BD Phoenix™ NMIC/ID-307 449289Recalls & reported problems
Becton, Dickinson And Company · BD Phoenix™ NMIC/ID-307 · Class 2
Data updated Nov 27, 2026 from official FDA records · Editorial policy
Quick summary
This BD Phoenix bacteria identification panel has 524 adverse event reports, with no deaths and no injuries. Almost all describe incorrect results, mainly bacteria being identified as the wrong species. Reports fell from 221 in the prior 12 months to 22 in the last 12 months. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 526
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 526 adverse event reports that list this exact device (DI 30382904492893), received between Mar 25, 2021 and Sep 3, 2026 — classified as 526 malfunctions.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (521 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Nov 27, 2026.
Reports received per month
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings521
- False Positive Result3
- Device Markings/Labelling Problem1
- False Negative Result1
Patient problems
- No Clinical Signs, Symptoms or Conditions524
- Insufficient Information2
Recent adverse event reports
The 25 most recent of 526 reports, in the reporter's own words as published by the FDA.
MalfunctionSep 3, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50588370 ↗MalfunctionSep 2, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50569354 ↗MalfunctionAug 11, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50213122 ↗MalfunctionAug 6, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50165357 ↗MalfunctionAug 6, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50165356 ↗MalfunctionAug 6, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50161556 ↗MalfunctionAug 6, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50156428 ↗MalfunctionAug 6, 2026Incorrect, Inadequate or Imprecise Result or Readings
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 50156381 ↗MalfunctionMay 19, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE WAS MISIDENTIFIED. THE USER SPECIFIED ORGANISMS CITROBACTER BRAAKII AND CITROBACTER FREUNDII WERE IDENTIFIED WITH EACH REPEAT OF THE ISOLATE. THE USER REPORTED CITROBACTER SPP. TO PROVIDERS. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25218564 ↗MalfunctionMay 19, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE WAS MISIDENTIFIED. THE USER SPECIFIED ORGANISMS CITROBACTER BRAAKII AND CITROBACTER FREUNDII WERE IDENTIFIED WITH EACH REPEAT OF THE ISOLATE. THE USER REPORTED CITROBACTER SPP. TO PROVIDERS. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25217910 ↗MalfunctionApr 3, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE (PROTEUS MIRABILIS) WAS MISIDENTIFIED AS ESCHERICHIA COLI. NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 2 OF 2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24778225 ↗MalfunctionMar 19, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE FROM A URINE CULTURE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS, THE USER ALSO NOTED PERFORMING REPEAT TESTING. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE OF PATIENT IMPACT. THIS IS REPORT 2 OF 2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24638624 ↗MalfunctionMar 19, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE FROM A URINE CULTURE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS, THE USER ALSO NOTED PERFORMING REPEAT TESTING. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE OF PATIENT IMPACT. THIS IS REPORT 2 OF 2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24638620 ↗MalfunctionMar 19, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE IF PATIENT TREATMENT WAS AFFECTED. THIS IS REPORT 2 OF 2.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24638619 ↗MalfunctionFeb 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC/ID-307 A PATIENT ISOLATE WAS MISIDENTIFIED AS SHIGELLA BOYDII. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER STATED, "SHIGELLA BOYDII WAS REPORTED AND SEEN BY THE DOCTOR. A REFERENCE LAB THAT THE ISOLATE WAS SENT TO, SENT BACK A RESULT OF "SHIGELLA NOT FOUND." THE CUSTOMER SUSPECTS THAT THE PATIENT WAS TREATED BUT DOES NOT HAVE DETAILS. ALL ANTIBIOTICS FOR THE ISOLATE WERE SENSITIVE." NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 2 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24364107 ↗MalfunctionFeb 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC/ID-307 A PATIENT ISOLATE WAS MISIDENTIFIED AS SHIGELLA BOYDII. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER STATED, "SHIGELLA BOYDII WAS REPORTED AND SEEN BY THE DOCTOR. A REFERENCE LAB THAT THE ISOLATE WAS SENT TO, SENT BACK A RESULT OF "SHIGELLA NOT FOUND." THE CUSTOMER SUSPECTS THAT THE PATIENT WAS TREATED BUT DOES NOT HAVE DETAILS. ALL ANTIBIOTICS FOR THE ISOLATE WERE SENSITIVE." NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 3 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24364106 ↗MalfunctionFeb 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC/ID-307 A PATIENT ISOLATE WAS MISIDENTIFIED AS SHIGELLA BOYDII. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER STATED, "SHIGELLA BOYDII WAS REPORTED AND SEEN BY THE DOCTOR. A REFERENCE LAB THAT THE ISOLATE WAS SENT TO, SENT BACK A RESULT OF "SHIGELLA NOT FOUND." THE CUSTOMER SUSPECTS THAT THE PATIENT WAS TREATED BUT DOES NOT HAVE DETAILS. ALL ANTIBIOTICS FOR THE ISOLATE WERE SENSITIVE." NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 3 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24364104 ↗MalfunctionFeb 17, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC/ID-307 A PATIENT ISOLATE WAS MISIDENTIFIED AS SHIGELLA BOYDII. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER STATED, "SHIGELLA BOYDII WAS REPORTED AND SEEN BY THE DOCTOR. A REFERENCE LAB THAT THE ISOLATE WAS SENT TO, SENT BACK A RESULT OF "SHIGELLA NOT FOUND." THE CUSTOMER SUSPECTS THAT THE PATIENT WAS TREATED BUT DOES NOT HAVE DETAILS. ALL ANTIBIOTICS FOR THE ISOLATE WERE SENSITIVE." NO HEALTH IMPACT OR CONSEQUENCE REPORTED. THIS IS REPORT 2 OF 3.