GENESIS II 71440190Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II FEMORAL IMPACTOR · Class 2
All models of GENESIS II (462)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew GENESIS II femoral impactor is a surgical tool used in joint replacement. Its 143 reports often describe parts breaking, cracking or flaking during knee replacement surgery, and in several cases surgeons switched to a backup tool. Most say the patient was not harmed. There are no deaths and 19 injuries. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 143
- Deaths reported
- 0
- Injuries reported
- 19
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 143 adverse event reports that list this exact device (DI 03596010214379), received between Nov 30, 2017 and May 19, 2026 — classified as 19 injuries, 124 malfunctions.
The most frequently reported specific device problem is “Break” (75 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break75
- Fracture34
- Material Fragmentation18
- Crack14
- Flaked5
- Naturally Worn4
- Appropriate Device Problem Term/Code Not Available2
- Insufficient Device Problem Information2
Patient problems
- No Clinical Signs, Symptoms or Conditions99
- No Consequences Or Impact To Patient16
- No Known Impact Or Consequence To Patient16
- No Patient Involvement7
- Foreign Body In Patient3
- No Information3
- Injury2
- Device Embedded In Tissue or Plaque1
Recent adverse event reports
The 25 most recent of 143 reports, in the reporter's own words as published by the FDA.
MalfunctionMay 19, 2026Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, A GII FEMORAL IMPACTOR BROKE OFF AT THE PAD. NO PIECE FELL IN PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25224763 ↗MalfunctionMay 19, 2026Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, A GII FEMORAL IMPACTOR BROKE OFF AT THE PAD. NO PIECE FELL IN PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 25224756 ↗MalfunctionMar 15, 2026Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, UPON THE FINAL MALLET STRIKE, THE PLASTIC OF ONE (1) GII FEMORAL IMPACTOR SNAPPED APART. ALL PIECES WERE ACCOUNTED FOR. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24601488 ↗MalfunctionNov 3, 2025Flaked; Flaked
IT WAS REPORTED THAT DURING A TKA, THE BLACK PLASTIC OF A GII FEMORAL IMPACTOR WAS NOTICED TO BE FLAKING. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23451008 ↗MalfunctionSep 18, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, ONE (1) GII FEMORAL IMPACTOR FRACTURED; THE PROCEDURE WAS PERFORMED AFTER A NON-SIGNIFICANT SURGICAL DELAY USING A S+N BACKUP DEVICE. THE PATIENT WAS NOT INJURED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23087724 ↗MalfunctionJun 2, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, THE PLASTIC OF ONE (1) GII FEMORAL IMPACTOR BROKE. IT IS UNKNOWN HOW THE PROCEDURE WAS FINISHED. NO DELAY OR INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22125120 ↗MalfunctionApr 1, 2025Fracture; Fracture
IT REPORTED THAT, DURING A TKA SURGERY, ONE (1) GII FEMORAL IMPACTOR BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21749049 ↗MalfunctionDec 10, 2024Degraded; Fracture; Degraded; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, THE BLACK PAD IN A GII FEMORAL IMPACTOR IS WEARING AWAY AND CHIPPING OFF PIECES. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20894267 ↗MalfunctionNov 25, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, PART OF THE GII FEMORAL IMPACTOR BROKE OFF WHEN THE ORIGINAL PROSTHESIS WAS IMPACTED. THE BROKEN PIECE WAS COMPLETELY RECOVERED AND DID NOT END UP IN THE PATIENT. NO PART HAS REMAINED IN THE BODY. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20776291 ↗MalfunctionNov 21, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA PROCEDURE, THE BLACK HEAD PART OF ONE (1) GII FEMORAL IMPACTOR BROKE OFF WHILE IMPACTING THE FEMUR. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20745615 ↗MalfunctionOct 23, 2024Fracture; Insufficient Device Problem Information; Fracture; Insufficient Device Problem Information
