Skip to content

REFLECTION 71362950Recalls & reported problems

Smith & Nephew, Inc. · REFLECTION FLEXIBLE DRILL 50MM · Class 2

Adverse event reports
6
Deaths reported
0
Injuries reported
0
Recalls naming it
0

Summary

The FDA's MAUDE database contains 6 adverse event reports that list this exact device (DI 03596010463005), received between Aug 3, 2020 and Apr 26, 2025 — classified as 6 malfunctions.

The most frequently reported specific device problem is “Material Twisted/Bent” (3 reports).

2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.

Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.

Reports received per month

Nov 2023: 0 reportsDec 2023: 0 reportsJan 2024: 0 reportsFeb 2024: 0 reportsMar 2024: 0 reportsApr 2024: 0 reportsMay 2024: 0 reportsJun 2024: 0 reportsJul 2024: 0 reportsAug 2024: 0 reportsSep 2024: 0 reportsOct 2024: 1 reportsNov 2024: 0 reportsDec 2024: 0 reportsJan 2025: 0 reportsFeb 2025: 0 reportsMar 2025: 0 reportsApr 2025: 1 reportsMay 2025: 0 reportsJun 2025: 0 reportsJul 2025: 0 reportsAug 2025: 0 reportsSep 2025: 0 reportsOct 2025: 0 reportsNov 2025: 0 reportsDec 2025: 0 reportsJan 2026: 0 reportsFeb 2026: 0 reportsMar 2026: 0 reportsApr 2026: 0 reportsMay 2026: 0 reportsJun 2026: 0 reportsJul 2026: 0 reportsAug 2026: 0 reportsSep 2026: 0 reportsOct 2026: 0 reports2024202520261
Last 36 months. The most recent month may be incomplete.

Most reported problems

Device problems

  • Material Twisted/Bent3
  • Break2
  • Fracture1

Patient problems

  • No Clinical Signs, Symptoms or Conditions3
  • No Known Impact Or Consequence To Patient3

Recent adverse event reports

The 6 most recent of 6 reports, in the reporter's own words as published by the FDA.

  • MalfunctionApr 26, 2025Break; Break

    IT WAS REPORTED THAT, DURING A THR SURGERY, ONE (1) FLEXIBLE DRILL 50MM BROKE WHILE THE SURGEON WAS PERFORMING THE DRILLING. THE PIECES WERE COMPLETELY REMOVED FROM THE PATIENT. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 21915680 ↗
  • MalfunctionOct 9, 2024Fracture; Fracture

    IT WAS REPORTED THAT DURING A HIP ARTHROSCOPY PROCEDURE, THE TIP OF THE FLEXIBLE DRILL 50MM BROKE OFF. SURGEON WAS ABLE TO REMOVE THE PIECE. THE PROCEDURE WAS PERFORMED, AFTER NO DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT HARMED AS CONSEQUENCE OF THIS PROBLEM.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 20409318 ↗
  • MalfunctionDec 13, 2021Break; Break

    IT WAS REPORTED THAT, DURING THA SURGERY, A FLEXIBLE DRILL 50MM BIT SNAPPED BEFORE IT GETS TO THE DESIRE DEPTH. ALL PIECES OF THE DRILL WERE RETRIEVED. SURGERY WAS COMPLETED, AFTER A DELAY OF 30 MINUTES OR LESS, WITH A SN BACK-UP DEVICE. NO HARM TO THE PATIENT OR FURTHER COMPLICATIONS REPORTED.

    Patient problems: No Clinical Signs, Symptoms or Conditions

    MAUDE report 12990815 ↗
  • MalfunctionOct 28, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE DRILL BIT WAS USED IN A TOTAL HIP REPLACEMENT PROCEDURE TO DRILL A HOLE FOR REFLECTION SCREW, AFTER THREE USES THE DRILL BENT. PROCEDURE FINISHED WITH A SMITH AND NEPHEW BACKUP DEVICE. PATIENT INJURIES WERE NOT REPORTED. SURGICAL DELAY LESS THAN 30 MINUTES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10749579 ↗
  • MalfunctionOct 28, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT THE DRILL BIT WAS USED IN A TOTAL HIP REPLACEMENT PROCEDURE TO DRILL A HOLE FOR REFLECTION SCREW, AFTER THREE USES THE DRILL BENT. PROCEDURE FINISHED WITH A SMITH AND NEPHEW BACKUP DEVICE. PATIENT INJURIES WERE NOT REPORTED. SURGICAL DELAY LESS THAN 30 MINUTES. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10749542 ↗
  • MalfunctionAug 3, 2020Material Twisted/Bent; Material Twisted/Bent

    IT WAS REPORTED THAT DURING A SURGERY, THE FLEXIBLE DRILL BIT BENT. A NEW S+N DEVICE HAD TO BE OPENED TO CONTINUE THE PROCEDURE WITH A DELAY OF LESS THAN 30 MINUTES. NO OTHER COMPLICATIONS WERE REPORTED.

    Patient problems: No Known Impact Or Consequence To Patient

    MAUDE report 10354581 ↗

Recalls under the same FDA submission

These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.

  • Z-0730-2019 · Jan 7, 2019 · Terminated
    SL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
  • Z-1008-2017 · Nov 16, 2016 · Terminated
    REFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116

Other Prosthesis, hip, femoral component, cemented, metal devices

All Prosthesis, hip, femoral component, cemented, metal devices →