ETEST® Benzylpenicillin (PG) (UDI 03573026549961)Recalls & reported problems
Biomerieux SA · Etest is a quantitative technique for determining susceptibility of clinically significant bacteria. It provides a gradient of 15 doubling dilutions for MIC determination. · Class 2
All models of ETEST Benzylpenicillin (PG) (3)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026549961), received between Mar 27, 2020 and Mar 27, 2020 — classified as 1 malfunction.
The most frequently reported specific device problem is “Incorrect, Inadequate or Imprecise Result or Readings” (1 report).
2 other recalls cover devices cleared under the same FDA submission (K981135); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Incorrect, Inadequate or Imprecise Result or Readings1
Patient problems
- No Patient Involvement1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 27, 2020Incorrect, Inadequate or Imprecise Result or Readings; Incorrect, Inadequate or Imprecise Result or Readings
A CUSTOMER IN THE UNITED STATES NOTIFIED BIOMERIEUX OF OUT-OF-RANGE HIGH BENZYLPENICILLIN RESULTS FOR ENTEROCOCCUS FAECALIS ATCC 29212 WHEN PERFORMING INTERNAL QUALITY CONTROL TESTING WITH THE ETEST BENZYLPENICIL PG 32 US S30 (REF. (B)(4), LOT 1007645640). THE CUSTOMER OBTAINED A MIC OF 32 MG/L (RESISTANT) AND THE EXPECTED RANGE FOR ENTEROCOCCUS FAECALIS ATCC 29212 IS 1 - 4 MG/L (SUSCEPTIBLE). AS THERE IS NO PATIENT ASSOCIATED WITH THIS QUALITY CONTROL STRAIN, THERE IS NO ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. A BIOMERIEUX INTERNAL INVESTIGATION HAS BEEN INITIATED.
Patient problems: No Patient Involvement
MAUDE report 9890707 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K981135. One submission often covers many models, so check the product and lot details in each record.
- Z-0820-2023 · Nov 9, 2022 · Open, ClassifiedETEST CLINICAL BENZYLPENICIL PG 256 US S30, CATALOG 412262
- Z-1527-2017 · Dec 15, 2016 · TerminatedETEST Benzyl Penicillin PG256 FOAM packaging, Product Name: Etest Benzylpenicillin PG 0.016-256 (high). ETEST is a quantitative technique for determining the antimicrobial suscepti
Other Manual Antimicrobial Susceptibility Test Systems devices
- ETEST® Imipenem/Relebactam (IPR)Biomerieux SA · Manual antimicrobial susceptibility test systems1 recall4 reports
- ETEST® Amikacin (AK)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall2 reports
- ETEST® Tobramycin (TM)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall1 report
- NA CP112-10Remel, Inc. · Manual Antimicrobial Susceptibility Test Systems1 recall0 reports
- ETEST® Vancomycin (VA)Biomerieux SA · Manual antimicrobial susceptibility test systems1 recall0 reports
- ETEST® Vancomycin (VA)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall0 reports
- ETEST® Erythromycin (EM)Biomerieux SA · Manual Antimicrobial Susceptibility Test Systems1 recall0 reports
- ETEST® Meropenem/Vaborbactam (MEV)Biomerieux SA · Manual antimicrobial susceptibility test systems1 recall0 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page