SYNERGY 71366709Recalls & reported problems
Smith & Nephew, Inc. · SYNERGY FULLY TOOTHED BROACH SIZE 9 · Class 2
All models of SYNERGY (52)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 0
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010429629), received between Sep 27, 2020 and Oct 19, 2020 — classified as 8 malfunctions.
The most frequently reported specific device problem is “Dull, Blunt” (6 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Most reported problems
Device problems
- Dull, Blunt6
- Nonstandard Device6
- Material Deformation3
- Connection Problem2
- Defective Device1
- Fitting Problem1
- Incomplete or Inadequate Connection1
Patient problems
- No Consequences Or Impact To Patient7
- No Clinical Signs, Symptoms or Conditions4
- No Patient Involvement1
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 19, 2020Nonstandard Device; Dull, Blunt; Material Deformation; Nonstandard Device; Dull, Blunt; Material Deformation
IT WAS REPORTED THAT, AFTER PROCEDURE, THIS BROACH WAS NOTICED TO BE DULL AND NOT TRUE TO SIZE. THE DEVICE HAD BEEN USED TO COMPLETE THE PROCEDURE AND NO ISSUES WERE REPORTED DURING CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10697006 ↗MalfunctionOct 19, 2020Nonstandard Device; Dull, Blunt; Material Deformation; Nonstandard Device; Dull, Blunt; Material Deformation
IT WAS REPORTED THAT, AFTER PROCEDURE, THIS BROACH WAS NOTICED TO BE DULL AND NOT TRUE TO SIZE. THE DEVICE HAD BEEN USED TO COMPLETE THE PROCEDURE AND NO ISSUES WERE REPORTED DURING CASE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10697002 ↗MalfunctionOct 19, 2020Nonstandard Device; Dull, Blunt; Nonstandard Device; Dull, Blunt
IT WAS REPORTED THAT, AFTER PROCEDURE, THIS BROACH WAS NOTICED TO BE DULL AND NOT TRUE TO SIZE. THE DEVICE HAD BEEN USED TO COMPLETE THE PROCEDURE AND NO ISSUES WERE REPORTED DURING CASE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10696997 ↗MalfunctionOct 19, 2020Nonstandard Device; Dull, Blunt; Nonstandard Device; Dull, Blunt
IT WAS REPORTED THAT, AFTER PROCEDURE, THIS BROACH WAS NOTICED TO BE DULL AND NOT TRUE TO SIZE. THE DEVICE HAD BEEN USED TO COMPLETE THE PROCEDURE AND NO ISSUES WERE REPORTED DURING CASE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10696994 ↗MalfunctionOct 19, 2020Nonstandard Device; Dull, Blunt; Nonstandard Device; Dull, Blunt
IT WAS REPORTED THAT, AFTER PROCEDURE, THIS BROACH WAS NOTICED TO BE DULL AND NOT TRUE TO SIZE. THE DEVICE HAD BEEN USED TO COMPLETE THE PROCEDURE AND NO ISSUES WERE REPORTED DURING CASE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 10696985 ↗MalfunctionOct 19, 2020Nonstandard Device; Dull, Blunt; Material Deformation; Nonstandard Device; Dull, Blunt; Material Deformation
IT WAS REPORTED THAT, AFTER PROCEDURE, THIS BROACH WAS NOTICED TO BE DULL AND NOT TRUE TO SIZE. THE DEVICE HAD BEEN USED TO COMPLETE THE PROCEDURE AND NO ISSUES WERE REPORTED DURING CASE.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10696977 ↗MalfunctionOct 7, 2020Defective Device; Connection Problem; Defective Device; Connection Problem
IT WAS REPORTED THAT SURGEON DOES NOT LIKE HOW THE BROACHES ARE SEATING. NO CASE RELATED, THEREFORE NO PATIENT INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Patient Involvement
MAUDE report 10645664 ↗MalfunctionSep 27, 2020Fitting Problem; Connection Problem; Incomplete or Inadequate Connection; Fitting Problem; Connection Problem; Incomplete or Inadequate Connection
IT WAS REPORTED THAT THE BROACHES WERE NOT SEATING PROPERLY. IT IS UNKNOWN WHEN THE ISSUE WAS DISCOVERED. IT IS UNKNOWN IF THERE WAS A DELAY OR IF A BACKUP DEVICE WAS AVAILABLE. NO PATIENT INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Consequences Or Impact To Patient
MAUDE report 10591330 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page