Skip to content

REFLECTION 71362279Recalls & reported problems

Smith & Nephew, Inc. · REFLECTION REAMER HANDLE · Class 2

Quick summary

This Smith & Nephew Reflection reamer handle is a surgical tool used to prepare the hip socket during hip replacement. Its 155 reports mostly describe the handle breaking or snapping during surgery, and many say the procedure continued with a similar backup tool. The reports include no deaths and 11 injuries, and 144 are malfunctions. No recalls name this device.

AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.

Adverse event reports
155
Deaths reported
0
Injuries reported
11
Recalls naming it
0

Summary

The FDA's MAUDE database contains 155 adverse event reports that list this exact device (DI 03596010456519), received between Feb 22, 2019 and Jun 22, 2026 — classified as 11 injuries, 144 malfunctions.

The most frequently reported specific device problem is “Break” (94 reports).

2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 1 reportsDec 2023: 2 reportsJan 2024: 1 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 1 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 1 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 1 reportsJun 2025: 0 reportsJul 2025: 2 reportsAug 2025: 0 reportsSep 2025: 1 reportsOct 2025: 0 reportsNov 2025: 1 reportsDec 2025: 1 reportsJan 2026: 1 reportsFeb 2026: 1 reportsMar 2026: 2 reportsApr 2026: 1 reportsMay 2026: 1 reportsJun 2026: 1 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520262
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break94
  • Fracture24
  • Detachment of Device or Device Component11
  • Connection Problem10
  • Incomplete or Inadequate Connection5
  • Material Deformation5
  • Component Missing3
  • Defective Device3

Patient problems

  • No Clinical Signs, Symptoms or Conditions124
  • No Consequences Or Impact To Patient12
  • No Known Impact Or Consequence To Patient10
  • No Patient Involvement10
  • Failure of Implant1
  • Injury1
  • Laceration(s)1
  • No Information1

Recent adverse event reports

The 25 most recent of 155 reports, in the reporter's own words as published by the FDA.

  • MalfunctionJun 22, 2026Fracture

    IT WAS REPORTED THAT, DURING A THA SURGERY, ONE (1) REAMER HANDLE BROKE AND PART DISASSOCIATED WHEN REAMING. NO PIECES FELL INTO THE PATIENT'S ANATOMY. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25573118 ↗
  • MalfunctionMay 11, 2026Fracture

    IT WAS REPORTED THAT DURING A THR, THE SHAFT OF A REAMER HANDLE BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25136347 ↗
  • MalfunctionApr 3, 2026Break

    IT WAS REPORTED THAT, DURING A THA, AN REAMER HANDLE SNAPPED AT THE ADAPTOR/HANDLE JUNCTION. NO PIECES WERE LOST. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24776555 ↗
  • MalfunctionMar 30, 2026Break

    IT WAS REPORTED THAT DURING A THR SURGERY, THE REAMER HANDLE BROKE UPON REAMING. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24724910 ↗
  • MalfunctionMar 18, 2026Break

    IT WAS REPORTED THAT, DURING A THA SURGERY, THE TIP OF THE REAMER HANDLE WAS BROKEN WHILE REAMING SOCKET. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24626494 ↗
  • MalfunctionFeb 8, 2026Break; Fracture; Material Deformation

    IT WAS REPORTED THAT DURING A THR SURGERY ONE (1) REAMER HANDLE BROKE UPON IMPACTION. THE PROCEDURE WAS COMPLETED, AFTER NO DELAY, WITH A SIMILAR S+N BACK UP PRODUCT. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24288434 ↗
  • MalfunctionJan 15, 2026Break

    IT WAS REPORTED THAT, DURING A THA SURGERY, EIGHT REAMER HANDLE ARE BROKEN DUE TO USE OVER TIME. IT IS UNKNOWN THE SECTION ON THE DEVICE FOR THE REPORTED DAMAGE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR S+N BACK-UP PRODUCT. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24086677 ↗
  • MalfunctionDec 1, 2025Fracture; Fracture

    IT WAS REPORTED THAT DURING A THR THE CONNECTION PIECE AT THE END OF A REAMER HANDLE BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23685507 ↗
  • MalfunctionNov 16, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING A THA SURGERY, ONE (1) REAMER HANDLE WAS BENT AND ANOTHER (1) REAMER HANDLE WAS MISSING THE LOCKING COLLAR. IT IS UNKNOWN HOW PROCEDURE WAS COMPLETED OR IF THIS CAUSED ANY DELAY. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23565745 ↗
  • MalfunctionSep 24, 2025Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING A THR SURGERY, IT WAS NOTICED THAT THE REAMER HEAD FASTENER OF ONE (1) REAMER HANDLE CAME APART DURING REAMING. ALL PARTS WERE RECOVERED. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY USING A S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 23130268 ↗
  • MalfunctionJul 21, 2025Fracture; Material Deformation; Fracture; Material Deformation

    IT WAS REPORTED THAT, DURING THR SURGERY, ONE (1) REAMER HANDLE WAS BROKEN. IT BROKE AT THE END THAT GOES INTO THE POWER ADAPTER/CHUCK. NOTHING FELL INTO THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22560617 ↗
  • MalfunctionJul 2, 2025Fracture; Fracture

    IT WAS REPORTED THAT, DURING A BIRMINGHAM HIP RESURFACING SURGERY, ONE (1) REAMER HANDLE BROKE NEAR THE CONNECTION POINT TO THE REAMER DOME AND COULD NO LONGER BE ATTACHED TO IT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 22391623 ↗
  • MalfunctionMay 9, 2025Break; Material Deformation; Break; Material Deformation

