NA 71362755Recalls & reported problems
Smith & Nephew, Inc. · REFLECTION REAMER DOME 55MM · Class 2
Data updated Oct 13, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 8
- Deaths reported
- 0
- Injuries reported
- 5
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 8 adverse event reports that list this exact device (DI 03596010456939), received between Jul 15, 2020 and Oct 19, 2023 — classified as 5 injuries, 3 malfunctions.
The most frequently reported specific device problem is “Break” (3 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break3
- Material Fragmentation2
- Flaked1
- Fracture1
- Solder Joint Fracture1
Patient problems
- No Clinical Signs, Symptoms or Conditions6
- No Information2
- No Known Impact Or Consequence To Patient2
Recent adverse event reports
The 8 most recent of 8 reports, in the reporter's own words as published by the FDA.
MalfunctionOct 19, 2023Fracture; Fracture
IT WAS REPORTED THAT, DURING A THR SURGERY, THE REAMER DOME 55MM ATTACHMENT ARMS SNAPPED DURING USE. ALL PARTS WERE RECOVERED. THE SURGERY WAS COMPLETED, WITHOUT ANY DELAY, WITH A S/N BACK-UP DEVICE. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17966111 ↗InjuryJan 4, 2022Break; Break
IT WAS REPORTED THAT, DURING A THR SURGERY, THE SURGEON WAS REAMING ACETABULUM AND THE REAMER DOME 55MM SNAPPED OFF FROM THE BASE. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 13152549 ↗MalfunctionFeb 9, 2021Solder Joint Fracture; Solder Joint Fracture
IT WAS REPORTED THAT THE REAMER DOME DEVICE WAS FOUND WITH WELDS BROKEN. NO CASE INVOLVED; THEREFORE, NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11300576 ↗MalfunctionDec 9, 2020Break; Break
IT WAS REPORTED THAT, DURING INSPECTION, THE REAMER DOME 55MM WAS NOTICED TO BE BROKEN. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10976323 ↗InjurySep 17, 2020Break; Break
IT WAS REPORTED THAT DURING A THR THE REAMER DOME 55MM WAS FOUND TO BE BROKEN, INTRAOPERATIVELY; THERE IS NO INFORMATION REGARDING HOW THE PROCEDURE WAS COMPLETED, NO DELAY OR OTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Known Impact Or Consequence To Patient
MAUDE report 10549379 ↗InjuryJul 18, 2020Flaked; Flaked
IT WAS REPORTED THAT AFTER THR PROCEDURE IT WAS NOTICED THAT THE REAMER WAS SHEDDING METAL INSIDE THE PATIENT. NO DELAY. THE PROCEDURE WAS FINISHED USING THE SAME DEVICE.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 10293459 ↗InjuryJul 15, 2020Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE REAMER CROSS SUPPORT BROKE OUT OF THE REAMER DOME 55MM WHILE REAMING DURING A THR. THE DEVICE BROKE INSIDE THE PATIENT AND THE DEVICES HAD TO BE RETRIEVED. AN S&N BACKUP WAS NOT REQUIRED AS THE SURGERY WAS COMPLETED WITH THE SAME DEVICE. NO DELAY REPORTED. THE PATIENT WAS NOT HARMED BEYOND THE REPORTED EVENT.
Patient problems: No Information
MAUDE report 10279222 ↗InjuryJul 15, 2020Material Fragmentation; Material Fragmentation
IT WAS REPORTED THAT THE CROSS SUPPORT BROKE OUT OF THE 55 MM REAMER DOME WHILE REAMING INSIDE THE PATIENT. ALL PIECES WERE RETRIEVED FROM THE PATIENT. AN S&N BACKUP WAS NOT REQUIRED AS THE PROCEDURE WAS FINISHED WITH THE SAME DEVICE. SURGERY WAS NOT DELAYED. THE PATIENT WAS NOT HARMED BEYOND THE DESCRIBED EVENT.
Patient problems: No Clinical Signs, Symptoms or Conditions; No Information
MAUDE report 10279160 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 13, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 13, 2026
- Page last rebuilt: Oct 13, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page