Skip to content

ACUFEX 7205517Recalls & reported problems

Smith & Nephew, Inc. · ACUFEX DIRECTOR DRILL GUIDE. REQUIRES 1 AIMER AND 1 BULLET NOT INCLUDED · Class 1

All models of ACUFEX (86)

Data updated Oct 16, 2026 from official FDA records · Editorial policy

Adverse event reports
4
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 4 adverse event reports that list this exact device (DI 03596010373250), received between Feb 13, 2019 and Mar 18, 2025 — classified as 4 malfunctions.

The most frequently reported specific device problem is “Break” (1 report).

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 1 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break1
  • Defective Device1
  • Device Markings/Labelling Problem1
  • Failure to Align1
  • Material Twisted/Bent1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient1
  • No Patient Involvement1

Recent adverse event reports

The 4 most recent of 4 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 18, 2025Break; Break

    IT WAS REPORTED THAT DURING AN ARTHROSCOPY, INSTRUMENTS OUTSIDE THE PATIENT, THE THREAD ON THE SCREW OF THE ACUFEX DIRECTOR DRILL GUIDE FUSED AND BROKE (THE SCREW HEAD BROKE). THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES USING THE SAME FAULTY DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21628260 ↗
  • MalfunctionDec 17, 2020Device Markings/Labelling Problem; Device Markings/Labelling Problem

    IT WAS REPORTED THAT THE ARROW ON THE ACL HANDLE HAS WORN OFF AND IT IS NO LONGER LEGIBLE; THEREFORE, IT IS IMPOSSIBLE TO SET THE ANGLE OF THE AIMER ACCURATELY. THE MALFUNCTION WAS FOUND DURING SET UP AND WAS SOLVED WITH NO DELAYS OR PATIENT HARM USING A BACK UP DEVICE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 11026659 ↗
  • MalfunctionNov 1, 2020Failure to Align; Material Twisted/Bent; Failure to Align; Material Twisted/Bent

    IT WAS REPORTED THAT DURING THE SET UP BEFORE THE SURGERY, IT WAS NOTICED AN INACCURACY OF THE TIBIA TIP AIMER, 4 POINT BULLET AND ACUFEX TIBIA TIP AIMER HAVE BEEN ALL SET UP, 2.4MM GUIDE PIN WAS USED AND IT WAS NOTICED THAT THE EXIT OF THE GUIDE DID NOT HIT THE TIP OF THE AIMER, IT SLIGHTLY DEVIATED TO THE TIP. NO PATIENT INJURIES OR DELAY REPORTED. SMITH AND NEPHEW BACK-UP DEVICE WAS AVAILABLE TO COMPLETE THE SURGERY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement

    MAUDE report 10767700 ↗
  • MalfunctionFeb 13, 2019Defective Device

    IT WAS REPORTED THAT THE DEVICE WAS DEFECTIVE. PRIOR TO DRILLING THE TIBIAL TUNNEL, WE COULD NOT ASSEMBLE THE GUIDE DUE TO THE FACT THAT THE BULLET WAS NOT ABLE TO SLIDE INTO THE GUIDE HANDLE. THE PROCEDURE WAS COMPLETED WITH A COMPETITOR DEVICE. NO PATIENT INJURIES WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 8336020 ↗

Other ACCESSORIES,ARTHROSCOPIC devices

All ACCESSORIES,ARTHROSCOPIC devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page