Accord 71360020Recalls & reported problems
Smith & Nephew, Inc. · ACCORD TENSIONER · Class 2
All models of Accord (14)
Data updated Oct 16, 2026 from official FDA records · Editorial policy
Quick summary
This Smith & Nephew Accord tensioner, a surgical tool used in hip and fracture surgery, has 108 adverse event reports, with no deaths and 11 injuries. Many reports describe the tensioner breaking or chipping during surgery, or its locking handle jamming, and surgeons often switched to a backup device. No recalls name this device.
AI-generated from the FDA records on this page and checked automatically against the numbers below. Read the details before relying on it.
- Adverse event reports
- 108
- Deaths reported
- 0
- Injuries reported
- 11
- Recalls naming it
- 0
Summary
The FDA's MAUDE database contains 108 adverse event reports that list this exact device (DI 03596010485359), received between Aug 6, 2018 and Jan 30, 2026 — classified as 11 injuries, 97 malfunctions.
The most frequently reported specific device problem is “Break” (28 reports).
2 other recalls cover devices cleared under the same FDA submission (K123598); they may or may not involve this specific model.
Reports come from manufacturers, hospitals and patients and are not verified by the FDA. A report does not prove the device caused the event, and counts can't be used to compare devices or estimate failure rates. What reports can and can't tell you · How this data works · Data updated Oct 16, 2026.
Reports received per month
Most reported problems
Device problems
- Break28
- Mechanical Jam18
- Mechanical Problem16
- Connection Problem15
- Defective Device12
- Incomplete or Inadequate Connection7
- Defective Component2
- Fracture2
Patient problems
- No Clinical Signs, Symptoms or Conditions68
- No Known Impact Or Consequence To Patient20
- No Consequences Or Impact To Patient15
- No Patient Involvement11
- No Information2
- Injury1
- Insufficient Information1
- Not Applicable1
Recent adverse event reports
The 25 most recent of 108 reports, in the reporter's own words as published by the FDA.
MalfunctionJan 30, 2026Break
IT WAS REPORTED THAT, DURING AN INTERNAL FIXATION SURGERY ONE (1) ACCORD TENSIONER BROKE. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK-UP DEVICE. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24220149 ↗MalfunctionJan 28, 2026Fracture
IT WAS REPORTED THAT DURING AN INTERNAL FIXATION SURGERY, IT WAS OBSERVED THAT THE TIP OF ONE (1) ACCORD TENSIONER WAS CHIPPED. NO PIECES FELL INSIDE THE PATIENT'S INCISION. THE PROCEDURE WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, USING AN EQUIVALENT S+N BACK UP. NO INJURY WAS REPORTED AS A CONSEQUENCE OF THIS ISSUE. THE CURRENT HEALTH STATUS OF THE PATIENT IS STABLE.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24206621 ↗MalfunctionJan 27, 2026Break
IT WAS REPORTED THAT, DURING A THR SURGERY, ONE (1) ACCORD TENSIONER BROKE. THE PROCEDURE WAS PERFORMED, WITHOUT ANY DELAY, USING AN EQUIVALENT S+N BACK- UP. NO FURTHER COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 24192248 ↗MalfunctionAug 15, 2024Separation Failure; Mechanical Jam; Separation Failure; Mechanical Jam
IT WAS REPORTED THAT, DURING A REVISION INTERNAL FIXATION SURGERY, TWO (2) ACCORD TENSIONER WERE USED BUT THEN THE LOCKING HANDLE BECAME LOCKED AND NOT ABLE TO UNLOCK. MULTIPLE PEOPLE TRIED OF VARIOUS SIZE AND BUILD BUT NO-ONE COULD UNLOCK IT. THE SURGERY WAS COMPLETED, AFTER A NON-SIGNIFICANT DELAY, CHANGING THE SURGICAL TECHNIQUE. NO INJURIES WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 19991310 ↗MalfunctionNov 20, 2023Fracture; Fracture
IT WAS REPORTED THAT DURING SURGERY, ONE (1) ACCORD TENSIONER WAS NOTICED TO BE FRACTURED. ALL THE PIECES WERE REMOVED USING TWEEZERS. THERE ARE NO OPERATIVE REPORTS OR X RAYS. THE PROCEDURE WAS RESUMED, WITHOUT ANY DELAY, USING A S+N BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 18171271 ↗MalfunctionOct 4, 2022Break; Mechanics Altered; Break; Mechanics Altered