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24364103 ↗MalfunctionFeb 16, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC/ID-307 A PATIENT ISOLATE WAS MISIDENTIFIED AS SHIGELLA BOYDII. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER STATED, "SHIGELLA BOYDII WAS REPORTED AND SEEN BY THE DOCTOR. A REFERENCE LAB THAT THE ISOLATE WAS SENT TO, SENT BACK A RESULT OF "SHIGELLA NOT FOUND." THE CUSTOMER SUSPECTS THAT THE PATIENT WAS TREATED BUT DOES NOT HAVE DETAILS. ALL ANTIBIOTICS FOR THE ISOLATE WERE SENSITIVE." NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24349557 ↗MalfunctionFeb 16, 2026Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE BD PHOENIX¿ NMIC/ID-307 A PATIENT ISOLATE WAS MISIDENTIFIED AS SHIGELLA BOYDII. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER STATED, "SHIGELLA BOYDII WAS REPORTED AND SEEN BY THE DOCTOR. A REFERENCE LAB THAT THE ISOLATE WAS SENT TO, SENT BACK A RESULT OF "SHIGELLA NOT FOUND." THE CUSTOMER SUSPECTS THAT THE PATIENT WAS TREATED BUT DOES NOT HAVE DETAILS. ALL ANTIBIOTICS FOR THE ISOLATE WERE SENSITIVE." NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24349553 ↗MalfunctionDec 17, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 AN UNSPECIFIED NUMBER OF PATIENT ISOLATES (ESCHERICHIA COLI) HAD A HIGH MIC (FALSE RESISTANT) RESULT FOR THE DRUG ERTAPENEM. THE USER VERIFIED THE FINAL RESULT USING E-TEST STRIP. THE USER ALSO NOTED THE FAILURE FOR QC RESULTS. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23830480 ↗MalfunctionNov 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE FROM A URINE CULTURE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS, THE USER ALSO NOTED PERFORMING REPEAT TESTING. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23609704 ↗MalfunctionNov 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE FROM A URINE CULTURE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS, THE USER ALSO NOTED PERFORMING REPEAT TESTING. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE OF PATIENT IMPACT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23609703 ↗MalfunctionNov 20, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE (ESCHERICHIA COLI) HAD HIGH MICS FOR THE CARBAPENEM DRUG CLASS. THE USER VERIFIED THE FINAL RESULT USING A REFERENCE LABORATORY. THE USER IS UNSURE IF PATIENT TREATMENT WAS AFFECTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23605380 ↗MalfunctionOct 29, 2025Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
IT WAS REPORTED WHILE USING THE PANEL PHOENIX NMIC/ID-307 A PATIENT ISOLATE (ESCHERICHIA COLI) WAS MISIDENTIFIED AS SHIGELLA SPP. THE USER VERIFIED THE FINAL RESULT USING MALDI. NO HEALTH IMPACT OR CONSEQUENCE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23415219 ↗
FDA clearance and classification
The FDA regulates product code LON (system, test, automated, antimicrobial susceptibility, short incubation) under 21 CFR 866.1645 as a Class II device. 21 CFR 866.1645 ↗
Devices of this type normally need 510(k) clearance before sale.
Common questions about the BD Phoenix™ NMIC/ID-307 449289
Has the BD Phoenix™ NMIC/ID-307 449289 been recalled?
No FDA recall record in our data names this exact device identifier (DI 30382904492893), based on the FDA recall file dated Nov 27, 2026. Recalls are sometimes announced by product family or lot, so the manufacturer’s own notices are worth checking too.
What problems have been reported with the BD Phoenix™ NMIC/ID-307 449289?
The FDA's MAUDE database lists 526 adverse event reports for this device, received between Mar 25, 2021 and Sep 3, 2026. In those reports, the most common specific device problems are “Incorrect, Inadequate or Imprecise Result or Readings” (521), “False Positive Result” (3) and “Device Markings/Labelling Problem” (1). Reports are not verified by the FDA and do not prove the device caused the problem.
Have deaths or injuries been reported with the BD Phoenix™ NMIC/ID-307 449289?
None of the reports are classified as deaths, 0 as injuries and 526 as malfunctions. The classification is chosen by whoever filed the report, and a death report does not mean the device caused the death.
Who makes the BD Phoenix™ NMIC/ID-307 449289, and is it still sold?
The FDA's device database lists Becton, Dickinson And Company as the company responsible for it (brand BD Phoenix™ NMIC/ID-307, model 449289). Its record shows it as in commercial distribution.
Answers are built automatically from the FDA records on this page. They are not medical advice.
Other SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION devices
- BD Phoenix™ M50 Automated Microbiology System 443624Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION3 recalls411 reports
- BD Phoenix™ PMIC/ID-107 448607Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION2 recalls169 reports
- BD Phoenix™ NMIC-306 449292Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION1 recall143 reports
- VITEK® 2 AST-GP67Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls83 reports
- BD Phoenix™ AST Broth 246003Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION182 reports
- VITEK® 2 AST-GP75Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls52 reports
- VITEK® 2 AST-GP78Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls41 reports
- VITEK® 2 AST-GN95Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls2 reports
Sources and updates
- Device record: FDA GUDID, file dated Nov 27, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Nov 27, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Nov 27, 2026
- Clearance and classification: FDA 510(k), PMA and product classification databases, files dated Nov 27, 2026
- Page last rebuilt: Nov 27, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page