24258287 ↗MalfunctionJan 6, 2026Break; Electrical /Electronic Property Problem
THE CUSTOMER RETURNED THE DEVICE STATING, ¿THE OUTLET WAS BROKEN¿; DURING DEVICE EVALUATION IT WAS FOUND THE AC PLUG BLADE WAS BROKEN AND BURN MARKS WERE FOUND. THE EVENT OCCURRED DURING PRIMING AT THE FACILITY. THERE WAS NO REPORTED PATIENT INVOLVEMENT/HARM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23975682 ↗MalfunctionDec 8, 2025No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT THE DEVICE HAD LOW TEMPERATURE READING AND NO AUDIBLE ALARM. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23751478 ↗MalfunctionNov 20, 2025Temperature Problem; Excessive Heating; Temperature Problem; Excessive Heating
IT WAS REPORTED THAT THE FLUID WARMER TEMPERATURE DISPLAY WENT UP TO 44 DEGREES. THERE WAS PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23610373 ↗MalfunctionNov 13, 2025Device Alarm System; Temperature Problem; Device Alarm System; Temperature Problem
IT WAS REPORTED THAT THERE IS A TEMPERATURE ALARM OCCURRED IN THE SURGERY ROOM. THERE WAS A THERAPY DELAY, AND THE DEVICE WAS REPLACED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23544463 ↗MalfunctionNov 6, 2025Temperature Problem; Temperature Problem
IT WAS REPORTED THAT DEVICE HAS FLUCTUATING TEMPERATURE DURING TESTING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23492951 ↗MalfunctionNov 3, 2025Overheating of Device; Smoking; Overheating of Device; Smoking
IT WAS REPORTED THAT DEVICE GAVE A BURNING SMELL AND WAS SMOKING. REPORTER STATED THAT THE PRODUCT WOULD NOT POWER ON AND STATED THAT A BURNT COMPONENT WAS OBSERVED ON THE BOARD DURING SET UP. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23450875 ↗MalfunctionOct 30, 2025Circuit Failure; Circuit Failure
IT WAS REPORTED THAT THE DEVICE HAD NO AUDIBLE ALARM. DURING TESTING, IT WAS FOUND THAT THERE WAS NO AUDIBLE SOUND. THERE WAS NO PATIENT INVOLVEMENT, NO INJURY WAS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23424404 ↗MalfunctionOct 28, 2025Device Alarm System; No Audible Alarm; Device Alarm System; No Audible Alarm
IT WAS REPORTED THAT THE ALARM DID NOT ACTIVATE DURING PRIMING. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23405873 ↗MalfunctionOct 28, 2025No Audible Alarm; Temperature Problem; No Audible Alarm; Temperature Problem
IT WAS REPORTED THAT THERE WERE NO ALARMS ON, UNLESS THE DISPLAY TEMP REACHES ABOVE 43.1 CELSIUS. THE EVENT OCCURRED DURING PREVENTIVE MAINTENANCE. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23402843 ↗MalfunctionOct 15, 2025Overheating of Device; Overheating of Device
IT WAS REPORTED THAT THE DEVICE WAS OVERHEATING AND HAD FRAYED WIRES ON THE POWER CORD. THE EVENT OCCURRED UPON POWER ON. THERE WAS NO REPORT OF PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23279291 ↗MalfunctionSep 8, 2025False Alarm; False Alarm
IT WAS REPORTED THAT THE DEVICE DETECTS LOW WATER LEVEL EVEN IS FULLY FILLED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22992202 ↗MalfunctionAug 25, 2025Fluid/Blood Leak; Failure to Infuse; Fluid/Blood Leak; Failure to Infuse
IT WAS REPORTED THAT THE DEVICE HAS TEMPERATURE ISSUES ALARM 69-DEGREE CELSIUS AND LEAKING DURING POWER ON. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22891195 ↗MalfunctionAug 19, 2025Inaccurate Information; Inaccurate Information
THE SCREEN ON THIS UNIT WAS DISPLAYING ¿06¿ WHILE RUNNING. AS IN THE SCREEN NEVER CHANGES EVEN THOUGH THEY CAN VALIDATE THE TEMPERATURE WAS CHANGING. THE EVENT OCCURRED DURING ROUTINE PREVENTIVE MAINTENANCE INSPECTION.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22843015 ↗MalfunctionAug 13, 2025No Audible Alarm; No Audible Alarm
IT WAS REPORTED THAT THE HANDLE OF THE POLE CLAMP WAS BROKEN AND THE HEAT TEST ALARM DOES NOT WORK. THERE WAS NO REPORTED PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22797618 ↗MalfunctionAug 12, 2025Failure to Recalibrate; Calibration Problem; Failure to Recalibrate; Calibration Problem
