GENESIS II 71440192Recalls & reported problems
Smith & Nephew, Inc. · GENESIS II TIBIAL BASEPLATE IMPACTOR · Class 2
All models of GENESIS II (462)
Data updated Oct 18, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew GENESIS II tibial baseplate impactor is a surgical tool used in knee replacement surgery. Its 99 adverse event reports include no deaths, 12 injuries and 87 malfunctions. The reports mostly describe the tool breaking, cracking or splintering during surgery, usually with the operation finished using a back-up tool. Many reports say the patient was not harmed. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 99
- Deaths reported
- 0
- Injuries reported
- 12
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 99 adverse event reports that list this exact device (DI 03596010214386), received between Dec 6, 2018 and Mar 20, 2026 — classified as 12 injuries, 87 malfunctions.
The most frequently reported specific device problem is “Break” (56 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 18, 2026.
Reports received per month
Most reported problems
Device problems
- Break56
- Crack16
- Fracture15
- Material Fragmentation8
- Naturally Worn6
- Detachment of Device or Device Component3
- Scratched Material2
- Component Missing1
Patient problems
- No Clinical Signs, Symptoms or Conditions69
- No Consequences Or Impact To Patient10
- No Known Impact Or Consequence To Patient10
- No Patient Involvement6
- Device Embedded In Tissue or Plaque2
- No Information2
- Failure of Implant1
- Insufficient Information1
Recent adverse event reports
The 25 most recent of 99 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 20, 2026Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) GII TIBIAL BASE IMPACTOR WAS DAMAGED. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24649247 ↗MalfunctionMar 20, 2026Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) GII TIBIAL BASE IMPACTOR WAS CRACKED. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24649167 ↗MalfunctionOct 2, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKR SURGERY, ONE (1) FEMORAL TIBIAL SPACER BROKE. ALL PIECES FELL OUTSIDE THE PATIENT AND WERE ACCOUNTED. THE PROCEDURE WAS COMPLETED WITH AN EQUIVALENT S+N BACK UP PRODUCT, AFTER A NON-SIGNIFICANT DELAY. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 23205666 ↗MalfunctionJun 30, 2025Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, AFTER IMPACTION IT WAS NOTICED THAT ONE (1) GII TIBIAL BASE IMPACTOR HAD BROKEN. THE SITE WAS CHECKED TO CONFIRM NO PIECES WERE IN THE WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22355173 ↗MalfunctionMar 17, 2025Break; Fracture; Detachment of Device or Device Component; Break; Fracture; Detachment of Device or Device Component
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) GII TIBIAL BASE IMPACTORS WERE BROKEN. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21619932 ↗MalfunctionDec 31, 2024Break; Fracture; Detachment of Device or Device Component; Break; Fracture; Detachment of Device or Device Component
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) GII TIBIAL BASE IMPACTORS WERE BROKEN. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21047264 ↗MalfunctionNov 11, 2024Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA, THE GII TIBIAL BASE IMPACTOR'S PLASTIC SPLINTERED DURING USE. ALL PLASTIC WAS REMOVED FROM THE WOUND, IN AN UNKNOWN WAY. THERE WAS NO DELAY AND THE PROCEDURE WAS FINISHED USING THE SAME DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20653187 ↗MalfunctionMay 8, 2024Fracture; Material Fragmentation; Fracture; Material Fragmentation
IT WAS REPORTED THAT DURING A TKA SURGERY, USING A GII TIBIAL BASE IMPACTOR, IT WAS FOUND THAT THE RUBBERIZED SURFACE WAS VERY BRITTLE AFTER A SHORT PERIOD OF USE AND WOULD SPLINTER IF HIT IN THE SURGICAL AREA. HOWEVER, AN ABDOMINAL DRAPE WAS USED TO ENSURE THAT NO SPLINTERS ENTERED THE BODY. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19274688 ↗MalfunctionApr 2, 2024Break; Break
