ACUFEX 7205524Recalls & reported problems
Smith & Nephew, Inc. · ACUFEX DIRECTOR ANGLED BULLET · Class 1
All models of ACUFEX (86)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 7
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010373267), received between Aug 3, 2017 and Apr 8, 2026 — classified as 2 injuries, 5 malfunctions.
The most frequently reported specific device problem is “Break” (2 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break2
- Illegible Information2
- Failure to Align1
- Flaked1
- Mechanical Jam1
Patient problems
- No Clinical Signs, Symptoms or Conditions2
- No Information2
- No Known Impact Or Consequence To Patient2
- Injury1
- Perforation1
Recent adverse event reports
The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.
MalfunctionApr 8, 2026Failure to Align
IT WAS REPORTED THAT, DURING A KNEE ARTHROSCOPY WITH MENISCAL SUTURE, WHEN USING THE DRILL BIT AND SLEEVE IT DID NOT PASS THROUGH ONE (1) ACUFEX DIRECTOR ANGLED BULLET. SURGEON HAD TO DRILL SEVERAL TIMES WITH THE GUIDE. THE PROCEDURE WAS PERFORMED, WITH A DELAY GREATER THAN 30 MIN, USING THE SAME DEVICE. ADDITIONAL ANESTHESIA WAS REQUIRED AS CONSEQUENCE OF THE SURGICAL PROLONGATION. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; Perforation
MAUDE report 24818285 ↗InjuryJan 8, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING A CRUCIATE LIGAMENT RECONSTRUCTION, WHEN USING THE TOOLS, DEVICE WAS UNABLE TO CATCH THE GUIDER WHEN USING THE POSITIONER GUIDE HANDLE. THIS RESULT IN THE HANDLE FALLING OFF. THERE WAS CHANGE IN THE SURGICAL TECHNIQUE AND A SIGNIFICANT DELAY. NO FURTHER COMPLICATIONS REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11140481 ↗MalfunctionJun 18, 2020Illegible Information; Illegible Information
IT WAS REPORTED THE BLT ASSY ANGLED DIRECTOR ACL HAD THE CALIBRATION MARKS NO LONGER LEGIBLE. THERE IS NOT A CASE REPORTED BUT THIS HAS BEEN AN ONGOING ISSUE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10168798 ↗MalfunctionJun 18, 2020Illegible Information; Illegible Information
IT WAS REPORTED THE BLT ASSY ANGLED DIRECTOR ACL HAD THE CALIBRATION MARKS NO LONGER LEGIBLE. THERE IS NOT A CASE REPORTED BUT THIS HAS BEEN AN ONGOING ISSUE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Information
MAUDE report 10168741 ↗InjuryMar 30, 2020Flaked; Flaked
IT WAS REPORTED THAT, DURING AN ACL RECONSTRUCTION, METAL SHEDDINGS WERE SEEN COMING OFF IN THE JOINT. THIS WAS NOTICED IN THE FEMORAL TUNNEL WHILE THE AXCUFEX'S 2.4 MM BULLET GUIDE WAS IN CONTACT WITH THE PASSING PIN. MOST OF THE METAL SHEDDINGS WERE REMOVED FROM THE KNEE JOINT WITH A MECHANICAL SHAVER. THE PROCEDURE WAS COMPLETED WITH THE SAME BULLET GUIDE. THERE WAS NO SIGNIFICANT DELAY REPORTED. THE PATIENT'S OUTCOME IS AS EXPECTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: Injury
MAUDE report 9901437 ↗MalfunctionMay 27, 2018Break
IT WAS REPORTED DURING A CRUCIATE LIGAMENT RECONSTRUCTION SURGERY, THE KIRSCHNER WIRE BROKE IN THE ANGLE DIRECTOR AND COULDN¿T BE PULLED OUT. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7547267 ↗MalfunctionAug 3, 2017Break
IT WAS REPORTED THAT THE DEVICE BROKE WHILE USING IT TO TRY TO REMOVE A SCREW. NO BACKUP DEVICE WAS AVAILABLE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 6762985 ↗
Other ACCESSORIES,ARTHROSCOPIC devices
- Arthrex® AR-12990Arthrex, Inc. · Accessories, arthroscopic409 reports
- NA 228302Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC287 reports
- FIRSTPASS 72290128Smith & Nephew, Inc. · ACCESSORIES,ARTHROSCOPIC232 reports
- NA 251723Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC145 reports
- COBRA 3910900093Stryker Corporation · Accessories, arthroscopic138 reports
- FirstPass 72290129Smith & Nephew, Inc. · Accessories, arthroscopic82 reports
- FirstPass 72290130Smith & Nephew, Inc. · Accessories, arthroscopic80 reports
- MITEK 242025Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC61 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.