Skip to content

ID color CatalaseRecalls & reported problems

Biomerieux SA · Manual reagent for catalase test. API®strips give accurate identifications based on extensive databases and are standardized, easy-to-use test systems. · Class 1

Data updated Oct 18, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03573026584139), received between Oct 14, 2020 and Oct 14, 2020 — classified as 2 malfunctions.

The most frequently reported specific device problem is “False Positive Result” (2 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.

Most reported problems

Device problems

  • False Positive Result2

Patient problems

  • No Known Impact Or Consequence To Patient2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionOct 14, 2020False Positive Result; False Positive Result

    A CUSTOMER FROM GERMANY NOTIFIED BIOMÉRIEUX OF FALSE POSITIVE RESULTS IN ASSOCIATION WITH ID COLOR CATALASE (REF. 55561, LOT 1007568120). THE CUSTOMER TESTED ATCC® 29212¿ ENTEROCOCCUS FAECALIS AND AN UNKNOWN NUMBER OF PATIENT ENTEROCOCCUS FAECALIS AND STREPTOCOCCI STRAINS WITH LOT 1007568120. THE NUMBER OF PATIENT STRAINS IMPACTED WAS NOT PROVIDED. THE CUSTOMER STATED THE FALSE POSITIVE CATALASE RESULT CAUSED THE INCORRECT VITEK® 2 CARDS TO BE USED FOR ANTIBIOTIC SUSCEPTIBILITY TESTING (AST) LEADING TO A DELAY IN OBTAINING AST RESULTS. THE CUSTOMER ALSO TESTED THE ISOLATES USING LOT 1007205060; FALSE POSITIVE RESULTS WERE ALSO OBTAINED. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE FALSE POSITIVE RESULT LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. BIOMÉRIEUX HAS INITIATED AN INTERNAL INVESTIGATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10678878 ↗
  • MalfunctionOct 14, 2020False Positive Result; False Positive Result

    A CUSTOMER FROM (B)(6) NOTIFIED BIOMÉRIEUX OF FALSE POSITIVE RESULTS IN ASSOCIATION WITH ID COLOR CATALASE (REF. 55561, LOT: 1007205060). THE CUSTOMER TESTED ATCC® 29212¿ ENTEROCOCCUS FAECALIS AND AN UNKNOWN NUMBER OF PATIENT ENTEROCOCCUS FAECALIS AND STREPTOCOCCI STRAINS WITH LOT:1007205060. THE NUMBER OF PATIENT STRAINS IMPACTED WAS NOT PROVIDED. THE CUSTOMER STATED THE FALSE POSITIVE CATALASE RESULT CAUSED THE INCORRECT VITEK® 2 CARDS TO BE USED FOR ANTIBIOTIC SUSCEPTIBILITY TESTING (AST) LEADING TO A DELAY IN OBTAINING AST RESULTS. THE CUSTOMER ALSO TESTED THE ISOLATES USING LOT 1007568120; FALSE POSITIVE RESULTS WERE ALSO OBTAINED. THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THE FALSE POSITIVE RESULT LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH. BIOMÉRIEUX HAS INITIATED AN INTERNAL INVESTIGATION.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10678758 ↗

Other DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION devices

All DISCS, STRIPS AND REAGENTS, MICROORGANISM DIFFERENTIATION devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page