Skip to content

NA 71362754Recalls & reported problems

Smith & Nephew, Inc. · REFLECTION REAMER DOME 54MM · Class 2

Data updated Oct 13, 2026 from official FDA records · Editorial policy

Adverse event reports
7
Deaths reported
0
Injuries reported
2
Recalls naming it
0

Summary

The FDA's MAUDE database contains 7 adverse event reports that list this exact device (DI 03596010456922), received between Jul 29, 2019 and May 18, 2026 — classified as 2 injuries, 5 malfunctions.

The most frequently reported specific device problem is “Break” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 1 reportsSep 2024: 0 reportsOct 2024: 0 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 0 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 1 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 1 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Break3
  • Fracture3
  • Detachment of Device or Device Component1

Patient problems

  • No Clinical Signs, Symptoms or Conditions5
  • No Consequences Or Impact To Patient1
  • No Known Impact Or Consequence To Patient1

Recent adverse event reports

The 7 most recent of 7 reports, in the reporter's own words as published by the FDA.

  • MalfunctionMay 18, 2026Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, ONE (1) REAMER DOME 54MM EXPERIENCED BREAKAGE OF THE X-SHAPED COMPONENT THAT CONNECTS THE REAMER DOME TO THE REAMER HANDLE WHILE REAMING. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK- UP. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 25211513 ↗
  • MalfunctionJan 22, 2026Fracture

    IT WAS REPORTED THAT, DURING A THR SURGERY, THE REAMER ATTACHMENT BARS OF ONE (1) REAMER DOME 54MM BROKE OFF. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED OR IF THIS CAUSED ANY DELAY. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 24145747 ↗
  • MalfunctionAug 12, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A THR, WHEN FRAZING THE ACEBULAR, THERE WAS A BREAKAGE OF THE REAMER DOME 54MM CUTTER, DUE TO HIGH RESISTANCE AND GOOD BONE QUALITY. PIECE WAS COMPLETELY REMOVED BY THE SURGEON. SURGERY WAS PERFORMED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. NO HARM TO PATIENT WAS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 19960964 ↗
  • InjuryJun 9, 2022Detachment of Device or Device Component; Detachment of Device or Device Component

    IT WAS REPORTED THAT, DURING THR SURGERY, THE PHYSICIAN WAS REAMING THE ACETABULUM WHEN THE CROSSBAR ON THE REAMER DOME 54MM BECAME DETACHED FROM THE REAMER. THE DETACHED PIECE WAS REMOVED FROM THE PATIENT. THE PROCEDURE WAS COMPLETED WITHOUT SURGICAL DELAY THE SAME DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 14640483 ↗
  • MalfunctionMar 28, 2022Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, THE REAMER DOME 54MM FRAME BROKE UPON INTRODUCING INTO ACETABULUM. NO PIECE FELL INTO THE PATIENT. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A SMITH AND NEPHEW BACK-UP DEVICE. PATIENT WAS NOT HARMED. NO FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 13927024 ↗
  • InjuryAug 10, 2020Break; Break

    IT WAS REPORTED THAT THE CROSSBARS OF THE REAMER DOME BROKE DURING REAMING. STAFF WAS ABLE TO GET ALL PIECES, RETRIEVE THE DOME, OPEN ANOTHER SET AND CONTINUE CASE. THE REAMER DOME WAS STUCK, BUT IT GOT OUT. SLIGHT DELAY TO OPEN ANOTHER REAMER TRAY. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10388407 ↗
  • MalfunctionJul 29, 2019Break; Break

    IT WAS REPORTED THAT DURING A THA PROCEDURE, REAMER BROKE DURING REAMING THE ACETABULUM. DELAY FORM 30 MINUTES TO 1 HOUR REPORTED. BACKUP DEVICE AVAILABLE. NO IMPACT OR INJURY TO PATIENT REPORTED.

    Patient problems: No Consequences Or Impact To Patient

    MAUDE report 8839842 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.

  • Z-0730-2019 · Jan 7, 2019 · Terminated
    SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
  • Z-1008-2017 · Nov 16, 2016 · Terminated
    REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

Other Prosthesis, hip, femoral component, cemented, metal devices

All Prosthesis, hip, femoral component, cemented, metal devices →

Sources and updates