Skip to content

VITEK® 2 Pipetter/Diluter Accessories Kit 21219Recalls & reported problems

Biomerieux , Inc. · VITEK 2 Pipettor / Diluter Accessory Kit containing straws for VITEK 2 and VITEK 2 XL needed to prepare sample suspensions. · Class 2

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026049782), received between Nov 14, 2025 and Nov 14, 2025 — classified as 1 malfunction.

The most frequently reported specific device problem is “Nonstandard Device” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Nonstandard Device1

Patient problems

  • No Clinical Signs, Symptoms or Conditions1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionNov 14, 2025Nonstandard Device; Nonstandard Device

    INTENDED USE: VITEK 2 PIPETTOR / DILUTER ACCESSORY KIT CONTAINING PIPETTES FOR VITEK 2 AND VITEK 2 XL NEEDED TO PREPARE SAMPLE SUSPENSIONS. ISSUE DESCRIPTION: A CUSTOMER IN FRANCE NOTIFIED BIOMERIEUX OF DAMAGE AND DEFECT OBSERVED ON ACCESSORY KIT VTK2 PIP/DIL (REF. 21219, LOT 1011213650, EXP DATE 05-MARCH-2027. THE PIPETTOR/DILUTER KIT IS SEPARATED WHERE THE FLEXIBLE TUBING MEETS THE SHOT-TUBE FITTING. THE FITTING IS BROKEN OFF. THIS CAUSES INABILITY TO DISPENSE SALINE INTO THE AST SPECIMEN TUBES. WITHOUT THE PIPETTOR/DILUTOR ASSEMBLY, THE INSTRUMENT CANNOT PROCESS NEW SAMPLES. THE CUSTOMER STATED THAT ADDITIONAL NEW PIP/DIL ASSEMBLIES IN THE SHIPPING CONTAINER ARE ALREADY BROKEN AS WELL. AT THE TIME OF THE ASSESSMENT, THERE IS NO INDICATION OR REPORT FROM THE CUSTOMER THAT THIS EVENT LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23553420 ↗

Other System, Test, Automated, Antimicrobial Susceptibility, Short Incubation devices

All System, Test, Automated, Antimicrobial Susceptibility, Short Incubation devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page