    IT WAS REPORTED THAT, DURING THR, ONE (1) REAMER HANDLE BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21996790 ↗
  • MalfunctionMar 25, 2025Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, THE CONNECTING PIECE OF ONE (1) REAMER HANDLE BROKE. THE PROCEDURE WAS RESUMED WITHOUT DELAY USING A SMITH AND NEPHEW BACKUP DEVICE, AND NO INJURIES WERE REPORTED AS A RESULT.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21693866 ↗
  • MalfunctionSep 26, 2024Unintended Movement; Unintended Movement

    IT WAS REPORTED THAT, DURING A THR, THE REAMER HANDLE WAS ROTATING OUTSIDE OF ITS ORIGINAL AXIS. THE INTENDED ACETABULAR DIAMETER WAS DAMAGED A LITTLE BEYOND THE DIAMETER DESCRIBED ON THE REAMER DOME. THE PLAN WAS GRADUALLY ADJUSTED AS THEY INCREASED THE DIAMETER OF THE REAMERS (CHANGING TO THE LARGER REAMER), WITHOUT AFFECTING THE PATIENT. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: Laceration(s)

    MAUDE report 20316070 ↗
  • MalfunctionJun 23, 2024Fracture; Fracture

    IT WAS REPORTED THAT, WHILE REAMING IN A THR SURGERY, THE PROXIMAL END TO THE DRILL OF ONE (1) REAMER HANDLE BROKE OFF. IT BROKE INTO TWO PIECES OF WHICH NONE GOT INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19591843 ↗
  • MalfunctionJan 15, 2024Fracture; Fracture

    IT WAS REPORTED THAT, DURING THR SURGERY, THE SHAFT OF A REAMER HANDLE BROKE NEAR THE QUICK CONNECT INSERTION END. IT HAPPENED WHILE THE CUP WAS BEING REAMED, SO THE REAMER AND REAMER SHAFT CONNECTOR WERE IN THE PATIENT AND THE END CONNECTED TO THE DRILL WERE OUT OF THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18511174 ↗
  • MalfunctionDec 27, 2023Fracture; Material Twisted/Bent; Fracture; Material Twisted/Bent

    IT WAS REPORTED THAT, DURING SURGERY, TWO (1) REAMER HANDLE WAS NOTICED TO BE BENT THROUGH REGULAR USE OF PRODUCT. REPLACEMENT REQUESTED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM. AS PER INVESTIGATION RESULTS, IT WAS FOUND THAT THE TIP OF THE REAMER HANDLE DEVICE IS FRACTURED OFF.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18398490 ↗
  • MalfunctionDec 26, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, A REAMER HANDLE BROKE WHILE REAMING. NO PIECES WERE LEFT BEHIND, ALL WERE MANUALLY REMOVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18392438 ↗
  • MalfunctionNov 27, 2023Fracture; Material Deformation; Fracture; Material Deformation

    IT WAS REPORTED THAT, DURING THR SURGERY, A REAMER HANDLE BROKE WHILE USING THE INSTRUMENT IN THE PATIENT. THE PROCEDURE WAS RESUMED, AFTER A SIGNIFICANT DELAY GREATER THAN 30 MIN, WITH THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 18212752 ↗
  • MalfunctionOct 17, 2023Break; Break

    IT WAS REPORTED THAT, DURING THR SURGERY, ONE (1) REAMER HANDLE BROKE WHERE IT MEETS DRILL ATTACHMENT. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17954460 ↗
  • MalfunctionSep 25, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, THE TIP OF A REAMER HANDLE BROKE OFF, WHEN REMOVING IT FROM THE DRILL. NO PIECES FELL INSIDE THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17810200 ↗
  • MalfunctionAug 15, 2023Break; Break

    IT WAS REPORTED THAT DURING THR SURGERY, ONE (X1) REAMER HANDLE BROKE WHILE REAMING THE ACETABULUM, NO PIECES OF THE BROKEN INSTRUMENT WERE LEFT IN THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 17553558 ↗
  • MalfunctionMar 21, 2023Fracture; Fracture

    IT WAS REPORTED THAT, DURING A TOTAL RIGHT HIP REPLACEMENT, WHILE PROCEEDING TO PERFORM THE ACETABULAR REAMING, THE HANDLE OF THE REAMER HANDLE SPLIT WHERE IT ATTACHES TO THE ENGINE. THE DEVICE BROKE INSIDE THE PATIENT, BUT THE PARTS WERE RECOVERED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. THE PATIENT'S HEALTH CONDITION IS STABLE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 16587723 ↗
  • MalfunctionDec 9, 2022Fracture; Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, IN THE ACT OF REAMING THE ACETABULUM, THE METAL STEM FROM THE REAMER HANDLE BROKE IN THE HUDSON ADAPTER OF THE POWER TOOL. THE ITEM WAS REMOVED FROM THE FIELD. NOTHING FELL INTO THE WOUND AND ALL PARTS WERE RECOVERED AFTER THE PROCEDURE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15948840 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.

  • Z-0730-2019 · Jan 7, 2019 · Terminated
    SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
  • Z-1008-2017 · Nov 16, 2016 · Terminated
    REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

Other Prosthesis, hip, femoral component, cemented, metal devices

All Prosthesis, hip, femoral component, cemented, metal devices →

Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.