IT WAS REPORTED THAT DURING A TKA, SURGEON NOTICED THE GII FEMORAL IMPACTOR WAS DAMAGED. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED. RESULTS OF INVESTIGATION OF THE GII FEMORAL IMPACTOR HAVE CONCLUDED THAT PARTS OF THE PLASTIC FEMORAL IMPACTOR BUMPER ARE FRACTURED OFF.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20517942 ↗MalfunctionOct 15, 2024Break; Break
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) GII FEMORAL IMPACTOR BROKE. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, USING A S+N BACKUP DEVICE. NO INJURY WAS REPORTED AS A RESULT OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20452235 ↗MalfunctionJul 22, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA, A CHUNK AT THE END OF THE GII FEMORAL IMPACTOR BROKE OFF THE PAD WHILE IMPACTING THE FEMORAL COMPONENT. NO PIECES FELL INTO PATIENT. THERE WAS NO DELAY, AND THE PROCEDURE WAS FINISHED USING A SMITH & NEPHEW BACK UP. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19794046 ↗MalfunctionJun 17, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY, IN A GII FEMORAL IMPACTOR, THE BLACK PLASTIC IS COMING OFF. THE PIECES WERE PICKED OUT OF THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19554525 ↗MalfunctionApr 25, 2024Break; Break
IT WAS REPORTED THAT, DURING A TKA SURGERY, THE GII FEMORAL IMPACTOR BROKE WHILE IMPLANTING THE FEMUR. THE SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH A S+N BACK-UP DEVICE. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19183452 ↗MalfunctionApr 2, 2024Break; Break
IT WAS REPORTED THAT DURING A TKA, THE PLASTIC BUMPERS ON THE GII FEMORAL IMPACTOR BROKE. ALL PIECES WERE RECOVERED. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19021363 ↗MalfunctionFeb 8, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA SURGERY ONE (1) GII FEMORAL IMPACTOR WAS WORN AND CHIPPED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18670313 ↗MalfunctionFeb 2, 2024Fracture; Fracture
IT WAS REPORTED THAT ONE (1) GII FEMORAL IMPACTOR HAD THE PS PORTION CHIPPED AWAY AND DIDN'T IMPACT AS IT SHOULD. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18635114 ↗MalfunctionNov 29, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING TKA SURGERY, A GII FEMORAL IMPACTOR BROKE WHILE THE SURGEON WAS IMPLANTING A FEMORAL COMPONENT, THE HEAD COMPLETELY BROKE OFF; NO PIECE(S) FELL INSIDE THE WOUND. THE PROCEDURE WAS RESUMED, AFTER A NON-SIGNIFICANT DELAY, WITH A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18234382 ↗MalfunctionOct 30, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA PROCEDURE, THE TIP OF A GII FEMORAL IMPACTOR BROKE, NO PIECES CAME INTO CONTACT WITH THE PATIENT, AND ALL WERE ACCOUNTED FOR. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18035879 ↗MalfunctionMay 23, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING TKA SURGERY THE BLACK 'BLOCK' PART OF THE GII FEMORAL IMPACTOR WAS FRACTURED IN HALF. NO PIECES FELL INTO THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. THE PATIENT WAS NOT HARMED AS A CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16989162 ↗MalfunctionMay 20, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA, THE IMPACTOR HEAD OF ONE (1) GII FEMORAL IMPACTOR WAS BROKEN WHILE IMPACTING THE FEMORAL COMPONENT. THE PIECE CRACKED IN HALF ON IMPACTION. NOTHING FELL INTO THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16970548 ↗MalfunctionMay 2, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) GII FEMORAL IMPACTOR BROKE DURING FINAL IMPACTION. BACK-UP DEVICE WAS AVAILABLE BUT NOT REQUIRED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16857716 ↗MalfunctionDec 5, 2022Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, THE GII FEMORAL IMPACTOR BROKE AT THE PIN/FIXATION AREA. ALL PIECES WERE RECOVERED FROM THE PATIENT. THE PROCEDURE WAS RESUMED, AFTER A NON SIGNIFICANT DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15921719 ↗MalfunctionNov 3, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING TKA SURGERY, THE PLASTIC PIECE OF THE GII FEMORAL IMPACTOR BROKE WHEN IMPACTING THE FEMUR. IT IS UNKNOWN HOW THE PROCEDURE WAS RESUMED OR IF THERE WAS ANY DELAY NOR INJURY REPORTED. IT IS UNKNOWN IF THERE WAS A S&N BACKUP DEVICE AVAILABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15721307 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.