IT WAS REPORTED THAT DURING A TKA, THE PLASTIC BUMPERS ON THE GII TIBIAL BASE IMPACTOR BROKE. ALL PIECES WERE RECOVERED. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19021342 ↗MalfunctionNov 27, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, THE GII TIBIAL BASE IMPACTOR FRACTURED DOWN THE CENTER. THE DEVICE FRACTURED WHILST IN USE, BUT THE PIECES REMAINED TOGETHER/INTACT. THE SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH THE SAME REPORTED DEVICE. NO INJURIES WERE REPORTED, NO PIECES FELL INSIDE THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18210050 ↗MalfunctionNov 27, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA PROCEDURE, WHEN IMPACTING THE TIBIAL IMPLANT A GII TIBIAL BASE IMPACTOR FRACTURED. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18209398 ↗MalfunctionJan 17, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A TKA SURGERY, ONE (1) GII TIBIAL BASE IMPACTOR FRACTURED WHILE IMPACTING TIBIA. THE PIECES DID NOT FELL INTO PATIENT, ALL OF THEM WERE RETRIEVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16189887 ↗InjuryOct 18, 2022Fracture; Fracture
IT WAS REPORTED THAT DURING A TKA, WHEN IMPACTING THE TIBIAL TRAY, SHARDS OF PLASTIC WERE COMING OFF THE GII TIBIAL BASE IMPACTOR AND FALLING IN THE WOUND. ALL BROKEN PIECES REMOVED WITH FORCEPS. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15626678 ↗MalfunctionJun 19, 2022Break; Break
IT WAS REPORTED THAT, FIVE (5) GII TIBIAL BASE IMPACTORS WERE FOUND WITH BLACK PLASTIC CHIPPING. AS THIS WAS NOTICE DURING SET UP OR INSPECTION, NOT PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14741402 ↗MalfunctionJun 13, 2022Break; Break
IT WAS REPORTED THAT, THERE WAS CHIP IN THE GII TIBIAL BASE IMPACTOR PLASTIC. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14678219 ↗MalfunctionMay 12, 2022Crack; Crack
IT WAS REPORTED THAT THE HEAD OF THE GII TIBIAL BASE IMPACTOR WAS CRACKED. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14380408 ↗MalfunctionMay 4, 2022Crack; Crack
IT WAS REPORTED THAT A GII TIBIAL BASE IMPACTOR CRACKED. NO CASE INVOLVED; THEREFORE, THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14283743 ↗InjuryApr 28, 2022Material Fragmentation; Naturally Worn; Material Fragmentation; Naturally Worn
IT WAS REPORTED THAT, DURING TKA SURGERY, THE GENESIS II TIBIAL BASEPLATE IMPACTOR IS LEAVING BLACK BITS IN THE PATIENT DURING USE, DUE TO WEAR AND TEAR. A THOROUGH WASHOUT WAS PERFORMED TO AVOID ANY BITS BEING LEFT IN THE PATIENT. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14246271 ↗InjuryApr 25, 2022Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT, DURING A TKA SURGERY, THE GII TIBIAL BASE IMPACTOR WAS PRODUCING THIRD BODY WEAR IN THE JOINT WHEN PLACING THE FINAL IMPLANT. THE JOINT WAS PULSE LAVAGED TO TRY AND REMOVE ANY PARTICULATES. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITHOUT DELAY. THE PATIENT WAS NOT HARMED BEYOND THE REPORTED PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14191580 ↗MalfunctionApr 21, 2022Crack; Crack
IT WAS REPORTED THAT THE GII TIBIAL BASE IMPACTOR WAS CRACKED. AS THIS WAS NOTICED UPON INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14177218 ↗MalfunctionApr 9, 2022Break; Break
IT WAS REPORTED THAT, THE BLACK TIP OF THE GII TIBIAL BASE IMPACTOR BROKE. AS THIS WAS NOTICE IN A FIELD INSPECTION, NOT PATIENT WAS INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14068644 ↗MalfunctionMar 31, 2022Break; Break
IT WAS REPORTED THAT THE TIP OF THE GII TIBIAL BASE IMPACTOR IS CHIPPING OFF. FURTHER INFORMATION IS UNKNOWN.
Patient problems: Insufficient Information
MAUDE report 13961973 ↗MalfunctionMar 29, 2022Crack; Crack
IT WAS REPORTED THAT A GENESIS II TIBIAL BASEPLATE IMPACTOR IS CRACKED. AS THIS WAS NOTICED UPON FIELD INSPECTION, THERE WAS NOT PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13934850 ↗MalfunctionMar 17, 2022Break; Break
IT WAS REPORTED THAT, AFTER A TKA SURGERY, IT WAS NOTICED THAT THE GII TIBIAL BASE IMPACTOR HAS BROKEN DOWN FROM REPEATED USE. AS THIS HAPPENED AFTER SURGERY, THE PATIENT WAS NO LONGER INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13795915 ↗InjuryMar 16, 2022Break; Break
IT WAS REPORTED THAT, DURING TOTAL KNEE ARTHROPLASTY, THE PLASTIC TIP OF A GENESIS II TIBIAL BASEPLATE IMPACTOR CHIPPED OFF WHILE IMPACTING, INSIDE THE PATIENT. THE PIECES WERE REMOVED BY HAND AND TWEEZERS. SURGERY WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED BEYOND THE PROBLEM REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13777795 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 18, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page