Skip to content

NA 71364008Recalls & reported problems

Smith & Nephew, Inc. · STEM INSERTER-FRAME · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
12
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 12 adverse event reports that list this exact device (DI 03596010199249), received between Aug 12, 2019 and Jul 15, 2026 — classified as 2 injuries, 10 malfunctions.

The most frequently reported specific device problem is “Break” (7 reports).

2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 1 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break7
  • Fracture2
  • Material Deformation1
  • Material Fragmentation1
  • Unstable1

Patient problems

  • No Clinical Signs, Symptoms or Conditions6
  • Foreign Body In Patient2
  • No Patient Involvement2
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 12 most recent of 12 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJul 15, 2026Fracture

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25821937 ↗
  • InjuryFeb 28, 2022Break; Break

    IT WAS REPORTED THAT, AFTER THR SURGERY, POST-OPERATIVE X RAY REVEALED A FOREIGN 7MM METAL OBJECT AROUND THE LESSER TROCHANTER OF THE PATIENT. AFTER REVIEWING THE SALES REP INSPECTED THE INSTRUMENT AND FOUND THAT THE TOOTH WAS MISSING FROM THE STEM INSERTER FRAME THAT IS PRESUMED THE FOREIGN OBJECT. A REVISION IS NOT SCHEDULED AT THIS TIME.

    Patient problems: Foreign Body In Patient

    MAUDE report 13620438 ↗
  • InjuryJan 25, 2022Break; Break

    IT WAS REPORTED THAT, DURING TOTAL HIP REPLACEMENT, A STEM INSERTER FRAME BROKE WHILE SEATING THE FINAL IMPLANT. THE BROKEN PIECE WAS RETRIEVED BY HAND FROM THE PATIENT. SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH A SMITH & NEPHEW BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13353197 ↗
  • MalfunctionSep 9, 2021Break; Break

    IT WAS REPORTED THAT, DURING SET UP, BEFORE A THR SURGERY, ONE STEM INSERTER FRAME BROKE OFF. THIS HAPPENED BEFORE USE IN PATIENT. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A SMITH AND NEPHEW BACK-UP DEVICE INSTEAD. NO HEALTH CONSEQUENCES WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12450547 ↗
  • MalfunctionJun 1, 2021Unstable; Unstable

    IT WAS REPORTED THAT, DURING THR SURGERY, A STEM IMPACTOR ROD (CASE-(B)(4)) AND A STEM INSERTER FRAME (CASE-(B)(4) WERE NOTICED TO BE LOOSE UPON IMPACTION, AND COULD NOT BE TIGHTENED ENOUGH TO GET FIRM FIXATION BETWEEN THEM. THERE WAS A SURGICAL DELAY LESS THAN 30 MINUTES REPORTED DUE TO THIS EVENT, AND THE PROCEDURE WAS CONCLUDED USING SMITH AND NEPHEW BACKUP DEVICES. NEITHER PATIENT INJURY NOR OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11914395 ↗
  • MalfunctionApr 21, 2021Material Deformation; Material Deformation

    IT WAS REPORTED THAT THE THREADS OF ONE STEM INSERTER FRAME AND ONE STEM IMPACTOR ROD ARE STRIPPED. IT IS UNKNOWN IF THIS WAS NOTICED DURING SURGERY OR NOT; THEREFORE, PATIENT INVOLVEMENT IS NOT CONFIRMED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11704317 ↗
  • MalfunctionApr 21, 2021Break; Break

    IT WAS REPORTED THAT THE STEM INSERTER FRAME IS BROKEN. NO CASE INVOLVED THEREFORE NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11704298 ↗
  • MalfunctionJul 15, 2020Material Fragmentation; Material Fragmentation

    IT WAS REPORTED THAT DURING A ROUTINE EFIP INSPECTION, THE STEM INSERTER FRAME WAS FOUND TO HAVE ONE OF THE TIPS BROKEN OFF. THERE WAS NO CASE INVOLVED.

    Patient problems: No Patient Involvement

    MAUDE report 10280219 ↗
  • MalfunctionMay 15, 2020Fracture; Fracture

    IT WAS REPORTED THAT, DURING AN INSTRUMENT INSPECTION, IT WAS NOTICED THAT THE END OF THE STEM INSERTER FRAME WAS BROKEN. NO CASE INVOLVED.

    Patient problems: No Patient Involvement

    MAUDE report 10063420 ↗
  • MalfunctionOct 14, 2019Break; Break

    IT WAS REPORTED THAT ONE OF THE 2 PIECES THAT ALLOWS THE SYNERGY FEMORAL STEM IMPLANT TO SEAT ON THE INSERTER FRAME HAS BROKEN OFF. NO DELAY REPORTED. NO PATIENT INJURIES REPORTED. AN S&N BACKUP WAS AVAILABLE.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 9187051 ↗
  • MalfunctionSep 19, 2019Break; Break

    IT WAS REPORTED THAT DURING SURGERY, WHILE IMPACTING, THE TIP OF THE DEVICE BROKE. PIECES REMAIN ON THE PATIENT. DELAY FROM FIFTEEN MINUTES REPORTED. NO BACKUP DEVICE WAS REQUIRED. NO IMPACT OR INJURY TO PATIENT REPORTED.

    Patient problems: Foreign Body In Patient

    MAUDE report 9088147 ↗
  • MalfunctionAug 12, 2019Break; Break

    IT WAS REPORTED THAT DURING SURGERY, INSTRUMENT BROKE WHILE BEING USED. NO INJURY REPORTED. NO DELAY AND PROCEDURE WAS CONCLUDED WITH A BACKUP FROM SMITH AND NEPHEW.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8884022 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.

  • Z-0730-2019 · Jan 7, 2019 · Terminated
    SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
  • Z-1008-2017 · Nov 16, 2016 · Terminated
    REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

Other Prosthesis, hip, femoral component, cemented, metal devices

All Prosthesis, hip, femoral component, cemented, metal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page