ARTHRO-PIERCE 7209497Recalls & reported problems
Smith & Nephew, Inc. · ARTHROPIERCE 45 DEG RIGHT · Class 1
All models of ARTHRO-PIERCE (9)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
- Adverse event reports
- 18
- Deaths reported
- 0
- Injuries reported
- 2
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 18 adverse event reports that list this exact device (DI 03596010474520), received between Feb 8, 2018 and Mar 30, 2026 — classified as 2 injuries, 16 malfunctions.
The most frequently reported specific device problem is “Break” (15 reports).
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break15
- Fracture2
- Detachment Of Device Component1
- Detachment of Device or Device Component1
- Material Separation1
- Scratched Material1
Patient problems
- No Clinical Signs, Symptoms or Conditions14
- No Known Impact Or Consequence To Patient3
- No Consequences Or Impact To Patient1
Recent adverse event reports
The 18 most recent of 18 reports, in the reporter's own words as published by the FDA.
MalfunctionMar 30, 2026Fracture; Detachment of Device or Device Component
IT WAS REPORTED THAT DURING A SHOULDER ARTHROSCOPY, IT WAS EVIDENT THAT THE JAW OF THE ARTHROPIERCE WAS FRACTURED.THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT SN BACK-UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24715432 ↗MalfunctionJul 3, 2025Break; Break
IT WAS REPORTED THAT, DURING A SHOULDER ARTHROSCOPY, THE TIP OF THE ARTHROPIERCE BROKE AND FELL INTO THE PATIENT. THE BROKEN PIECE WAS SUCCESSFULLY REMOVED WITH FORCEPS. SURGERY WAS COMPLETED WITH A BACK-UP DEVICE, INSTEAD. THERE WAS NO SURGICAL DELAY, AND NO FURTHER COMPLICATIONS WERE REPORTED. PATIENT'S STATUS IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 22391858 ↗MalfunctionFeb 21, 2025Break; Scratched Material; Break; Scratched Material
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, WHEN THE ARTHROPIERCE WAS PASSED, IT BROKE. NO DAMAGE CAUSED, THE PATIENT'S HEALTH CONDITION IS STABLE. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SURGICAL DELAY USING A SMITH AND NEPHEW BACK UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED. EVERY PIECE WAS REMOVED FROM THE PATIENT USING A GRASPER.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21437124 ↗MalfunctionFeb 21, 2025Break; Break
IT WAS REPORTED THAT DURING A SHOULDER INSTABILITY PROCEDURE THE ARTHROPIECE FORCEPS WERE BROKEN. THE PIECES WERE REMOVED. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE WITH NO SURGICAL DELAY. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21433205 ↗MalfunctionFeb 6, 2025Break; Break
IT WAS REPORTED THAT THE CLAMP OF THE ARTHROPIERCE BROKE, BUT IT IS UNKNOWN WHEN DURING THE SURGERY THE DEVICE FAILED. THE CLAMP WAS NOT SEARCHED FOR INSIDE THE PATIENT DUE TO THE UNCLEAR TIMING OF THE BREAKAGE. SURGERY WAS COMPLETED WITHOUT SURGICAL DELAY USING THE SAME DEVICE, AND NO FURTHER COMPLICATIONS WERE REPORTED. THE PATIENT'S STATUS IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21319315 ↗MalfunctionJan 20, 2025Fracture; Fracture
IT WAS REPORTED THAT DURING AN ARTHROSCOPY, THE TIP OF THE ARTHROPIERCE WAS FRACTURED WHILE TRYING TO PASS THE TISSUE. THE PROCEDURE WAS COMPLETED USING A BACK-UP DEVICE. THERE WAS A DELAY LESS THAN 30 MINUTES AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 21191269 ↗MalfunctionOct 2, 2024Break; Break
IT WAS REPORTED THAT, DURING A SHOULDER ARTHROSCOPY, THE ARTHROPIERCE BROKE AT THE TIP. THE BROKEN PIECE WAS REMOVED FROM THE PATIENT. IT IS UNKNOWN IF A BACK-UP DEVICE WAS AVAILABLE. NO DELAY OR FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 20352591 ↗MalfunctionFeb 13, 2024Break; Break
IT WAS REPORTED THAT DURING SET UP FOR AN ARTHROSCOPY, THE ARTHROPIERCE 45° RIGHT WAS FOUND DETACHED IN THE MIDDLE. IN ADDITION, A PART OF THE JAW BROKE OFF. THERE WAS NO SURGICAL DELAY REPORTED. THERE WAS NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18695437 ↗MalfunctionNov 22, 2023Break; Break
