NA 71362748Recalls & reported problems
Smith & Nephew, Inc. · REFLECTION REAMER DOME 48MM · Class 2
- Adverse event reports
- 2
- Deaths reported
- 0
- Injuries reported
- 1
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 2 adverse event reports that list this exact device (DI 03596010456861), received between Oct 8, 2020 and Apr 12, 2021 — classified as 1 injury, 1 malfunction.
The most frequently reported specific device problem is “Break” (2 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Most reported problems
Device problems
- Break2
Patient problems
- No Clinical Signs, Symptoms or Conditions1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 2 most recent of 2 reports, in the reporter's own words as published by the FDA.
InjuryApr 12, 2021Break; Break
IT WAS REPORTED THAT DURING TOTAL HIP REPLACEMENT THE REAMER CROSSBAR FAILED AT ITS WELD WITH THE REAMER BASKET DUE TO FATIGUE. NO DELAY OR INJURY OCCURRED. THE PROCEDURE WAS FINISHED USING A SMITH AND NEPHEW BACK UP DEVICE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11648456 ↗MalfunctionOct 8, 2020Break; Break
IT WAS REPORTED THAT THE REAMER DOME CROSS BAR BROKE OFF THE REAMER DOME. IT IS UNKNOWN IF THIS OCCURRED DURING THE PROCEDURE OR IF THERE WAS ANY DELAY. NO PATIENT HARM WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10652660 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources: FDA GUDID, MAUDE and Medical Device Recalls via openFDA. openFDA data is not validated for clinical or regulatory use.