ARTHRO-PIERCE 7209496Recalls & reported problems
Smith & Nephew, Inc. · ARTHRO-PIERCE STRAIGHT · Class 1
- Adverse event reports
- 9
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 9 adverse event reports that list this exact device (DI 03596010474513), received between Jan 10, 2018 and Mar 2, 2026 — classified as 2 injuries, 7 malfunctions.
The most frequently reported specific device problem is “Break” (9 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. How this data works · Data updated Oct 13, 2026.
Reports received per month
Most reported problems
Device problems
- Break9
Patient problems
- No Clinical Signs, Symptoms or Conditions7
- Foreign Body In Patient1
- No Known Impact Or Consequence To Patient1
Recent adverse event reports
The 9 most recent of 9 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 2, 2026Break
IT WAS REPORTED THAT, DURING A SHOULDER ARTHROSCOPY, WHEN THE DOCTOR TRIED TO PASS AN ARTHROPIERCE STRAIGHT THROUGH THE TENDON, IT BROKE INSIDE THE PATIENT. THE BROKEN PIECE WAS REMOVED. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A SIMILAR S+N BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24479950 ↗InjuryFeb 14, 2026Break
IT WAS REPORTED THAT, DURING A LEFT SHOULDER ARTHROSCOPY WITH ROTATOR CUFF REPAIR PROCEDURE, THE ARTHROPIERCE BROKE AND THE TIP GOT LOST IN THE PATIENT. THE PROCEDURE WAS RESUMED USING AN EQUIVALENT S+N BACK-UP. IT IS UNKNOWN IF THERE WAS A SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: Foreign Body In Patient
MAUDE report 24341291 ↗MalfunctionAug 26, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE ARTHROPIERCE CLAMP BROKE WHEN IT WAS REMOVED FROM THE JOINT. THE BROKEN PART WAS SEARCHED INTERNALLY AND EXTERNALLY ON THE PATIENT BUT WAS NOT FOUND. THE SURGERY WAS COMPLETED WITH A SMITH AND NEPHEW BACK-UP DEVICE. THERE WAS A DELAY OF LESS THAN 30 MINUTES, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22901395 ↗MalfunctionJul 21, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE ARTHROPIERCE GOT BROKEN, THE PROCEDURE WAS SUCCESSFULLY COMPLETED USING A SMITH AND NEPHEW BACKUP DEVICE WITH A SURGICAL DELAY OF LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22562201 ↗MalfunctionJul 20, 2025Break; Break
IT WAS REPORTED THAT DURING AN ARTHROSCOPY PROCEDURE, THE TIP OF THE ARTHROPIERCE GOT BROKEN. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITH A DELAY OF LESS THAN 30 MINUTES USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22552268 ↗MalfunctionDec 2, 2024Break; Break
IT WAS REPORTED THAT, DURING A ROTATOR CUFF REPAIR, THE ARTHROPIERCE STRAIGHT TIP WAS FRACTURED. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE INSTEAD. THERE WAS NO SURGICAL DELAY, AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20826495 ↗MalfunctionDec 7, 2023Break; Break
IT WAS REPORTED THAT, DURING AN ARTHROSCOPY, THE TIP OF TWO (2) ARTHROPIERCE STRAIGHT BROKE INSIDE THE PATIENT'S JOINT. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE INSTEAD. THERE WAS A DELAY OF LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED. PATIENT'S STATUS IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18283391 ↗InjuryMar 1, 2023Break; Break
IT WAS REPORTED THAT DURING A SHOULDER INSTABILITY PROCEDURE, THE ARTHROPIERCE STRAIGHT BRID FORCEPS BROKE. IT IS UNKNOWN HOW THE PROCEDURE WAS COMPLETED; HOWEVER, NO SURGICAL DELAY WAS REPORTED. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 16464418 ↗MalfunctionJan 10, 2018Break
IT WAS REPORTED THAT THE JAW BROKE DURING THE SURGERY NO PARTS WERE MISSING. NO PATIENT INJURY WAS REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7177701 ↗
Other ACCESSORIES,ARTHROSCOPIC devices
- Arthrex® AR-12990Arthrex, Inc. · Accessories, arthroscopic409 reports
- NA 228302Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC287 reports
- FIRSTPASS 72290128Smith & Nephew, Inc. · ACCESSORIES,ARTHROSCOPIC232 reports
- NA 251723Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC145 reports
- COBRA 3910900093Stryker Corporation · Accessories, arthroscopic138 reports
- FirstPass 72290129Smith & Nephew, Inc. · Accessories, arthroscopic82 reports
- FirstPass 72290130Smith & Nephew, Inc. · Accessories, arthroscopic80 reports
- MITEK 242025Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC61 reports