Skip to content

TRIGEN 71118230Recalls & reported problems

Smith & Nephew, Inc. · 11.0mm SCULPTOR FIXED ENDCUTTING REAMER · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
2
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03596010436634), received between Dec 8, 2022 and Mar 3, 2026 — classified as 2 malfunctions.

The most frequently reported specific device problem is “Fracture” (2 reports).

2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 1 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Fracture2

Patient problems

  • No Clinical Signs, Symptoms or Conditions2

Recent adverse event reports

The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMar 3, 2026Fracture

    IT WAS REPORTED THAT DURING AN INTRAMEDULLARY NAIL PROCEDURE, WHILE REAMING WAS UNDERWAY, THE 11.0MM FIXED ENDCUTTING RMR FRACTURED. NO PIECES FELL INTO THE PATIENT'S WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A CHANGE OF THE SURGICAL TECHNIQUE. THE SURGICAL TECHNIQUE WAS CHANGED TO REAM UNTIL THE HALFWAY POINT. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24502173 ↗
  • MalfunctionDec 8, 2022Fracture; Fracture

    IT WAS REPORTED THAT DURING SURGERY ONE (1) 11.0MM FIXED ENDCUTTING RMR HEAD HIT A BLOCKING WIRE CAUSING THE SHAFT OF THE REAMER TO SPIRAL SPLIT ALL THE WAY UP. NO PIECES FELL INTO THE WOUND. THE SHAFT BROKE OUTSIDE THE PATIENT WHILE THE REAMER HEAD WAS INSIDE PATIENT SO NOTHING BROKE INTO THE PATIENTS WOUND. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING THE SAME DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 15946681 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.

  • Z-0730-2019 · Jan 7, 2019 · Terminated
    SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
  • Z-1008-2017 · Nov 16, 2016 · Terminated
    REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

Other Prosthesis, hip, femoral component, cemented, metal devices

All Prosthesis, hip, femoral component, cemented, metal devices →

Sources and updates

How reports and recalls are matched · Editorial policy · Report an error on this page