Skip to content

Nephros SSU-D UltraFilter 70-0243DRecalls & reported problems

Nephros, Inc. · The Nephros SSU-D UltraFilter is intended to be used to filter water or bicarbonate concentrate used in hemodialysis devices. It assists in providing hemodialysis quality water or bicarbonate concentrate. The device is not a complete water treatment system, but serves to remove biological contaminants. Therefore it must be used in conjunction with other water treatment equipment (RO, DI, etc.). · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 00896241002220), received between Apr 3, 2026 and Apr 3, 2026 — classified as 1 malfunction.

The most frequently reported specific device problem is “Filtration Problem” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 1 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Filtration Problem1
  • Material Deformation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 3, 2026Filtration Problem; Material Deformation

    ON (B)(6) 2026 IT WAS REPORTED THAT A HEALTHCARE FACILITY RECEIVED A POSITIVE ENDOTOXIN CULTURE DOWNSTREAM OF AN SSU-D (70-0243D) FILTER ON ONE OF THEIR RO MACHINES AND OBSERVED THE DEVICE HAD DEFORMED. THERE WAS NO REPORTED ADVERSE EVENT. THE NEPHROS SSU-D IS INTENDED TO BE USED TO FILTER WATER OR BICARBONATE CONCENTRATE USED IN HEMODIALYSIS DEVICES. IT ASSISTS IN PROVIDING HEMODIALYSIS QUALITY WATER OR BICARBONATE CONCENTRATE. THE DEVICE IS NOT A COMPLETE WATER TREATMENT SYSTEM BUT SERVES TO REMOVE BIOLOGICAL CONTAMINANTS. THEREFORE, IT MUST BE USED IN CONJUNCTION WITH OTHER WATER TREATMENT EQUIPMENT (RO, DI, ETC.). THE USER FACILITY ROUTINELY TESTS THEIR WATER TO ENSURE THE REMOVAL OF ENDOTOXINS. A POSITIVE ENDOTOXIN CULTURE CAN BE CAUSED BY IMPROPER INSTALLATION, CROSS CONTAMINATION AND/OR A BREACH OF THE FILTER MEMBRANE. COMPREHENSIVE REVIEW OF THE SPECIFIC FILTER DEVICE HISTORY RECORD AND MANUFACTURING INSPECTIONS DID NOT IDENTIFY ANY PRODUCT ISSUES AND THE INVESTIGATION IS UNDERWAY TO DETERMINE THE ROOT CAUSE. ONCE THE INVESTIGATION IS COMPLETED, A SUPPLEMENTAL/FOLLOWUP REPORT WILL BE SUBMITTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24781038 ↗

Other Subsystem, Water Purification devices

All Subsystem, Water Purification devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page