IT WAS REPORTED THAT DURING SURGERY, THE CLAMP OF THE ACCORD TENSIONER WOULD TIGHTEN. THE PIECE IS BROKEN. A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NO INJURY TO PATIENT OR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 15535446 ↗InjuryApr 9, 2022Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT, DURING THR REVISION, THE ACCORD TENSIONER WAS GETTING STUCK AND NOT RESETTING. PROCEDURE WAS COMPLETED, WITHOUT ANY DELAY, WITH THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 14068769 ↗MalfunctionJun 22, 2021Incomplete or Inadequate Connection; Incomplete or Inadequate Connection
IT WAS REPORTED THAT AN ACCORD TENSIONER DID NOT TENSION CABLE TIGHT ENOUGH. SURGERY WAS RESUMED, WITHOUT ANY DELAY, WITH A BACK-UP DEVICE. PATIENT WAS NOT INJURED AS CONSEQUENCE OF THIS PROBLEM.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12038842 ↗MalfunctionJun 17, 2021Failure to Reset; Failure to Reset
IT WAS REPORTED THAT AN ACCORD TENSIONER RESET SWITCH DOES NOT WORK CONSISTENTLY. AS THIS WAS NOTICED IN A NON-SURGICAL ENVIRONMENT, THERE WAS NOT ANY PATIENT INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 12018352 ↗MalfunctionJun 4, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT, DURING SURGERY, THE ACCORD TENSIONER WAS NOT FEEDING THE CABLE THROUGH PROPERLY. THE PROCEDURE WAS FINISHED WITH A SMITH AND NEPHEW BACK UP DEVICE WITHOUT DELAYS. THE PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11943746 ↗MalfunctionJun 1, 2021Break; Break
IT WAS REPORTED THAT ONE OF THE METAL PIECES ON THE HANDLE OF AN ACCORD TENSIONER BROKE OFF AND WAS FULLY RECOVERED. IT IS UNKNOWN IF THIS HAPPENED DURING SURGERY OR NOT; THEREFORE, PATIENT INVOLVEMENT IS NOT CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11916522 ↗MalfunctionMay 31, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT, DURING SURGERY, AN ACCORD TENSIONER HAD A WIRE STUCK IN IT. DEVICE OUTSIDE THE PATIENT. IT IS UNKNOWN HOW SURGERY WAS COMPLETED AND IF THERE WAS DELAYED. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11909296 ↗MalfunctionMay 28, 2021Device Damaged by Another Device; Mechanical Jam; Device Damaged by Another Device; Mechanical Jam
IT WAS REPORTED THAT, DURING THR SURGERY, AN ACCORD TENSIONER WAS UNABLE TO BE USED AS INTENDED BECAUSE A WIRE WAS STUCK IN IT. NO SIGNIFICANT SURGICAL DELAY WAS REPORTED, AS A SMITH AND NEPHEW BACKUP DEVICE WAS AVAILABLE. NEITHER PATIENT INJURY NOR OTHER COMPLICATIONS WERE REPORTED. ALL AVAILABLE INFORMATION HAS BEEN DISCLOSED. IF ADDITIONAL INFORMATION SHOULD BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED ACCORDINGLY
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11903087 ↗MalfunctionMay 20, 2021Material Deformation; Material Deformation
IT WAS REPORTED THAT, DURING A THR SURGERY, THE ACCORD TENSIONER DID NOT RETRACT ((B)(4)) AND THE ACCORD DUAL ENDED HEX DRIVER WAS STRIPPED ((B)(4)). THERE WERE NO DELAYS AND THE PROCEDURE WAS FINISHED WITH SMITH AND NEPHEW BACK UP DEVICES. THE PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11853598 ↗MalfunctionMay 19, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT, DURING A THR SURGERY THE ACCORD TENSIONER CABLE DID NOT TIGHTEN WHEN THE SLACK WAS PULLED OUT. THERE WERE NO BROKEN PIECES. THE PROCEDURE WAS SUCCESSFULLY COMPLETED WITHOUT DELAY USING THE SAME DEVICE. PATIENT WAS NOT HARMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11846687 ↗MalfunctionMay 17, 2021Defective Device; Defective Device
IT WAS REPORTED THAT DURING THR SURGERY, ONE ACCORD TENSIONER WAS SLIPPING THE OTHER ACCORD TENSIONER WOULD NOT TENSION ABOVE 50. A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NO INJURY TO PATIENT OR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11832488 ↗MalfunctionMay 17, 2021Device Slipped; Device Slipped
IT WAS REPORTED THAT DURING THR SURGERY, ONE ACCORD TENSIONER WAS SLIPPING THE OTHER ACCORD TENSIONER WOULD NOT TENSION ABOVE 50. A SMITH AND NEPHEW BACK UP DEVICE WAS AVAILABLE. NO INJURY TO PATIENT OR SIGNIFICANT DELAY REPORTED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11832475 ↗MalfunctionMay 11, 2021Break; Break
IT WAS REPORTED THAT A ACCORD TENSIONER WAS BROKEN. IT IS UNKNOWN IF THIS WAS NOTICED DURING SURGERY OR NOT; THEREFORE, PATIENT INVOLVEMENT IS NOT CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11806040 ↗MalfunctionMay 7, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT THE ACCORD TENSIONER GOT STUCK. AS THIS WAS DISCOVERED DURING INSPECTION ON THE WAREHOUSE, NO PATIENT INVOLVEMENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11794721 ↗MalfunctionMay 5, 2021Break; Break
