VIDAS® RUB IgG 30226Recalls & reported problems
Biomerieux SA · VIDAS ® RUB IgG (RBG) assay is a quantitative IA intended for use on VIDAS family instrument in the determination of IgG antibodies to rubella virus in human serum. · Class 2
Data updated Oct 16, 2026 from official FDA records · Editorial policy
2 open recalls name this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 0
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 2
Summary
No MAUDE adverse event reports in our data reference this device identifier directly.
2 FDA recall records name this device's identifier, 2 still open.
3 other recalls cover devices cleared under the same FDA submission (K141133); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Recalls naming this device
- Nov 9, 2022Class IIOpen, Classified
As the temperature and time out of range was exceeded, product performances cannot be guaranteed
Z-0797-2023 · Root cause: Storage · Recall details · FDA record ↗
- Sep 22, 2021Class IIOpen, Classified
bioMerieux has been receiving an increasing number of complaints linked to a VIDAS substrate error. It prevents the test from being run, therefore leads to potential delayed results as the user must run another test.
Z-0432-2022 · Root cause: Nonconforming Material/Component · Recall details · FDA record ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K141133. One submission often covers many models, so check the product and lot details in each record.
- Z-0864-2023 · Nov 9, 2022 · Open, ClassifiedVIDAS CLINICAL VIDAS HCG 60 TESTS, CATALOG 30405-01
- Z-0427-2022 · Sep 22, 2021 · Open, ClassifiedVIDAS DEX@, Dimer Exclusion II, REF 30455-01
- Z-0928-2020 · Oct 22, 2019 · TerminatedBIOMERIEUX VIDAS 3, Compact immunoanalyzer with full traceability and automation
Other Enzyme Linked Immunoabsorbent Assay, Rubella devices
- Elecsys Rubella IgG 07027770190Roche Diagnostics GmbH · ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA21 reports
- ADVIA Centaur® Rub G 10310283Siemens Healthcare Diagnostics Inc. · ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA16 reports
- Atellica IM Rub G 10995670Siemens Healthcare Diagnostics Inc. · ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA4 reports
- Elecsys Rubella IgM 07027796190Roche Diagnostics GmbH · ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA4 reports
- Atellica IM Rub M 10995668Siemens Healthcare Diagnostics Inc. · ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA1 report
- Elecsys Rubella IgG Immunoassay 4618793160Roche Diagnostics GmbH · ENZYME LINKED IMMUNOABSORBENT ASSAY, RUBELLA1 report
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page