Skip to content

ETEST® Benzylpenicillin (PG) (UDI 03573026549961)Recalls & reported problems

Biomerieux SA · Etest is a quantitative technique for determining susceptibility of clinically significant bacteria. It provides a gradient of 15 doubling dilutions for MIC determination. · Class 2

All models of ETEST Benzylpenicillin (PG) (3)

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026549961), received between Mar 27, 2020 and Mar 27, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K981135); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • Incorrect, Inadequate or Imprecise Result or Readings1

Patient problems

  • No Patient Involvement1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 27, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings

    A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMERIEUX OF OUT-OF-RANGE HIGH BENZYLPENICILLIN RESULTS FOR ENTEROCOCCUS FAECALIS ATCC 29212 WHEN PERFORMING INTERNAL QUALITY CONTROL TESTING WITH THE ETEST BENZYLPENICIL PG 32 US S30 (REF. (B)(4), LOT 1007645640). THE CUSTOMER OBTAINED A MIC OF 32 MG/L (RESISTANT) AND THE EXPECTED RANGE FOR ENTEROCOCCUS FAECALIS ATCC 29212 IS 1 - 4 MG/L (SUSCEPTIBLE). AS THERE IS NO PATIENT ASSOCIATED WITH THIS QUALITY CONTROL STRAIN, THERE IS NO ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. A BIOMERIEUX INTERNAL INVESTIGATION HAS BEEN INITIATED.

    Patient problems: No Patient Involvement

    MAUDE report 9890707 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K981135. One submission often covers many models, so check the product and lot details in each record.

  • Z-0820-2023 · Nov 9, 2022 · Open, Classified
    ETEST CLINICAL BENZYLPENICIL PG 256 US S30, CATALOG 412262
  • Z-1527-2017 · Dec 15, 2016 · Terminated
    ETEST Benzyl Penicillin PG256 FOAM packaging, Product Name: Etest Benzylpenicillin PG 0.016-256 (high). ETEST is a quantitative technique for determining the antimicrobial suscepti

Other Manual Antimicrobial Susceptibility Test Systems devices

All Manual Antimicrobial Susceptibility Test Systems devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page