Adverse event reports
8
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010033192), received between Apr 22, 2021 and Apr 29, 2024 — classified as 8 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 1 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fracture2
  • Incomplete or Inadequate Connection1
  • Loose or Intermittent Connection1
  • Mechanical Jam1

Patient problems

  • No Clinical Signs, Symptoms or Conditions8

Recent adverse event reports

The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 29, 2024Break; Break

    IT WAS REPORTED THAT, DURING A CHS PROCEDURE, THE LOCKING MECHANISM FOR THE DEPTH STOP OF ONE (1) POWER COMBINATION REAMER BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19202556 ↗
  • MalfunctionOct 10, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING INTERNAL FIXATION SURGERY, THE TIP AND PART OF THE HANDLE OF A POWER COMBINATION REAMER BROKE OFF. SURGERY WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, WITH THE SAME DEVICE. NO INJURIES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17905572 ↗
  • MalfunctionApr 18, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING SURGERY, THE CONNECTION TO THE DRILL OF A POWER COMBINATION REAMER SNAPPED OFF DUE TO WEAR AND TEAR. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16764578 ↗
  • MalfunctionMay 5, 2022Break; Break

    IT WAS REPORTED THAT A POWER COMBINATION REAMER BROKEN OFF FROM SHAFT. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14285768 ↗
  • MalfunctionNov 17, 2021Break; Break

    IT WAS REPORTED THAT, DURING SET UP FOR AN INTERNAL FIXATION SURGERY, A POWER COMBINATION REAMER WAS FOUND BROKEN. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE INSTEAD. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12833439 ↗
  • MalfunctionJun 1, 2021Mechanical Jam; Mechanical Jam

    IT WAS REPORTED THAT, A POWER COMBINATION REAMER IS JAMMED ON TO AN OTHER COMPANY REAMER. AS THIS WAS NOTICED AFTER PROCEDURE, NO DELAY WAS REPORTED AND PATIENT WAS NOT IMPACTED BY THIS EVENT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11916864 ↗
  • MalfunctionMay 6, 2021Incomplete or Inadequate Connection; Incomplete or Inadequate Connection

    IT WAS REPORTED THAT, DURING A FEMORAL FRACTURE SURGERY, THE REAMER CHANNEL OF A POWER COMBINATION REAMER DID NOT ALLOW THE SMALL DIAMETER REAMER TO THE PASS THROUGH. NO SIGNIFICANT SURGICAL DELAY WAS REPORTED AS A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NO PATIENT INJURY WAS REPORTED. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11784132 ↗
  • MalfunctionApr 22, 2021Loose or Intermittent Connection; Loose or Intermittent Connection

    IT WAS REPORTED THAT UPON INSPECTION WAS FOUND THAT A POWER COMBINATION REAMER (CASE-(B)(4)) DOES NOT TIGHTEN PROPERLY AND A COMBINATION DRILL GUIDE 3.5M (CASE-(B)(4)) IS STUCK. NO CASE WAS INVOLVED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11711042 ↗