IT WAS REPORTED THAT AN ACCORD TENSIONER BROKE. IT IS UNKNOWN IF THIS WAS NOTICED DURING SURGERY OR NOT; THEREFORE, PATIENT INVOLVEMENT HAS NOT BEEN CONFIRMED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11781283 ↗InjuryApr 26, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING SURGERY AFTER FASTENING IT TO THE FEMUR WITH A PASSER WITH AN ACCORD TENSIONER, THE LOCKING HANDLE DID NOT MOVE WHEN THE FIRST TENSIONER WAS USED TO TIGHTEN IT UP. WHEN THE SECOND TENSIONER WAS USED TO TIGHTEN UP THE LOCKING HANDLE STOP MOVING. THE PROCEDURE WAS COMPLETED WITH 1 HOUR DELAY USING A COMPETITOR BACK-UP DEVICE. NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11722534 ↗InjuryApr 26, 2021Mechanical Jam; Mechanical Jam
IT WAS REPORTED THAT DURING SURGERY AFTER FASTENING IT TO THE FEMUR WITH A PASSER WITH AN ACCORD TENSIONER, THE LOCKING HANDLE DID NOT MOVE WHEN THE FIRST TENSIONER WAS USED TO TIGHTEN IT UP. WHEN THE SECOND TENSIONER WAS USED TO TIGHTEN UP THE LOCKING HANDLE STOP MOVING. THE PROCEDURE WAS COMPLETED WITH 1 HOUR DELAY USING A COMPETITOR BACK-UP DEVICE. NO PATIENT INJURY.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11722533 ↗InjuryApr 23, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT, DURING A THR PROCEDURE, BOTH TENSIONERS IN THE SET WEREN'T TIGHTEN CORRECTLY AND NEITHER WOULD FEED THE CABLE CORRECTLY. NEED TO BE REPLACED. IT IS UNKNOWN IF THERE WAS A DELAY. PROCEDURE WAS COMPLETED WITHOUT A CABLE. INSTRUMENTS WERE OUTSIDE THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11720927 ↗InjuryApr 23, 2021Connection Problem; Connection Problem
IT WAS REPORTED THAT, DURING A THR PROCEDURE, BOTH TENSIONERS IN THE SET WEREN'T TIGHTEN CORRECTLY AND NEITHER WOULD FEED THE CABLE CORRECTLY. NEED TO BE REPLACED. IT IS UNKNOWN IF THERE WAS A DELAY. PROCEDURE WAS COMPLETED WITHOUT A CABLE. INSTRUMENTS WERE OUTSIDE THE PATIENT.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11720917 ↗MalfunctionApr 21, 2021Thickening of Material; Thickening of Material
IT WAS REPORTED THAT DURING AN INSPECTION, IT WAS NOTICED THAT AN ACCORD TENSIONER HAD WEAR AND TEAR PROVOKED BY USE AND IT WAS STARTING TO STICK. NO CASE INVOLVED.
Patient problems: No Clinical Signs, Symptoms or Conditions
MAUDE report 11704305 ↗
Recalls under the same FDA submission
These recalls involve devices cleared under K123598. One submission often covers many models, so check the product and lot details in each record.
- Z-0730-2019 · Jan 7, 2019 · TerminatedSL PLUS MIA DOUBLE OFFSET ADAPTER RIGHT 60/25MM, S&N 75004613
- Z-1008-2017 · Nov 16, 2016 · TerminatedREFLECTION(R) DEAD BLOW MALLET, REF 71362106, QTY: (1), NON-STERILE, Smith & Nephew, Inc,. Memphis, TN 38116
Other Prosthesis, hip, femoral component, cemented, metal devices
- GENESIS II 71440005Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal599 reports
- NA 71364073Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal436 reports
- Peri-Loc 71173377Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal401 reports
- R3 71368569Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal398 reports
- NA 71364006Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal197 reports
- NA 71364011Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal176 reports
- Accord 71360046Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal171 reports
- REFLECTION 71362279Smith & Nephew, Inc. · Prosthesis, hip, femoral component, cemented, metal155 reports
Sources and updates
- Device record: FDA GUDID, file dated Oct 13, 2026 · View this device in AccessGUDID ↗
- Adverse event reports: FDA MAUDE, file dated Oct 14, 2026 · each report above links to its original FDA record
- Recalls: FDA Medical Device Recalls, file dated Oct 16, 2026
- Page last rebuilt: Oct 16, 2026. Data downloaded through openFDA, which notes it is not validated for clinical or regulatory use.
How reports and recalls are matched · Editorial policy · Report an error on this page