SYNERGY 71366414Recalls & reported problems
Smith & Nephew, Inc. · SYNERGY TRIAL NECK SIZE 14-18 · Class 2
All models of SYNERGY (52)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010199904), received between Feb 21, 2021 and Mar 17, 2022 — classified as 9 malfunctions.
The most frequently reported specific device problem is “Break” (6 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Most reported problems
Device problems
- Break6
- Corroded5
- Peeled/Delaminated2
- Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions9
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 17, 2022Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, THE MAGNET OF A SYN TRIAL NECK SZ 14-18 BROKE OFF WHILE SURGEON WAS USING IT. SURGERY WAS FINISHED WITHOUT ANY DELAY, WITH A SN BACK-UP DEVICE. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13792029 ↗MalfunctionFeb 28, 2022Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT, THE MAGNET OF A SYN TRIAL NECK SZ 14-18 APPEARS TO BE PEELING AWAY. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13624796 ↗MalfunctionFeb 28, 2022Peeled/Delaminated; Peeled/Delaminated
IT WAS REPORTED THAT, THE MAGNET A SYN TRIAL NECK SZ 14-18 APPEARS TO BE PEELING AWAY. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13624781 ↗MalfunctionMay 21, 2021Break; Break
IT WAS REPORTED THAT, DURING A THR, A SYN TRIAL NECK SZ 14-18 BROKE. DEVICE OUTSIDE THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT DELAY USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11864072 ↗MalfunctionMar 8, 2021Break; Corroded; Break; Corroded
IT WAS REPORTED THAT DURING SERVICE/PM, MAGNET ON THE SYNERGY TRIAL NECK SIZE 14-18 WAS FOUND TO BE CHIPPED AND CORRODED. THERE WAS NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11431130 ↗MalfunctionFeb 28, 2021Break; Corroded; Break; Corroded
IT WAS REPORTED THAT THE MAGNETS ON THE DEVICE ARE CORRODED/CHIPPED. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11390968 ↗MalfunctionFeb 26, 2021Break; Corroded; Break; Corroded
IT WAS REPORTED THAT, THE MAGNETS OF THE SYN HO TRIAL NECK SZ 14-18 ARE CORRODED, CHIPPED, AND FALLING OUT. INCIDENT OCCURRED DURING SERVICE; THEREFORE, NO PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11386084 ↗MalfunctionFeb 25, 2021Break; Corroded; Break; Corroded
IT WAS REPORTED THAT, DURING PREVENTIVE MAINTENANCE SERVICE, THE MAGNETS OF TWO SYNERGY TRIAL NECK SIZE 8-13 WERE FOUND CHIPPED/CORRODED/FALLING OUT. NO CASE REPORTED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11376948 ↗MalfunctionFeb 21, 2021Corroded; Fracture; Corroded; Fracture
IT WAS REPORTED THAT THE MAGNETS ON THE SYNERGY TRIAL ARE CORRODED/FALLING OUT. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11356765 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page