Skip to content

TRIGEN 71118242Recalls & reported problems

Smith & Nephew, Inc. · 14.0MM PILOT NOSE INTERCHANGE REAMER HEAD · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
1
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 1 adverse event report that list this exact device (DI 03596010409492), received between Sep 10, 2020 and Sep 10, 2020 — classified as 1 malfunction.

The most frequently reported specific device problem is “Defective Device” (1 report).

2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Most reported problems

Device problems

  • Defective Device1

Patient problems

  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 1 most recent of 1 reports, in the reporter's own words as published by the FDA.

  • MalfunctionSep 10, 2020Defective Device; Defective Device

    IT WAS REPORTED THAT THERE ARE NO NUMERICAL MARKINGS VISIBLE ON THE FLEX REAMER ENDS AND NURSING OFTEN MIX UP THE REAMER ENDS WHEN THEY ARE BEING ASKED FOR BY THE SURGEON. EVENT OCCURRED DURING SURGERY, WITH INSTRUMENTS OUTSIDE THE PATIENT. PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. A SURGICAL DELAY OF 30 MIN OR LESS WAS REPORTED. NO HARM TO PATIENT WAS REPORTED DUE TO THIS FAILURE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10518224 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.

  • Z-0730-2019 · Jan 7, 2019 · Terminated
    SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
  • Z-1008-2017 · Nov 16, 2016 · Terminated
    REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

Other Prosthesis, hip, femoral component, cemented, metal devices

All Prosthesis, hip, femoral component, cemented, metal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page