IT WAS REPORTED THAT THE DEVICE WOULD NOT CALIBRATE. THE NORMAL TEMPERATURE WOULD NOT ADJUST. THE EVENT OCCURRED DURING PREVENTIVE MAINTENANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22784861 ↗MalfunctionAug 12, 2025Calibration Problem; Calibration Problem
IT WAS REPORTED THAT THE DEVICE WOULD NOT CALIBRATE. THE NORMAL TEMP WOULD NOT ADJUST. THE EVENT OCCURRED DURING PREVENTIVE MAINTENANCE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22784483 ↗MalfunctionAug 12, 2025Device Alarm System; Device Alarm System
IT WAS REPORTED THAT WHEN POWERING, THE UNIT ON IT KEEPS BEEPING. THE ISSUE OCCURRED DURING TESTING.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22783791 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K911383. One submission often covers many models, so check the product and lot details in each record.
- Z-2759-2026 · Jun 24, 2026 · Open, ClassifiedHotline Fluid Warming Set, single use device, is designed to warm blood and intravenous fluids and deliver them to the patient's intravenous access site at normothermic temperature
- Z-1444-2014 · Mar 17, 2014 · TerminatedSmiths Medical HOTLINE Fluid Warming Set L-80. Product Usage: intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures under gravity condit
- Z-1443-2014 · Mar 17, 2014 · TerminatedSmiths Medical HOTLINE Fluid Warming Set L-70. Product Usage: intended for the delivery of warmed blood and intravenous fluids at normothermic temperatures under gravity conditio
- Z-1979-2012 · Jun 27, 2012 · TerminatedSmiths Medical HOTLINE 3 Blood and Fluid Warmer, Reorder Numbers: HL-390 Product Usage: The Micro-switch is used on the HOTLINE 3 Blood and Fluid Warmer. This product, for use in
- Z-0423-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Codes L-70NI Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE
- Z-0428-2012 · Nov 21, 2011 · TerminatedLevel 1 Normothermic I.V. Fluid Administration Sets, Product Code DI-60HL , Product Usage: Single patient use sets designed for use by trained medical professionals, in conjuncti
- Z-0424-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Code L-270 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE
- Z-0425-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Code L-370 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE
- Z-0427-2012 · Nov 21, 2011 · TerminatedLevel 1 Normothermic I.V. Fluid Administration Sets, Product Code D-60HL , Product Usage: Single patient use sets designed for use by trained medical professionals, in conjunctio
- Z-0426-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Code L-80 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE R
- Z-0422-2012 · Nov 21, 2011 · TerminatedHOTLlNE Disposable Administration Sets, Product Codes L-70 Product Usage: Single patient use Sets designed for use by trained medical professionals, in conjunction with HOTLlNE R
Other Warmer, thermal, infusion fluid devices
- Level 1 HL-390Icu Medical, Inc. · Warmer, thermal, infusion fluid634 reports
- Level 1 L-70Icu Medical, Inc. · Warmer, thermal, infusion fluid1 recall385 reports
- 3-Spike Disposable Set (1 Piece) 903-00006PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall1 death97 reports
- Rapid Infuser RI-2, 120V, 750 ml/min 903-00039Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid132 reports
- Level 1 L-70NIIcu Medical, Inc. · Warmer, thermal, infusion fluid114 reports
- Rapid Infuser RI-2, 120V, 1000ml/min 903-00037Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 death76 reports
- Hyperthermia Pump Procedure Kit 902-00045Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall4 reports
- Straight Inflow/Bifurcated Outflow Pt Line Kit 902-00038Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall3 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.