IT WAS REPORTED THAT DURING AN UNSPECIFIED ARTHROSCOPY, THE ARTHROPIERCE´S TIP BROKE WHEN PASSING THROUGH THE PATIENT´S LABRAL LESION. THE PIECES WERE COMPLETELY REMOVED. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. THERE WAS NO SURGICAL DELAY AND NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18196270 ↗MalfunctionJun 6, 2023Break; Break
IT WAS REPORTED, DURING AN ARTHROSCOPY, THE ARTHROPIERCE BROKE. ALL BROKEN PARTS WERE REMOVED FROM THE PATIENT USING ARTHROSCOPIC FORCEPS. SURGERY WAS COMPLETED WITHOUT DELAY, USING A CLEVERHOOK FROM DEPUY MITEK (COMPETITOR'S DEVICE). NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 17072675 ↗InjuryMar 22, 2021Break; Break
IT WAS REPORTED THAT DURING SURGERY, WHILE SUTURING THE SOFT TISSUES IN THE SHOULDER, THE SURGEON SAW THAT THE BRANCH OF THE ARTHROPIERCE DEVICE DID NOT LOCK. WHEN IT WAS REMOVED FROM THE JOINT, IT WAS NOTICED THAT THE JAMMED BRANCH BROKE OFF. A BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE WITH NO SIGNIFICANT DELAY OR OTHER COMPLICATIONS. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11547766 ↗MalfunctionMar 10, 2021Break; Break
IT WAS REPORTED THAT THE TIP OF THE ARTHROPIERCE 45 RIGHT WAS BROKEN. NO CASE INVOLVED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11456658 ↗InjuryJan 11, 2021Break; Break
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, WHEN THE SOFT TISSUES WERE SUTURED IN THE SHOULDER, THE ARTHROPIERCE BROKE OFF. BACKUP DEVICE WAS AVAILABLE TO COMPLETE THE PROCEDURE. NO DELAY AND NO PATIENT COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11152615 ↗MalfunctionDec 4, 2020Break; Break
IT WAS REPORTED THAT UPPER MOVABLE BRANCH OF THE ARTHRO-PIERCE WAS BROKEN. INCIDENT OCCURRED BEFORE THE PROCEDURE. THERE WAS NO DELAY AND A BACK-UP DEVICE WAS AVAILABLE. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 10944440 ↗MalfunctionJan 7, 2020Material Separation; Material Separation
IT WAS REPORTED THAT DURING INSTABILITY SURGERY THE TIP OF THE ARTHROPIERCE BROKE INSIDE THE PATIENT, BROKEN PIECES WERE REMOVED WITH FORCEPS AND SUCTION. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT SIGNIFICANT DELAY USING A BACK-UP DEVICE. NO PATIENT INJURY OR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 9560548 ↗MalfunctionMay 14, 2019Break
IT WAS REPORTED THAT DURING PROCEDURE, THE TIP OF THE DEVICE WAS BROKEN OUTSIDE THE PATIENT AND NO DELAY OR PATIENT INJURIES WERE REPORTED. THE PROCEDURE WAS COMPLETED WITH THE COMPETITOR DEVICE.
Patient problems: No Consequences Or Impact To Patient
MAUDE report 8608152 ↗MalfunctionMay 22, 2018Break; Detachment Of Device Component
THE TIP OF THE ARTHROPIERCE BROKE OFF INSIDE THE JOINT WHILE TRYING TO PASS SUTURE DURING AN ARTHROSCOPIC LABRAL REPAIR. THERE WAS ABOUT AN HOUR DELAY. NO PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7535469 ↗MalfunctionFeb 8, 2018Break
IT WAS REPORTED THAT THE MOVABLE BITE BROKE INSIDE THE PATIENT. PIECE WAS RECOVERED. THERE WAS NO DELAY OR PATIENT INJURY REPORTED.
Patient problems: No Known Impact Or Consequence To Patient
MAUDE report 7252880 ↗
Other ACCESSORIES,ARTHROSCOPIC devices
- Arthrex® AR-12990Arthrex, Inc. · Accessories, arthroscopic409 reports
- NA 228302Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC287 reports
- FIRSTPASS 72290128Smith & Nephew, Inc. · ACCESSORIES,ARTHROSCOPIC232 reports
- NA 251723Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC145 reports
- COBRA 3910900093Stryker Corporation · Accessories, arthroscopic138 reports
- FirstPass 72290129Smith & Nephew, Inc. · Accessories, arthroscopic82 reports
- FirstPass 72290130Smith & Nephew, Inc. · Accessories, arthroscopic80 reports
- MITEK 242025Medos International Sàrl · ACCESSORIES,ARTHROSCOPIC61 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page