Skip to content

NA 011021Recalls & reported problems

Smith & Nephew, Inc. · GRSPR ALGTR UPBTR · Class 1

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
5
Deaths reported
0
Injuries reported
1
Recalls naming it
0

Summary

The FDA's MAUDE database contains 5 adverse event reports that list this exact device (DI 03596010022998), received between Mar 25, 2019 and Mar 5, 2026 — classified as 1 injury, 4 malfunctions.

The most frequently reported specific device problem is “Break” (5 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break5
  • Material Separation1

Patient problems

  • No Clinical Signs, Symptoms or Conditions2
  • No Known Impact Or Consequence To Patient2
  • No Consequences Or Impact To Patient1
  • No Information1

Recent adverse event reports

The 5 most recent of 5 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 5, 2026Break

    IT WAS REPORTED THAT, DURING A MENISCUS SURGERY, THE UPPER CLAW OF AN ALLIGATOR GRASPER BROKE INSIDE THE PATIENT. THE DEVICE WAS REMOVED BY SUCTION. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT'S STATUS IS OKAY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24518607 ↗
  • InjurySep 24, 2020Break; Material Separation; Break; Material Separation

    IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE BODY OF THE GRASPER BROKE APART. IT IS UNKNOWN IF A BACKUP DEVICE WAS AVAILABLE AND IF A DELAY HAPPENED IN THE PROCEDURE. HOWEVER, NO PATIENT COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Information

    MAUDE report 10583568 ↗
  • MalfunctionMay 7, 2019Break

    IT WAS REPORTED THAT DURING A PROCEDURE, THE DEVICE BROKE DOWN. BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO DELAY NOR PATIENT INJURIES WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8588462 ↗
  • MalfunctionMar 25, 2019Break

    IT WAS REPORTED THAT DURING A MENISCUS RESECTION, THE MOBILE PART OF THE WORKING PART OF THE DEVICE BROKE OFF. THE PIECE REMAINED IN THE KNEE JOINT. THERE WAS A DELAY OF MORE THAN 60 MINUTES. THE DOCTOR DID NOT HAVE A BACKUP DEVICE AVAILABLE, SO AN ARTHROTOMY WAS EXECUTED. THE EXTRACTION OF THE FRAGMENT WAS NOT CARRIED OUT, THE PATIENT IS DOING WELL.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8448302 ↗
  • MalfunctionMar 25, 2019Break

    IT WAS REPORTED THAT DURING THE OPERATION THE CLAMP BROKE. THE OPERATION COMPLETED SUCCESSFULLY. THERE WAS NO HARM TO THE PATIENT. IT IS UNKNOWN IF THERE WAS A BACK UP DEVICE AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8446977 ↗

Other ACCESSORIES,ARTHROSCOPIC devices

All ACCESSORIES,ARTHROSCOPIC devices →

Sources and updates