VITEK® 2 Software 09.02 9.02.0Recalls & reported problems
Biomerieux , Inc. · CD with VITEK® 2 09.02 software version · Class 2
Data updated Oct 18, 2026 from official FDA records · Editorial policy
1 open recall names this device. See the recall details and contact the manufacturer or your healthcare provider.
- Adverse event reports
- 1
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 1
Summary
The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03573026609399), received between Jan 15, 2021 and Jan 15, 2021 — classified as 1 malfunction.
The most frequently reported specific device problem is “Computer Software Problem” (1 report).
1 FDA recall record name this device's identifier, 1 still open.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Recalls naming this device
- Jul 13, 2022Class IIOpen, Classified
There are 7 reported software anomalies that may affect use of the device.
Z-1640-2022 · Root cause: Software design · Recall details · FDA record ↗
Most reported problems
Device problems
- Computer Software Problem1
Patient problems
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 15, 2021Computer Software Problem
A CUSTOMER IN (B)(6) NOTIFIED BIOMÉRIEUX OF A SOFTWARE ISSUE WHILE USING THE VITEK® 2 SYSTEM (REFERENCE # 1950204-2021-00010). THE CUSTOMER UPDATED THEIR SYSTEM FROM (B)(4) IN (B)(6) OF 2020. HOWEVER, IT IS NOW RUNNING A (B)(4) DB ON A (B)(4). THERE WAS NO NOTIFICATION, ANOMALY OR ERROR FROM THE SYSTEM ALERTING THE USER OF THIS CHANGE. PER GLOBAL CUSTOMER SERVICE (GCS), "THIS COMPLAINT WAS CODED PRE DUE TO THE VITEK2 SOFTWARE BACKUP/RESTORE APPLICATION AT (B)(4), ALLOWING A (B)(4) DB TO BE RESTORED. BY DESIGN THE FUNCTION SHOULD HAVE PREVENTED THIS FROM OCCURRING AS THE (B)(4) VERSION WAS NOT A ¿DOU¿ (DATA ONLY UPDATE). IN THE CONFIGURATION SETTINGS IN THE VITEK 2 SOFTWARE, THE VERSION (B)(4) (UDI CODE CONFIRMS) IS DISPLAYED AND IS OPERATING WITH KNOWLEDGE BASE VERSION (B)(4). THE DISPARITY OF VERSIONS COULD RESULT IN THE SYSTEM ANALYSIS USING AN OLDER VERSION OF BREAKPOINTS, AND/OR LIMITATIONS, AND NEGLECT IMPLEMENTATION OF IMPROVEMENTS EXPECTED IN THE (B)(4) KNOWLEDGE BASE."THERE IS NO INDICATION OR REPORT FROM THE LABORATORY THAT THIS SOFTWARE ISSUE HAS LED TO ANY ADVERSE EVENT RELATED TO ANY PATIENT'S STATE OF HEALTH.A BIOMÉRIEUX INTERNAL INVESTIGATION HAS BEEN INITIATED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 11181301 ↗
Other SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION devices
- BD Phoenix™ M50 Automated Microbiology System 443624Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION3 recalls417 reports
- BD Phoenix™ NMIC/ID-307 449289Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION529 reports
- BD Phoenix™ PMIC/ID-107 448607Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION2 recalls169 reports
- BD Phoenix™ NMIC-306 449292Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION1 recall143 reports
- VITEK® 2 AST-GP67Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls83 reports
- BD Phoenix™ AST Broth 246003Becton, Dickinson And Company · SYSTEM, TEST, AUTOMATED, ANTIMICROBIAL SUSCEPTIBILITY, SHORT INCUBATION182 reports
- VITEK® 2 AST-GP75Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls52 reports
- VITEK® 2 AST-GP78Biomerieux , Inc. · System, Test, Automated, Antimicrobial Susceptibility, Short Incubation2 recalls41 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026 · each recall links to the FDA recall notice
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page